Abstract
OBJECTIVE: To examine the ability to use Quantitative Ultrasonography (QUS) densitometer for screening of osteoporosis and osteopenia by comparing QUS values obtained at the calcaneus region to bone mineral density (BMD) values measured at the spine and the neck of the femur using Dual Energy X-ray Absorbemetry (DXA).Entities:
Keywords: BMD; DXA; QUS; SI; Saudi Female; T-score
Year: 2016 PMID: 27182248 PMCID: PMC4859031 DOI: 10.12669/pjms.322.9757
Source DB: PubMed Journal: Pak J Med Sci ISSN: 1681-715X Impact factor: 1.088
Reliability and precision of the three parameters measured by QUS the calcaneus area.
| Parameter | Variability within subjects (MSw) | Variability between subjects (Msb) | Significance | Coefficient of variation %CV | Reliability R% | Precision |
|---|---|---|---|---|---|---|
| SI | 5.050 | 1047.102 | P<0.001 | 2.176% | 98.6 | 1.7743 |
| T- score | 0.071 | 4.203 | P<0.001 | -1.396% | 95.1 | 1.7779 |
Descriptive measures mean and (SD) for age, height, weight, BMI, QUS parameter in the calcaneus area and DXA parameters for both the spine and the neck of the femur (No of subjects =101).
| Age (Yrs.) | Height (m) | Weight (kg) | BMI | SICalcaneus | T-scorecalcaneus | BMDSpine | T-scoreSpine | BMDN.femur | T-scoreN.femur |
|---|---|---|---|---|---|---|---|---|---|
| 21.35 (0.83) | 1.59 (0.07) | 56.04 (9.47) | 22.27 (3.65) | 93.22 (19.4) | -0.48 (1.11) | 1.1 (0.2) | - 0.77 (1.06) | 0.97 (0.18) | -0.65 (0.88) |
Pearson correlation coefficient of QUS and DXA parameters.
| SI | T-scorecalcaneus | BMDSpine | T-scoreSpine | BMDN.femur | T-scoreN.femur | |
|---|---|---|---|---|---|---|
| T-scoreCalcaneus | 0.79 | |||||
| BMDSpine | 0.29 | 0.29 | ||||
| T-scoreSpine | 0.49 | 0.5 | 0.78 | |||
| BMDN,femur | 0.25 | 0.32 | 0.47 | 0.55 | ||
| T-scoreN.femur | 0.5 | 0.58 | 0.63 | 0.77 | 0.63 |
Correlation is significant at the 0.05 level.
Correlation is significant at the 0.01 level.