Abstract
Uncertainties in source color measurements with a tristimulus colorimeter are estimated for calibration factors determined, based on a known source spectral distribution or on accurate measurements of the spectral responsivities of the colorimeter channels. Application is to the National Institute of Standards and Technology (NIST) colorimeter and an International Commission on Illumination (CIE) Illuminant A calibration. Detector-based calibration factors generally have lower uncertainties than source-based calibration factors. Uncertainties are also estimated for calculations of spectral mismatch factors. Where both spectral responsivities of the colorimeter channels and the spectral power distributions of the calibration and test sources are known, uncertainties are lowest if the colorimeter calibration factors are recalculated for the test source; this process also avoids correlations between the CIE Source A calibration factors and the spectral mismatch factors.Entities:
Keywords: radiometry colorimetry; tristimulus colorimeter; uncertainty
Year: 2007 PMID: 27110460 PMCID: PMC4656001 DOI: 10.6028/jres.112.010
Source DB: PubMed Journal: J Res Natl Inst Stand Technol ISSN: 1044-677X
Fig. 1Spectral responsivity of the NIST colorimeter channels compared to the CIE color-matching functions (dashed lines). (Courtesy G. Eppeldauer, NIST)
Uncertainties in (x,y) chromaticities and CCT for a true CIE Illuminant A source measured with the NIST colori-meter propagated from uncertainty components for responsivity measurements on the SIRCUS facility
| Responsivity Uncertainty Component | Component Value | |||
|---|---|---|---|---|
| Relative, random | 0.0006 | 0.00010 | 0.00013 | 0.8 |
| Offset, random (relative to | 6 · 10−6 | 0.00001 | 0.00001 | – |
| Offset, systematic (relative to | 2 · 10−5 | 0.00001 | 0.00001 | – |
| Total: | 0.00011 | 0.00013 | 0.8 |
Uncertainties in (x,y) chromaticities and CCT for a realized CIE Source A for typical spectral irradiance uncertainty components
| Uncertainty Component | Type | Component Value | |||
|---|---|---|---|---|---|
| Reference 2800 K blackbody | 2 K in temperature 0.9 correlation coefficient | 0.00020 | 0.00014 | 3.12 | |
| Reference signal offset (noise) | Random | 0.1 % of signal maximum | 0.00013 | 0.00015 | 1.74 |
| Reference signal offset (drift) | Systematic | 0.03 % of signal maximum | 0.00012 | 0.00008 | 1.25 |
| Test signal offset (noise) | Random | 0.1 % of signal maximum | 0.00010 | 0.00012 | 1.43 |
| Test signal offset (drift) | Systematic | 0.03 % of signal maximum | 0.00009 | 0.00006 | 0.98 |
| Source noise (relative) | Random | 0.1 % (in signal ratio) | 0.00003 | 0.00003 | 0.60 |
| Wavelength | Random | 0.3 nm | 0.00001 | 0.00000 | 0.09 |
| Wavelength offset | Systematic | 0.1 nm | 0.00000 | 0.00000 | 0.01 |
| Sum of all components | 0.00030 | 0.00026 | 4.2 |
Summary of uncertainty calculations: I = current component (random, 0.1 %); R = responsivity components (see Table 1); S = spectral irradiance components (see Table 2)
| Detector calibration | Source calibration | ||||||
|---|---|---|---|---|---|---|---|
| I | R | S | |||||
|
| |||||||
| 0.00040 | 0.00011 | 0.00030 | |||||
| 0.00053 | 0.00013 | 0.00026 | |||||
| (K) | 1.66 | 0.83 | 4.20 | ||||
|
| |||||||
| 5500 K blackbody source | |||||||
|
| |||||||
| Detector calibration | Source calibration (mismatch Source A uncorrelated) | Source | |||||
|
| |||||||
| Mismatch | |||||||
|
| |||||||
| I | R | S | Base | R | S | S | |
|
| |||||||
| 0.00017 | 0.00005 | 0.00020 | 0.00035 | 0.00002 | 0.00020 | 0.00035 | |
| 0.00018 | 0.00005 | 0.00026 | 0.00047 | 0.00004 | 0.00026 | 0.00047 | |
| (K) | 1.22 | 0.76 | 3.16 | 4.20 | 0.48 | 3.16 | 14.6 |