Flow rates through intravenous access devices: an in vitro study.

Saleem I Khoyratty1, Pushpaj R Gajendragadkar2, Kiran Polisetty3, Sue Ward3, Tim Skinner3, Parag R Gajendragadkar4.   

Abstract

STUDY
OBJECTIVE: Fluid administration using intravenous (IV) access devices is required in many settings. There are a lack of quantitative data comparing traditional cannulas and modern access devices. We aimed to investigate flow rates through modern intravenous access devices using an in vitro system.
DESIGN: This is an experimental study. SETTING AND MEASUREMENTS: Rates of flow of intravenous fluids (crystalloid and colloid) were measured through various access devices using a uroflowmeter. Standardized conditions and repeat measurements ensured validity. Fluid was administered with or without the addition of a pressure bag and needle-free valve. MAIN
RESULTS: Increasing the size of cannulas improved flow. Fourteen-gauge cannulas had significantly higher mean flow rates compared to 14G central venous lines in all conditions (136% higher with no pressure bag/valve; 95% CI, +130% to +152%; P < .001). Both the emergency infusion device and rapid infusion catheter produced significantly increased mean flows compared to a 14G cannula (12% higher for emergency infusion catheter; 95% CI, +7% to +15%; P = .008, and 15% higher for rapid infusion catheter; 95% CI, +12% to +21%; P = .004). The needle-free valve significantly impaired flow on 16G and wider IV access devices (36% lower with no pressure bag using 14G cannula; 95% CI, -29% to -46%; P = .003), but flow reductions in narrower IV access were insignificant. Pressure bags significantly improved flow in all devices, in all combinations.
CONCLUSIONS: Flow rates in IV devices can be maximized by pressure bag use and removal of needle-free valves. The rapid infusion catheter and emergency infusion catheter allow some increase in flow over a 14G cannula. Familiarity with varying flow rates across IV access devices could better inform clinical decisions. Crown
Copyright © 2016. Published by Elsevier Inc. All rights reserved.

Keywords:  Catheters; Drug administration routes; Drug therapy; Equipment and supplies; Therapeutics; Vascular access devices

Mesh:

Substances:

Year:  2016        PMID: 27185686     DOI: 10.1016/j.jclinane.2016.01.048

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  3 in total

1.  Piezoresistor-Embedded Multifunctional Magnetic Microactuators for Implantable Self-Clearing Catheter.

Authors:  Qi Yang; Albert Lee; R Timothy Bentley; Hyowon Lee
Journal:  IEEE Sens J       Date:  2019-02-15       Impact factor: 3.301

2.  Comparison of the performance of battery-operated fluid warmers.

Authors:  Amit Lehavi; Avraham Yitzhak; Refael Jarassy; Rami Heizler; Yeshayahu Shai Katz; Aeyal Raz
Journal:  Emerg Med J       Date:  2018-06-07       Impact factor: 2.740

3.  The Effects of Intravenous Fluid Viscosity on the Accuracy of Intravenous Infusion Flow Regulators.

Authors:  Eunji Ko; Yeon Jae Song; Kwanyoung Choe; Yongdoo Park; Sung Yang; Choon Hak Lim
Journal:  J Korean Med Sci       Date:  2022-03-07       Impact factor: 2.153

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.