Feasibility, accuracy, and safety of 3-dimensional electroanatomic mapping without fluoroscopy in patients with congenital heart defects.

Apinya Bharmanee1, Srinath Gowda2, Harinder R Singh3.   

Abstract

BACKGROUND: Use of nonfluoroscopic 3-dimensional electroanatomic mapping (NF-3DEAM) in patients with congenital heart defects (CHDs) is not well reported.
OBJECTIVE: The purpose of this study was to evaluate the feasibility, accuracy, and safety of NF-3DEAM in patients with CHDs.
METHODS: Retrospective review of electrophysiologic study (EPS) in patients with CHD from 2009 to 2013 was conducted. Patients undergoing EPS with NF-3DEAM using the EnSite NavX system (St. Jude Medical) were included and those with fluoroscopy were excluded. Cardiac angiography performed in close proximity was matched with 3DEAM by proper aspect ratio using manual alignment by overlay images and comparing distances between cardiac landmarks using Pearson correlation coefficient and intraclass correlation. Success and complications were reviewed.
RESULTS: Three-dimensional electroanatomic mapping and angiograms were compared in 48 EPS (45 patients); 13 patients with intracardiac leads, quiescent chambers indicating significant scarring, and who required transseptal puncture were excluded. Indications for EPS included documented tachyarrhythmia and preoperative mapping. Mean age was 25.3 ± 9.8 years. Simple CHDs (isolated shunt lesions or valvular lesions) were identified in 13 studies (27%) and complex CHDs (others) were identified in 35 studies (73%). Ablations were performed in 25 studies (52%). Average time to obtain right atrial geometry was 25.3 minutes (range 14-47 minutes) and right ventricular geometry was 22.8 minutes (range 12-35 minutes). Pearson correlation coefficient and intraclass correlation of cardiac landmarks were 0.90 and 0.80, respectively. Anatomic landmarks, mapping, and ablation were accurate in all 3DEAMs. No complications were recorded.
CONCLUSION: NF-3DEAM is feasible, safe, and accurate in CHD patients without extensive scarring, intracardiac leads, and need for transseptal puncture.
Copyright © 2016 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Congenital heart defect; Feasibility; Safety; Three-dimensional electroanatomic mapping; accuracy

Mesh:

Year:  2016        PMID: 27108940     DOI: 10.1016/j.hrthm.2016.04.010

Source DB:  PubMed          Journal:  Heart Rhythm        ISSN: 1547-5271            Impact factor:   6.343


  2 in total

1.  Feasibility of a novel mapping system combined with remote magnetic navigation for catheter ablation of premature ventricular contractions.

Authors:  Shipeng Dang; Christian Jons; Peter Karl Jacobsen; Steen Pehrson; Xu Chen
Journal:  J Arrhythm       Date:  2019-01-31

Review 2.  ALARA in Pediatric Electrophysiology Laboratory.

Authors:  Harinder R Singh
Journal:  Children (Basel)       Date:  2022-06-10
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.