Literature DB >> 27124827

Hemorrhagic-Induced Cardiovascular Complications During Reamer-Irrigator-Aspirator-Assisted Femoral Nonunion Treatment.

Johanna C E Donders1, Holger M Baumann, Markus F Stevens, Peter Kloen.   

Abstract

UNLABELLED: The reamer-irrigator-aspirator (RIA) is a device that has recently been developed to retrieve bone graft from the medullary canal of the femur and the tibia. As for most new surgical procedures, complications arise as part of the learning curve and/or imperfections in the design or technique. To increase awareness of potential complications and to prevent unnecessary harm to the patient, new complications need to be reported. We present case reports of 2 patients who developed severe intraoperative cardiac events presumably caused by excessive and rapid blood loss during RIA bone graft harvesting. This complication has not been described before. Increased awareness of excessive blood loss and subsequent cardiac events during an RIA procedure is warranted. LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.

Entities:  

Mesh:

Year:  2016        PMID: 27124827     DOI: 10.1097/BOT.0000000000000612

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  2 in total

Review 1.  Complications associated using the reamer-irrigator -aspirator (RIA) system: a systematic review and meta-analysis.

Authors:  Markus Laubach; Lucas P Weimer; Felix M Bläsius; Frank Hildebrand; Philipp Kobbe; Dietmar W Hutmacher
Journal:  Arch Orthop Trauma Surg       Date:  2022-09-17       Impact factor: 2.928

2.  Challenges of bone tissue engineering in orthopaedic patients.

Authors:  Enrique Guerado; Enrique Caso
Journal:  World J Orthop       Date:  2017-02-18
  2 in total