Literature DB >> 27171358

A comprehensive spectrometry study of a stray neutron radiation field in scanning proton therapy.

Vladimir Mares1, Maite Romero-Expósito, Jad Farah, Sebastian Trinkl, Carles Domingo, Martin Dommert, Liliana Stolarczyk, Laurent Van Ryckeghem, Marek Wielunski, Pawel Olko, Roger M Harrison.   

Abstract

The purpose of this study is to characterize the stray neutron radiation field in scanning proton therapy considering a pediatric anthropomorphic phantom and a clinically-relevant beam condition. Using two extended-range Bonner sphere spectrometry systems (ERBSS), Working Group 9 of the European Radiation Dosimetry Group measured neutron spectra at ten different positions around a pediatric anthropomorphic phantom irradiated for a brain tumor with a scanning proton beam. This study compares the different systems and unfolding codes as well as neutron spectra measured in similar conditions around a water tank phantom. The ten spectra measured with two ERBSS systems show a generally similar thermal component regardless of the position around the phantom while high energy neutrons (above 20 MeV) were only registered at positions near the beam axis (at 0°, 329° and 355°). Neutron spectra, fluence and ambient dose equivalent, H (*)(10), values of both systems were in good agreement (<15%) while the unfolding code proved to have a limited effect. The highest H (*)(10) value of 2.7 μSv Gy(-1) was measured at 329° to the beam axis and 1.63 m from the isocenter where high-energy neutrons (E  ⩾  20 MeV) contribute with about 53%. The neutron mapping within the gantry room showed that H (*)(10) values significantly decreased with distance and angular position with respect to the beam axis dropping to 0.52 μSv Gy(-1) at 90° and 3.35 m. Spectra at angles of 45° and 135° with respect to the beam axis measured here with an anthropomorphic phantom showed a similar peak structure at the thermal, fast and high energy range as in the previous water-tank experiments. Meanwhile, at 90°, small differences at the high-energy range were observed. Using ERBSS systems, neutron spectra mapping was performed to characterize the exposure of scanning proton therapy patients. The ten measured spectra provide precise information about the exposure of healthy organs to thermal, epithermal, evaporation and intra-nuclear cascade neutrons. This comprehensive spectrometry analysis can also help in understanding the tremendous literature data based rem-counters while also being of great value for general neutron shielding and radiation safety studies.

Entities:  

Mesh:

Year:  2016        PMID: 27171358     DOI: 10.1088/0031-9155/61/11/4127

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  5 in total

1.  Excessive applicator radiation leakage for a common therapeutic kilovoltage system.

Authors:  Brad Beeksma; Joerg Lehmann
Journal:  Br J Radiol       Date:  2018-11-15       Impact factor: 3.039

2.  Peripheral Organ Equivalent Dose Estimation Procedure in Proton Therapy.

Authors:  Carles Domingo; Juan Ignacio Lagares; Maite Romero-Expósito; Beatriz Sánchez-Nieto; Jaime J Nieto-Camero; Jose Antonio Terrón; Leticia Irazola; Alexandru Dasu; Francisco Sánchez-Doblado
Journal:  Front Oncol       Date:  2022-05-25       Impact factor: 5.738

3.  Modelling and measurements of distributions in an adult human phantom undergoing proton scanning beam radiotherapy: lung- and prostate-located tumours.

Authors:  Monika Puchalska
Journal:  Radiat Environ Biophys       Date:  2021-03-02       Impact factor: 1.925

4.  Neutron Radiation Dose Measurements in a Scanning Proton Therapy Room: Can Parents Remain Near Their Children During Treatment?

Authors:  Vladimir Mares; Jad Farah; Marijke De Saint-Hubert; Szymon Domański; Carles Domingo; Martin Dommert; Magdalena Kłodowska; Katarzyna Krzempek; Michał Kuć; Immaculada Martínez-Rovira; Edyta Michaś; Natalia Mojżeszek; Łukasz Murawski; Ondrej Ploc; Maite Romero-Expósito; Marco Tisi; François Trompier; Olivier Van Hoey; Laurent Van Ryckeghem; Marek Wielunski; Roger M Harrison; Liliana Stolarczyk; Pawel Olko
Journal:  Front Oncol       Date:  2022-07-14       Impact factor: 5.738

5.  Simulation and experimental verification of ambient neutron doses in a pencil beam scanning proton therapy room as a function of treatment plan parameters.

Authors:  Olivier Van Hoey; Liliana Stolarczyk; Jan Lillhök; Linda Eliasson; Natalia Mojzeszek; Malgorzata Liszka; Ali Alkhiat; Vladimir Mares; François Trompier; Sebastian Trinkl; Immaculada Martínez-Rovira; Maite Romero-Expósito; Carles Domingo; Ondrej Ploc; Roger Harrison; Pawel Olko
Journal:  Front Oncol       Date:  2022-09-08       Impact factor: 5.738

  5 in total