Literature DB >> 27147321

Technical Note: Dosimetry of Leipzig and Valencia applicators without the plastic cap.

D Granero1, C Candela-Juan2, J Vijande3, F Ballester4, J Perez-Calatayud5, D Jacob6, F Mourtada6.   

Abstract

PURPOSE: High dose rate (HDR) brachytherapy for treatment of small skin lesions using the Leipzig and Valencia applicators is a widely used technique. These applicators are equipped with an attachable plastic cap to be placed during fraction delivery to ensure electronic equilibrium and to prevent secondary electrons from reaching the skin surface. The purpose of this study is to report on the dosimetric impact of the cap being absent during HDR fraction delivery, which has not been explored previously in the literature.
METHODS: geant4 Monte Carlo simulations (version 10.0) have been performed for the Leipzig and Valencia applicators with and without the plastic cap. In order to validate the Monte Carlo simulations, experimental measurements using radiochromic films have been done.
RESULTS: Dose absorbed within 1 mm of the skin surface increases by a factor of 1500% for the Leipzig applicators and of 180% for the Valencia applicators. Deeper than 1 mm, the overdosage flattens up to a 10% increase.
CONCLUSIONS: Differences of treating with or without the plastic cap are significant. Users must check always that the plastic cap is in place before any treatment in order to avoid overdosage of the skin. Prior to skin HDR fraction delivery, the timeout checklist should include verification of the cap placement.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27147321     DOI: 10.1118/1.4944784

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Recommendations of the Spanish brachytherapy group (GEB) of Spanish Society of Radiation Oncology (SEOR) and the Spanish Society of Medical Physics (SEFM) for high-dose rate (HDR) non melanoma skin cancer brachytherapy.

Authors:  S Rodríguez; M Arenas; C Gutierrez; J Richart; J Perez-Calatayud; F Celada; M Santos; A Rovirosa
Journal:  Clin Transl Oncol       Date:  2017-08-14       Impact factor: 3.405

Review 2.  Novel simple templates for reproducible positioning of skin applicators in brachytherapy.

Authors:  Silvia Rodríguez Villalba; Maria Jose Perez-Calatayud; Juan Antonio Bautista; Vicente Carmona; Francisco Celada; Alejandro Tormo; Teresa García-Martinez; José Richart; Manuel Santos Ortega; Magda Silla; Facundo Ballester; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2016-08-09

Review 3.  Commissioning and quality assurance procedures for the HDR Valencia skin applicators.

Authors:  Domingo Granero; Cristian Candela-Juan; Facundo Ballester; Zoubir Ouhib; Javier Vijande; Jose Richart; Jose Perez-Calatayud
Journal:  J Contemp Brachytherapy       Date:  2016-11-04
  3 in total