Evaluation of Three Hemoglobin A1c Point-of-Care Instruments.

Marieke Criel, Stijn Jonckheere, Michel Langlois.   

Abstract

BACKGROUND: Hemoglobin A1c is currently used both for monitoring and diagnosing patients with diabetes mellitus. Point-of-care (POC) instruments have been proposed to improve glycemic control of patients with diabetes by providing a rapid result if the analytical performance is acceptable. However, the performance of most current POC devices is unsatisfactory. In this study, we evaluated 3 POC devices: cobas b101 (Roche), Afinion (Alere) and B-Analyst (Menarini).
METHODS: The CLSI protocols EP-5 and EP-9 were applied to investigate precision, accuracy, and bias. Bias was compared with an ion-exchange chromatography method (Adams Arkray HA-8160, Menarini) of which the accuracy was assessed by three secondary reference methods (SRMs) and a capillary zone electrophoresis-based laboratory method (Capillarys Flex Piercing 2, Sebia).
RESULTS: Using the IFCC units (mmol/mol), total CV's for cobas b101, Afinion and B-Analyst ranged from 2.0 to 3.9%, from 2.3 to 3.7% and from 2.3 to 2.9%, respectively, compared to a total CV ranging from 0.9 to 1.2% for the Adams Arkray HA-8160. A high correlation was found between the HA-8160 and the three POC devices (r = 0.992, r = 0.968, r = 0.998 for cobas b101, Afinion and B-Analyst, respectively). However, all three showed a significant bias of -2.2 mmol/mol, -2.2 mmol/mol and 0.5 mmol/mol, respectively.
CONCLUSIONS: Only the B-Analyst met the current HbA1c quality specifications for precision, whereas cobas b101 and Afinion failed on the low level. Significant bias was observed for all three POC instruments. As low precision and bias are required for both monitoring and diagnosis of diabetes, we believe that POC devices should be carefully selected.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27156315     DOI: 10.7754/clin.lab.2015.150301

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  3 in total

1.  Analysis: Investigating the quality of POCT devices for HbA1c, what are our next steps?

Authors:  Erna Lenters-Westra; Emma English
Journal:  J Diabetes Sci Technol       Date:  2019-05-21

2.  Evaluation of Point of Care Analyzer for Hemoglobin A1c.

Authors:  Anne Marie Dupuy; Stéphanie Badiou; Anne Sophie Bargnoux; Jennifer Russelo; Jean Paul Cristol
Journal:  J Diabetes Sci Technol       Date:  2018-11-27

Review 3.  Impact of HbA1c Testing at Point of Care on Diabetes Management.

Authors:  Oliver Schnell; J Benjamin Crocker; Jianping Weng
Journal:  J Diabetes Sci Technol       Date:  2016-11-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.