Traceability Assessment and Performance Evaluation of Results for Measurement of Abbott Clinical Chemistry Assays on 4 Chemistry Analyzers.

Jinsook Lim, Kyung Eun Song, Sang Hoon Song, Hyun-Jung Choi, Sun Hoe Koo, Gye Choel Kwon1.   

Abstract

CONTEXT: -The traceability of clinical results to internationally recognized and accepted reference materials and reference measurement procedures has become increasingly important. Therefore, the establishment of traceability has become a mandatory requirement for all in vitro diagnostics devices.
OBJECTIVES: -To evaluate the traceability of the Abbott Architect c8000 system (Abbott Laboratories, Abbott Park, Illinois), consisting of calibrators and reagents, across 4 different chemistry analyzers, and to evaluate its general performance on the Toshiba 2000FR NEO (Toshiba Medical Systems Corporation, Otawara-shi, Tochigi-ken, Japan).
DESIGN: -For assessment of traceability, secondary reference materials were evaluated 5 times, and then bias was calculated. Precision, linearity, and carryover were determined according to the guidelines of the Clinical and Laboratory Standards Institute (Wayne, Pennsylvania).
RESULTS: -The biases from 4 different analyzers ranged from -2.33% to 2.70% on the Toshiba 2000FR NEO, -2.33% to 5.12% on the Roche Hitachi 7600 (Roche Diagnostics International, Basel, Switzerland), -0.93% to 2.87% on the Roche Modular, and -2.16% to 2.86% on the Abbott Architect c16000. The total coefficients of variance of all analytes were less than 5%. The coefficients of determination (R(2)) were more than 0.9900. The carryover rate ranged from -0.54% to 0.17%.
CONCLUSIONS: -Abbott clinical chemistry assays met the performance criteria based on desirable biological variation for precision, bias, and total error. They also showed excellent linearity and carryover. Therefore, these clinical chemistry assays were found to be accurate and reliable and are readily applicable on the various platforms used in this study.

Mesh:

Year:  2016        PMID: 27128304     DOI: 10.5858/arpa.2015-0088-OA

Source DB:  PubMed          Journal:  Arch Pathol Lab Med        ISSN: 0003-9985            Impact factor:   5.534


  1 in total

1.  Performance evaluation between two automated biochemical analyzer systems: Roche Cobas 8000 and Mindray BS2000M.

Authors:  Mingxing Chen; Simeng Qin; Sitao Yang; Huaping Chen; Liuyi Lu; Xue Qin
Journal:  J Med Biochem       Date:  2022-07-29       Impact factor: 2.157

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.