A Comparison of a New Ultrasound-Based Whole Blood Viscoelastic Test (SEER Sonorheometry) Versus Thromboelastography in Cardiac Surgery.

Penny S Reynolds1, Paul Middleton, Harry McCarthy, Bruce D Spiess.   

Abstract

BACKGROUND: Viscoelastic thromboelastography tests such as TEG™ are now routine for assessing the coagulation status of cardiac surgery patients. We compared TEG™ with a new technology, sonic estimation of elasticity via resonance (SEER) sonorheometry, to compare measures of coagulation dynamics of whole blood and assess its potential for rapid, near-point-of-care monitoring of hemostasis during cardiac surgery.
METHODS: Whole blood coagulation assessment of a prospective cohort of 50 cardiac surgery patients was performed using SEER sonorheometry and blood samples collected at 4 time points during cardiac surgery: baseline before anesthetic induction, during cardiopulmonary bypass on rewarming, 10 minutes after heparin reversal by protamine, and on patient transfer to the intensive care unit. Clot strength trajectories (G, measured by TEG™; and clot stiffness measured by SEER sonorheometry) and clot times were assessed by repeated-measures mixed models. Strength of association between the 2 methods (clot stiffness and clot times) was assessed using a modified Bland-Altman method for repeated measures; Deming (orthogonal) regression was used to quantify method concordance (constant and proportional bias).
RESULTS: Clot strength/stiffness and clot time measures for both techniques showed similar tracking of trajectories. Strength of association between methods was acceptable (correlations, 0.8-0.9); however, Deming regression detected substantial deviation (bias) between techniques. SEER clot stiffness values averaged approximately 10 hPa higher than corresponding G at all time points. Reaction time (TEG™) was 1 to 2.5 minutes longer than corresponding clot times (SEER). Laboratory times (from sample drop-off to results) were substantially less for SEER sonorheometry (median time, 11-17 minutes) compared with nonautomated kaolin TEG™ (median time, 42 minutes).
CONCLUSIONS: Currently, no viscoelastic hemostatic analyzer system can be considered the "gold standard"; therefore, differences observed between TEG™ and SEER are of importance only because they show that the methods are not perfectly substitutable. Measurements of clot stiffness determined by the 2 methods were correlated but not interchangeable. Reasons for discrepancies include the substantial difference in the physical methods of inducing coagulation activation in samples and the mathematical assumptions underlying calculations of G. Future studies will be required to evaluate SEER sonorheometry's abilities to identify bleeding diatheses (sensitivity/specificity) or to develop treatment algorithms based on the new tests.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27159065     DOI: 10.1213/ANE.0000000000001362

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  7 in total

Review 1.  Viscoelastic Testing in Pediatric Mechanical Circulatory Support.

Authors:  Katherine Regling; Arun Saini; Katherine Cashen
Journal:  Front Med (Lausanne)       Date:  2022-05-06

2.  Hemostatis Analyzer-Supported Hemotherapy Algorithm in Cardiac Surgery: Protocol for a Randomized Controlled Monocentric Trial.

Authors:  Sophie Michel; Florian Piekarski; Jan-Hendrik Fischer; Vanessa Hettler; Elisabeth Hannah Adam; Lars Holzer; Gösta Lotz; Thomas Walther; Kai Zacharowski; Florian Jürgen Raimann
Journal:  JMIR Res Protoc       Date:  2020-04-21

Review 3.  Hemostatic Testing in Critically Ill Infants and Children.

Authors:  Alison B Nair; Robert I Parker
Journal:  Front Pediatr       Date:  2021-01-08       Impact factor: 3.418

Review 4.  Viscoelastic Hemostatic Assays: A Primer on Legacy and New Generation Devices.

Authors:  Oksana Volod; Connor M Bunch; Nuha Zackariya; Ernest E Moore; Hunter B Moore; Hau C Kwaan; Matthew D Neal; Mahmoud D Al-Fadhl; Shivani S Patel; Grant Wiarda; Hamid D Al-Fadhl; Max L McCoy; Anthony V Thomas; Scott G Thomas; Laura Gillespie; Rashid Z Khan; Mahmud Zamlut; Peter Kamphues; Dietmar Fries; Mark M Walsh
Journal:  J Clin Med       Date:  2022-02-07       Impact factor: 4.241

5.  Comparison of the resonance sonorheometry based Quantra® system with rotational thromboelastometry ROTEM® sigma in cardiac surgery - a prospective observational study.

Authors:  Werner Baulig; Samira Akbas; Michael Thomas Ganter; Oliver M Theusinger; Philipp K Schütt; Wolfgang Keul; Marija Jovic; Pascal Berdat; Stefanie von Felten; Klaus Steigmiller
Journal:  BMC Anesthesiol       Date:  2021-10-28       Impact factor: 2.217

6.  Comparison of Coagulation Parameters in Arterial and Venous Blood in Cardiac Surgery Measured Using the Quantra System.

Authors:  Danja S Groves; Deborah A Winegar; Lucas G Fernandez; Julie L Huffmyer; Francesco Viola
Journal:  J Cardiothorac Vasc Anesth       Date:  2018-08-31       Impact factor: 2.628

7.  Initial clinical experience with the Quantra QStat System in adult trauma patients.

Authors:  Edward A Michelson; Michael W Cripps; Bradford Ray; Deborah A Winegar; Francesco Viola
Journal:  Trauma Surg Acute Care Open       Date:  2020-11-03
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.