Literature DB >> 27103024

Comprehensive Quality Control of the ITG 68Ge/68Ga Generator and Synthesis of 68Ga-DOTATOC and 68Ga-PSMA-HBED-CC for Clinical Imaging.

Alejandro Amor-Coarasa1, Megan Schoendorf1, Marian Meckel2, Shankar Vallabhajosula1, John W Babich3.   

Abstract

UNLABELLED: A good-manufacturing-practices (GMP) (68)Ge/(68)Ga generator that uses modified dodecyl-3,4,5-trihydroxybenzoate hydrophobically bound to a octadecyl silica resin (C-18) as an adsorbent has been developed that allows for dilute HCl (0.05N) to efficiently elute metal-impurity-free (68)Ga(3+) ready for peptide labeling. We characterized the performance of this generator system over a year in conjunction with the production of (68)Ga-labeled DOTATOC and Glu-NH-CO-NH-Lys(Ahx)-HBED-CC (PSMA-HBED-CC) intended for clinical studies and established protocols for batch release.
METHODS: A 2,040-MBq self-shielded (68)Ge/(68)Ga generator provided metal-free (68)GaCl3 ready for peptide labeling in the fluidic labeling module after elution with 4 mL of 0.05N HCl. The compact system was readily housed in a laminar flow cabinet allowing an ISO class-5 environment. (68)Ga labeling of peptides using GMP kits was performed in 15-20 min, and the total production time was 45-50 min. Batch release quality control specifications were established to meet investigational new drug submission and institutional review board approval standards.
RESULTS: Over a period of 12 mo, (68)Ga elution yields from the generator averaged 80% (range, 72.0%-95.1%), and (68)Ge breakthrough was less than 0.006%, initially decreasing with time to 0.001% (expressed as percentage of (68)Ge activity present in the generator at the time of elution), a unique characteristic of this generator. The radiochemical purity of both (68)Ga-DOTATOC and (68)Ga-PSMA-HBED-CC determined by high-performance liquid chromatography analysis was greater than 98%, with a minimum specific activity of 12.6 and 42 GBq/μmol, respectively. The radionuclidic ((68)Ge) impurity was 0.00001% or less (under the detection limit). Final sterile, pyrogen-free formulation was provided in physiologic saline with 5%-7% ethanol.
CONCLUSION: The GMP-certified (68)Ge/(68)Ga generator system was studied for a year. The generator system is contained within the fluidic labeling module, and it is compact, self-shielded, and easy to operate using simple manual techniques. The system provides radiolabeled peptides with high (>98%) radiochemical purity and greater than 80% radiochemical yield. The (68)Ge levels in the final drug products were under the detection limits at all times. (68)Ga-DOTATOC and (68)Ga-PSMA-HBED-CC investigational radiopharmaceuticals are currently being studied clinically under investigational new drug (IND) applications submitted to the U.S. Food and Drug Administration.
© 2016 by the Society of Nuclear Medicine and Molecular Imaging, Inc.

Entities:  

Keywords:  68Ga generator; 68Ga-DOTATOC; 68Ga-PSMA-HBED-CC; quality control

Mesh:

Substances:

Year:  2016        PMID: 27103024     DOI: 10.2967/jnumed.115.171249

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  12 in total

1.  Analysis of Pros and Cons in Using [68Ga]Ga-PSMA-11 and [18F]PSMA-1007: Production, Costs, and PET/CT Applications in Patients with Prostate Cancer.

Authors:  Costantina Maisto; Michela Aurilio; Anna Morisco; Roberta de Marino; Monica Josefa Buonanno Recchimuzzo; Luciano Carideo; Laura D'Ambrosio; Francesca Di Gennaro; Aureliana Esposito; Paolo Gaballo; Valentina Pirozzi Palmese; Valentina Porfidia; Marco Raddi; Alfredo Rossi; Elisabetta Squame; Secondo Lastoria
Journal:  Molecules       Date:  2022-06-16       Impact factor: 4.927

2.  Evaluation of RECIST, PERCIST, EORTC, and MDA Criteria for Assessing Treatment Response with Ga68-PSMA PET-CT in Metastatic Prostate Cancer Patient with Biochemical Progression: a Comparative Study.

