Literature DB >> 27180127

Evaluating the utility of "3D Slicer" as a fast and independent tool to assess intrafractional organ dose variations in gynecological brachytherapy.

Zahra Siavashpour1, Mahmoud Reza Aghamiri2, Ramin Jaberi3, Hamid Reza Dehghan-Manshadi4, Mahbod Sedaghat5, Christian Kirisits6.   

Abstract

PURPOSE: To demonstrate the utility of "3D Slicer" for easy treatment verification by comparing dose-volume histograms (DVHs) calculated on pretreatment and posttreatment images. METHODS AND MATERIALS: Thirty cervical cancer patients were CT scanned twice: first for treatment planning and a second time after the dose delivery. The initial plan was manually duplicated on the posttreatment image set in Flexiplan treatment planning system, and DVH parameters were calculated. Pretreatment and posttreatment images, organ structures, and plan data were exported from the treatment planning system to 3D Slicer to validate DVH parameter calculation with 3D Slicer. The gamma analysis was used to compare Flexiplan and 3D Slicer DVHs. Posttreatment images were rigidly fused on the initial CT to automatically transfer the data of the pretreatment plan onto the posttreatment images. DVH parameters were calculated in 3D Slicer for both image sets, and their relative variations were compared.
RESULTS: In calculating DVH parameter variations, no significant differences were observed between Flexiplan and 3D Slicer. Where the registration accuracy was better than 0.03, they returned similar results for D2 cm(3) of bladder, rectum, and sigmoid. Mean and standard deviation of DVH parameters were calculated on pretreatment and posttreatment images for several organs; both the manually duplicated plan and the automatically registered plan in SlicerRT returned comparable relative variations of these parameters. For 88% of the organs, more than 95% of the DVH dose bins passed the gamma analysis.
CONCLUSIONS: We tested an automated DVH assessment method with an imaging freeware, 3D Slicer, for use in image-guided adaptive brachytherapy. SlicerRT is a viable verification tool to report and detect DVH variations between different contoured images series.
Copyright © 2016 American Brachytherapy Society. All rights reserved.

Entities:  

Keywords:  3D Slicer; IGABT; Intrafractional dose variations

Mesh:

Year:  2016        PMID: 27180127     DOI: 10.1016/j.brachy.2016.03.009

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  5 in total

1.  18F-FDG PET/CT radiomic predictors of pathologic complete response (pCR) to neoadjuvant chemotherapy in breast cancer patients.

Authors:  Panli Li; Xiuying Wang; Chongrui Xu; Cheng Liu; Chaojie Zheng; Michael J Fulham; Dagan Feng; Lisheng Wang; Shaoli Song; Gang Huang
Journal:  Eur J Nucl Med Mol Imaging       Date:  2020-01-25       Impact factor: 9.236

2.  Evaluation of deformable image registration algorithm for determination of accumulated dose for brachytherapy of cervical cancer patients.

Authors:  Reza Mohammadi; Seied Rabi Mahdavi; Ramin Jaberi; Zahra Siavashpour; Leila Janani; Ali Soleimani Meigooni; Reza Reiazi
Journal:  J Contemp Brachytherapy       Date:  2019-10-30

3.  3D in vivo dosimetry of HDR gynecological brachytherapy using micro silica bead TLDs.

Authors:  Ramin Jaberi; Somayyeh Babaloui; Zahra Siavashpour; Maryam Moshtaghi; Alireza Shirazi; Musa Joya; Mohammad Hadi Gholami; Shakardokht Jafari
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

Review 4.  Dose Summation Strategies for External Beam Radiation Therapy and Brachytherapy in Gynecologic Malignancy: A Review from the NRG Oncology and NCTN Medical Physics Subcommittees.

Authors:  Hayeon Kim; Yongsook C Lee; Stanley H Benedict; Brandon Dyer; Michael Price; Yi Rong; Ananth Ravi; Eric Leung; Sushil Beriwal; Mark E Bernard; Jyoti Mayadev; Jessica R L Leif; Ying Xiao
Journal:  Int J Radiat Oncol Biol Phys       Date:  2021-06-17       Impact factor: 7.038

5.  Artificial neural network based gynaecological image-guided adaptive brachytherapy treatment planning correction of intra-fractional organs at risk dose variation.

Authors:  Ramin Jaberi; Zahra Siavashpour; Mahmoud Reza Aghamiri; Christian Kirisits; Reza Ghaderi
Journal:  J Contemp Brachytherapy       Date:  2017-12-30
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.