| Literature DB >> 27138593 |
Ruth Lapid-Gortzak1,2, Roberto Traversari3, Jan Willem van der Linden4, Sarit Y Lesnik Oberstein5, Oren Lapid6, Reinier O Schlingemann5.
Abstract
The aim of this study is to determine whether the use of a mobile ultra-clean laminar airflow screen reduces the air-borne particle counts in the setting of a simulated procedure of an intra-vitreal injection. A mobile ultra-clean unidirectional airflow (UDF) screen was tested in a simulated procedure for intra-vitreal injections in a treatment room without mechanical ventilation. One UDF was passed over the instrument tray and the surgical area. The concentration of particles was measured in the background, over the instrument table, and next to the ocular area. The degree of protection was calculated at the instrument table and at the surgical site. Use of the UDF mobile screen reduced the mean particle concentration (particles > 0.3 microns) on the instrument table by a factor of at least 100.000 (p < 0.05), and over the patient's eye by at least a factor of 436 (p < 0.05), which in clinical practice translates into significantly reduced air contamination. Mobile UDF screen reduces the mean particle concentration substantially. The mobile UDF screen may therefore allow for a safer procedural environment for ambulatory care procedures such as intra-vitreal injections in treatment rooms.Entities:
Keywords: Contamination; Endophthalmitis prevention; HEPA filters; Intra-vitreal injections; Unidirectional ultra-clean laminar airflow
Mesh:
Substances:
Year: 2016 PMID: 27138593 PMCID: PMC5306071 DOI: 10.1007/s10792-016-0236-1
Source DB: PubMed Journal: Int Ophthalmol ISSN: 0165-5701 Impact factor: 2.031
Fig. 1This figure schematically renders the setup of the measurements in the operating theater. The larger black dot is the site of particle emission and measurement of background particles, while the smaller white dots are the sites where the measurements were done on the instrument table and near the operative site, i.e., the intra-vitreal injection site
Fig. 2The Toul 400 mobile ultra-clean laminar air flow unit. On top of the standard equipped with 4 wheels is the flow unit with the HEPA filter. Positioning of the flow is done by placing the unidirectional laminar flow device adjacent to the area that needs to be treated
The measured degree of protection with the UDF mobile and the 95 % confidential interval during the at rest and operational conditions is shown here
| Particle size | ≥0.3 µm | ≥0.5 µm | ≥1.0 µm | ≥2.5 µm | ≥5.0 µm | ≥10.0 µm |
|---|---|---|---|---|---|---|
| Instrument table (operational, | ||||||
| Mean value | 5.00a,b | 5.00b | 5.00b | 5.00b | 5.00b | 5.00b |
| Standard deviation | ||||||
| 95 % CI upper boundary | ||||||
| 95 % CI lower boundary | ||||||
| Instrument table (at rest, | ||||||
| Mean value | 5.25 | 5.00b | 5.00b | 5.00b | 5.00b | 5.00b |
| Standard deviation | 0.69 | |||||
| 95 % CI upper boundary | 6.34 | |||||
| 95 % CI lower boundary | 4.15 | |||||
| Ocular area (operational, | ||||||
| Mean value | 2.64 | 2.70 | 3.39 | 4.31 | 4.57a | 5.00b |
| Standard deviation | 0.45 | 0.50 | 1.39 | 1.46 | 1.37 | |
| 95 % CI upper boundary | 2.96 | 3.05 | 4.38 | 5.35 | 5.55 | |
| 95 % CI lower boundary | 2.31 | 2.34 | 2.39 | 3.27 | 3.58 | |
| Ocular area (at rest, | ||||||
| Mean value | 2.63 | 2.63 | 2.70 | 3.19 | 2.78a | 3.43 |
| Standard deviation | 0.83 | 0.84 | 0.82 | 1.34 | 1.49 | 1.83 |
| 95 % CI upper boundary | 3.95 | 3.97 | 4.00 | 5.32 | 5.15 | 6.33 |
| 95 % CI lower boundary | 1.31 | 1.28 | 1.39 | 1.07 | 0.41 | 0.52 |
CI confidence interval
aValues differ significantly (P < .05)
bConstant truncated value (no particles ware detected)
Fig. 3This figure shows the degree of protection at the ocular area and on the instrument table, comparing the working state to the resting state
Fig. 4This figure shows the background concentration of particles both during the sham procedure and at rest. The smallest particles are present in larger amounts, compared to to the larger particles