Timon Geib1, Lekha Sleno2, Rabea A Hall3, Caroline S Stokes3, Dietrich A Volmer4.
Abstract
We describe a systematic comparison of high and low resolution LC-MS/MS assays for quantification of 25-hydroxyvitamin D3 in human serum. Identical sample preparation, chromatography separations, electrospray ionization sources, precursor ion selection, and ion activation were used; the two assays differed only in the implemented final mass analyzer stage; viz. high resolution quadrupole-quadrupole-time-of-flight (QqTOF) versus low resolution triple quadrupole instruments. The results were assessed against measured concentration levels from a routine clinical chemiluminescence immunoassay. Isobaric interferences prevented the simple use of TOF-MS spectra for extraction of accurate masses and necessitated the application of collision-induced dissociation on the QqTOF platform. The two mass spectrometry assays provided very similar analytical figures of merit, reflecting the lack of relevant isobaric interferences in the MS/MS domain, and were successfully applied to determine the levels of 25-hydroxyvitamin D for patients with chronic liver disease. Graphical Abstract ᅟ.Entities:
Keywords: Collision-induced dissociation; Electrospray ionization; High resolution mass spectrometry; Isobaric interferences; Vitamin D
Mesh:
Substances:
Year: 2016 PMID: 27154021 DOI: 10.1007/s13361-016-1412-2