Luminescence imaging of water during carbon-ion irradiation for range estimation.

Seiichi Yamamoto1, Masataka Komori1, Takashi Akagi2, Tomohiro Yamashita2, Shuji Koyama1, Yuki Morishita1, Eri Sekihara1, Toshiyuki Toshito3.   

Abstract

PURPOSE: The authors previously reported successful luminescence imaging of water during proton irradiation and its application to range estimation. However, since the feasibility of this approach for carbon-ion irradiation remained unclear, the authors conducted luminescence imaging during carbon-ion irradiation and estimated the ranges.
METHODS: The authors placed a pure-water phantom on the patient couch of a carbon-ion therapy system and measured the luminescence images with a high-sensitivity, cooled charge-coupled device camera during carbon-ion irradiation. The authors also carried out imaging of three types of phantoms (tap-water, an acrylic block, and a plastic scintillator) and compared their intensities and distributions with those of a phantom containing pure-water.
RESULTS: The luminescence images of pure-water phantoms during carbon-ion irradiation showed clear Bragg peaks, and the measured carbon-ion ranges from the images were almost the same as those obtained by simulation. The image of the tap-water phantom showed almost the same distribution as that of the pure-water phantom. The acrylic block phantom's luminescence image produced seven times higher luminescence and had a 13% shorter range than that of the water phantoms; the range with the acrylic phantom generally matched the calculated value. The plastic scintillator showed ∼15 000 times higher light than that of water.
CONCLUSIONS: Luminescence imaging during carbon-ion irradiation of water is not only possible but also a promising method for range estimation in carbon-ion therapy.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27147356     DOI: 10.1118/1.4946821

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

Review 1.  Photon counting detectors and their applications ranging from particle physics experiments to environmental radiation monitoring and medical imaging.

Authors:  Ryosuke Ota
Journal:  Radiol Phys Technol       Date:  2021-03-19

2.  Estimation of the fractions of luminescence of water at higher energy than Cerenkov-light threshold for various types of radiation.

Authors:  Yoshiyuki Hirano; Seiichi Yamamoto
Journal:  J Biomed Opt       Date:  2019-06       Impact factor: 3.170

3.  Optical imaging of muons.

Authors:  Seiichi Yamamoto; Kazuhiko Ninomiya; Naritoshi Kawamura; Yoshiyuki Hirano
Journal:  Sci Rep       Date:  2020-11-26       Impact factor: 4.379

4.  Increase in the intensity of an optical signal with fluorescein during proton and carbon-ion irradiation.

Authors:  Seiichi Yamamoto; Takuya Yabe; Takashi Akagi
Journal:  J Appl Clin Med Phys       Date:  2021-06-14       Impact factor: 2.102

5.  Detection of luminescence of radicals from TiO2 plate during alpha particle irradiation.

Authors:  Seiichi Yamamoto
Journal:  J Biomed Opt       Date:  2020-09       Impact factor: 3.170

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.