Sang-Hun Park1, Kyung-Soon Park1, Sung-Am Cho1.
Abstract
PURPOSE: The purpose of this study was to compare the removal torques of a chemically modified SLActive implant and a blasted, laser-treated (BLT) implant, which were soaked in saline for 2 weeks after their surface modifications. The removal torques of the two implants were measured 4 weeks after their implantation into the bone defect area in rabbit tibias with concentrated growth factor (CGF) application.Entities:
Keywords: Concentrated growth factor (CGF); Laser-treated implant; Removal torque; SLActive
Year: 2016 PMID: 27141254 PMCID: PMC4852262 DOI: 10.4047/jap.2016.8.2.110
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Fig. 1In group A, BLT implant was installed on CGF applied bone defect. In group B, BLT implant was installed on non-CGF applied bone defect. In group C, SLActive implant was installed on CGF applied bone defect. In group D, SLActive implant was installed on non-CGF applied bone defect.
Insertion torque
| Group A | Group B | Group C | Group D | |
|---|---|---|---|---|
| 1 | 12.0 | 3.2 | 1.2 | 11.9 |
| 2 | 0.5 | 6.1 | 9.2 | 5.2 |
| 3 | 5.6 | 5.8 | 1.7 | 3.2 |
| 4 | 4.1 | 6.1 | 2.1 | 7.6 |
| 5 | 3.6 | 7.2 | 2.4 | 6.0 |
| 6 | 3.5 | 4.3 | 1.4 | 5.4 |
| 7 | 0.3 | 2.2 | 5.6 | 2.3 |
| 8 | 3.1 | 4.0 | 4.1 | 4.7 |
| mean ± S.D (N/cm) | 4.1 ± 3.7 | 4.9 ± 1.7 | 3.5 ± 2.8 | 5.8 ± 3.0 |
Fig. 2Insertion torque and standard deviation.
Removal torque
| Group A | Group B | Group C | Group D | |
|---|---|---|---|---|
| 1 | 85.0 | 20.0 | 47.0 | 61.4 |
| 2 | 21.6 | 47.2 | 37.9 | 19.2 |
| 3 | 12.0 | 36.8 | 31.2 | 29.7 |
| 4 | 25.7 | 45.5 | 37.7 | 25.2 |
| 5 | 40.8 | 38.9 | 32.0 | 59.2 |
| 6 | 40.0 | 50.7 | 54.5 | 64.8 |
| 7 | 41.6 | 31.7 | 38.7 | 31.2 |
| 8 | 25.8 | 62.7 | 39.7 | 52.4 |
| mean ± S.D (N/cm) | 36.6 ± 22.2 | 41.7 ± 12.9 | 39.8 ± 7.7 | 42.9 ± 18.4 |
(Horizontal line means groups. In group A, BLT implant was installed on CGF applied bone defect. In group B, BLT implant was installed on unfilled bone defect. In group C, SLActive implant was installed on CGF applied bone defect. In group D, SLActive implant was installed on unfilled bone defect. Vertical line means rabbit numbers.)
Fig. 3Removal torque and standard deviation.