Authors:  Manoj Gupta; Partha Sarathi Choudhury; Sudhir Rawal; Harish Chandra Goel; S Avinash Rao
Journal:  Nucl Med Mol Imaging       Date:  2018-09-26

3.  Synthesis and pre-clinical evaluation of a new class of high-affinity 18F-labeled PSMA ligands for detection of prostate cancer by PET imaging.

Authors:  James Kelly; Alejandro Amor-Coarasa; Anastasia Nikolopoulou; Dohyun Kim; Clarence Williams; Shashikanth Ponnala; John W Babich
Journal:  Eur J Nucl Med Mol Imaging       Date:  2016-11-15       Impact factor: 9.236

4.  Initial risk stratification and staging in prostate cancer with prostatic-specific membrane antigen positron emission tomography/computed tomography: A first-stop-shop.

Authors:  Manoj Gupta; Partha Sarathi Choudhury; Sudhir Rawal; Harish Chandra Goel; Vineet Talwar; Amitabh Singh; Saroj Kumar Sahoo
Journal:  World J Nucl Med       Date:  2018 Oct-Dec

5.  Development of a 68Ge/68Ga Generator System Using Polysaccharide Polymers and Its Application in PET Imaging of Tropical Infectious Diseases.

Authors:  Takeshi Fuchigami; Hokuto Ono; Kohta Oyadomari; Mayumi Iwatake; Daisuke Hayasaka; Masoud Akbari; Katsuyuki Yui; Kodai Nishi; Takashi Kudo; Sakura Yoshida; Mamoru Haratake; Morio Nakayama
Journal:  ACS Omega       Date:  2017-04-11

6.  Evaluation of response in patients of metastatic castration resistant prostate cancer undergoing systemic radiotherapy with lutetium177-prostate-specific membrane antigen: A comparison between response evaluation criteria in solid tumors, positron-emission tomography response criteria in solid tumors, European organization for research and treatment of cancer, and MDA criteria assessed by gallium 68-prostate-specific membrane antigen positron-emission tomography-computed tomography.

Authors:  Manoj Gupta; Partha Sarathi Choudhury; Sudhir Rawal; Harish Chandra Goel; Shriram Avinash Rao
Journal:  Urol Ann       Date:  2019 Apr-Jun

7.  Otto: a 4.04 GBq (109 mCi) 68Ge/68Ga generator, first of its kind - extended quality control and performance evaluation in the clinical production of [68Ga]Ga-PSMA-11.

Authors:  Nicole N Waterhouse; Alejandro Amor-Coarasa; Anastasia Nikolopoulou; John W Babich
Journal:  EJNMMI Radiopharm Chem       Date:  2020-02-03

8.  68Ga-NOTA PET imaging for gastric emptying assessment in mice.

Authors:  Xueyan Chen; Yu Liu; Donghui Pan; Maoyu Cao; Xinyu Wang; Lizhen Wang; Yuping Xu; Yan Wang; Junjie Yan; Juan Liu; Min Yang
Journal:  BMC Gastroenterol       Date:  2021-02-13       Impact factor: 3.067

9.  Incremental value of 68-gallium-prostate-specific membrane antigen positron emission tomography/computed tomography in patients with abnormal prostate-specific antigen and benign transrectal ultrasound biopsy.

Authors:  Manoj Gupta; Partha Sarathi Choudhury; Sudhir Rawal; Gurudutt Gupta
Journal:  Urol Ann       Date:  2018 Apr-Jun

10.  3D-printed automation for optimized PET radiochemistry.

Authors:  Alejandro Amor-Coarasa; James M Kelly; John W Babich
Journal:  Sci Adv       Date:  2019-09-13       Impact factor: 14.136

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.