Radiolucent 4D Ultrasound Imaging: System Design and Application to Radiotherapy Guidance.

Jeffrey Schlosser, Dimitre Hristov.   

Abstract

Four-dimensional (4D) ultrasound (US) is an attractive modality for image guidance due to its real-time, non-ionizing, volumetric imaging capability with high soft tissue contrast. However, existing 4D US imaging systems contain large volumes of metal which interfere with diagnostic and therapeutic ionizing radiation in procedures such as CT imaging and radiation therapy. This study aimed to design and characterize a novel 4D Radiolucent Remotely-Actuated UltraSound Scanning (RRUSS) device that overcomes this limitation. In a phantom, we evaluated the imaging performance of the RRUSS device including frame rate, resolution, spatial integrity, and motion tracking accuracy. To evaluate compatibility with radiation therapy workflow, we evaluated device-induced CT imaging artifacts, US tracking performance during beam delivery, and device compatibility with commercial radiotherapy planning software. The RRUSS device produced 4D volumes at 0.1-3.0 Hz with 60° lateral field of view (FOV), 50° maximum elevational FOV, and 200 mm maximum depth. Imaging resolution (-3 dB point spread width) was 1.2-7.9 mm at depths up to 100 mm and motion tracking accuracy was ≤ 0.3±0.5 mm. No significant effect of the RRUSS device on CT image integrity was found, and RRUSS device performance was not affected by radiotherapy beam exposure. Agreement within ±3.0% / 2.0 mm was achieved between computed and measured radiotherapy dose delivered directly through the RRUSS device at 6 MV and 15 MV. In vivo liver, kidney, and prostate images were successfully acquired. Our investigations suggest that a RRUSS device can offer non-interfering 4D guidance for radiation therapy and other diagnostic and therapeutic procedures.

Entities:  

Mesh:

Year:  2016        PMID: 27164579     DOI: 10.1109/TMI.2016.2559499

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  8 in total

1.  Analysis and optimization of the robot setup for robotic-ultrasound-guided radiation therapy.

Authors:  Matthias Schlüter; Stefan Gerlach; Christoph Fürweger; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2019-06-06       Impact factor: 2.924

2.  Impact of robotic ultrasound image guidance on plan quality in SBRT of the prostate.

Authors:  Stefan Gerlach; Ivo Kuhlemann; Floris Ernst; Christoph Fürweger; Alexander Schlaefer
Journal:  Br J Radiol       Date:  2017-07-27       Impact factor: 3.039

3.  Robotic ultrasound-guided SBRT of the prostate: feasibility with respect to plan quality.

Authors:  Stefan Gerlach; Ivo Kuhlemann; Philipp Jauer; Ralf Bruder; Floris Ernst; Christoph Fürweger; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-07-12       Impact factor: 2.924

4.  Robotic intrafractional US guidance for liver SABR: System design, beam avoidance, and clinical imaging.

Authors:  Jeffrey Schlosser; Ren Hui Gong; Ralf Bruder; Achim Schweikard; Sungjune Jang; John Henrie; Aya Kamaya; Albert Koong; Daniel T Chang; Dimitre Hristov
Journal:  Med Phys       Date:  2016-11       Impact factor: 4.071

5.  Case Report: Treatment Planning Study to Demonstrate Feasibility of Transthoracic Ultrasound Guidance to Facilitate Ventricular Tachycardia Ablation With Protons.

Authors:  Rosalind Perrin; Patrick Maguire; Adriano Garonna; Georg Weidlich; Shelley Bulling; Marie Fargier-Voiron; Cedric De Marco; Eleonora Rossi; Mario Ciocca; Viviana Vitolo; Alfredo Mirandola
Journal:  Front Cardiovasc Med       Date:  2022-05-04

6.  Systematic analysis of volumetric ultrasound parameters for markerless 4D motion tracking.

Authors:  Johanna Sprenger; Marcel Bengs; Stefan Gerlach; Maximilian Neidhardt; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-05-21       Impact factor: 3.421

7.  AI-based optimization for US-guided radiation therapy of the prostate.

Authors:  Stefan Gerlach; Theresa Hofmann; Christoph Fürweger; Alexander Schlaefer
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-05-20       Impact factor: 3.421

Review 8.  The Use of Ultrasound Imaging in the External Beam Radiotherapy Workflow of Prostate Cancer Patients.

Authors:  Saskia M Camps; Davide Fontanarosa; Peter H N de With; Frank Verhaegen; Ben G L Vanneste
Journal:  Biomed Res Int       Date:  2018-01-24       Impact factor: 3.411

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.