Alejandro Rendon1, Tahir Hamid2, Gajen Kanaganayagam3, Devinda Karunaratne2, Vaikom S Mahadevan4.
Abstract
OBJECTIVE: Transcatheter aortic valve replacement (TAVR) has been established as an alternative therapy for patients with severe aortic stenosis who are unfit for the surgical aortic valve replacements. Pre and periprocedural imaging for the TAVR procedure is the key to procedural success. Currently transesophageal echocardiography (TOE), including real-time three-dimensional (RT-3D) imaging TOE, has been used for peri-interventional monitoring and guidance for TAVR. We describe our initial experience with real-time three-dimensional intracardiac echocardiography (RT-3DICE), imaging technology for the use in the TAVR procedure.Entities:
Year: 2016 PMID: 27158522 PMCID: PMC4854149 DOI: 10.1136/openhrt-2015-000316
Source DB: PubMed Journal: Open Heart ISSN: 2053-3624
3D-ICE results, linear and planimetry measurements were obtain in PLAX and SAX (equivalent to 130–150 TOE views)
| RT-3D ICE | MSCT | ||||||
|---|---|---|---|---|---|---|---|
| Case | Ao annulus (cm) | Ao sinus (cm) | Ao sinus (cm) SAX | Annulus area | Ao annulus (cm) | Ao sinus (cm) | Annulus perimeter (mm2) |
| 1 | 2 | 3.2 | 3.0×2.0 | 500 | 3.0×2.2 | 3.2×2.9 | 513 |
| 2 | 2.4 | 3.1 | 3.0×3.0 | 510 | 2.3×2.2 | 3.1×3.0 | 530 |
| 3 | 2.2 | 3 | 3.3×2.2 | 527 | 3.3×2.2 | 3.7×3.0 | 554 |
| 4 | 2.2 | 3.6 | 3.0×3.0 | 498 | 2.8×2.1 | 3.3×3.2 | 513 |
| 5 | 2.3 | 3.6 | 3.6×3.2 | 580 | 3.0×2.6 | 3.7×3.2 | 629 |
MSCT,multislice CT; RT-3D ICE,real-time three-dimensional intracardiac echocardiography; TOE, two-dimensional transoesophageal echocardiogram.
Figure 1The image shows intracardiac echocardiography (ICE) probe in a TAVI procedure. TAVI, transcatheter aortic valve implantation.
Figure 2Pre-TAVI procedure annular area. 3D-ICE annulus area by planimetry versus MSCT annulus area. TAVI, transcatheter aortic valve implantation; MSCT, multislice CT; 3D-ICE, three-dimensional intracardiac echocardiography.