Assessment of an image-guided neurosurgery system using a head phantom.

P Ballesteros-Zebadúa1, O A García-Garduño1, O O Galván de la Cruz2, A Arellano-Reynoso3, J M Lárraga-Gutiérrez1, M A Celis2,3.   

Abstract

OBJECTIVE: To acknowledge the challenges and limitations of image-guided neurosurgery systems, we compared the application accuracy of two different image registration methods for one commercial system. (VectorVision, BrainLab, Germany).
METHODS: We used an anthropomorphic head phantom for radiosurgery and a custom built add-on to simulate surgical targets inside the brain during an image-guided neurosurgery. We used two image registration methods, fiducial registration using attachable surface markers for computed tomography (CT) and surface registration using infrared laser face scanning. After simulation, we calculated the three-dimensional (3D) distance between the predicted position of a target, and its actual position using a registered pointer and an infrared camera. Deviations were measured for both superficial fiducial markers and internal surgical targets by five different users.
RESULTS: Deviations from the location of fiducial markers after each registration method were 2.15 ± 0.93 mm after CT surface marker registration and 1.25 ± 0.64 mm after infrared face scanner registration. The mean target registration errors were 2.95 ± 1.4 mm using fiducial registration and 2.90 ± 1.3 mm using surface registration. The largest deviations (6.2 mm) were found for the targets in the skull base and posterior cranial fossa. Fiducial deviations and target registration errors were statistically uncorrelated. The total application accuracy was 4.87 ± 0.97 mm after CT surface marker registration and 4.14 ± 0.64 mm after infrared face scanner registration.
CONCLUSIONS: Despite others have reported differences, we did not find significant variations between both registration methods for the target registration error, although application accuracy was slightly better after surface face registration. Superficial registration errors, but not the target registration error, can be routinely evaluated in the operating room. Since both errors were uncorrelated, surgeons may neglect the achievable accuracy of the procedure. The described method is recommended to assess application accuracy in the operating room.

Keywords:  Accuracy; head phantom; image-guided neurosurgery; registration method; target registration error

Mesh:

Year:  2016        PMID: 27101082     DOI: 10.3109/02688697.2016.1173188

Source DB:  PubMed          Journal:  Br J Neurosurg        ISSN: 0268-8697            Impact factor:   1.596


  7 in total

1.  The Asterion-to-Transverse Process of the Atlas Line as a Surgical Landmark.

Authors:  Jaafar Basma; Dom E Mahoney; Christos Anagnostopoulos; L Madison Michael; Jeffrey M Sorenson; David G Porter; Angelo Pichierri
Journal:  J Neurol Surg B Skull Base       Date:  2021-07-05

2.  Virtual splint registration for electromagnetic and optical navigation in orbital and craniofacial surgery.

Authors:  Ruud Schreurs; F Baan; C Klop; L Dubois; L F M Beenen; P E M H Habets; A G Becking; T J J Maal
Journal:  Sci Rep       Date:  2021-05-17       Impact factor: 4.379

3.  Current accuracy of surface matching compared to adhesive markers in patient-to-image registration.

Authors:  Mireli A Mongen; Peter W A Willems
Journal:  Acta Neurochir (Wien)       Date:  2019-03-16       Impact factor: 2.216

4.  Infrared needle mapping to assist biopsy procedures and training.

Authors:  Bruce Shar; John Leis; John Coucher
Journal:  Healthc Technol Lett       Date:  2018-01-26

5.  Stereotactic Neuro-Navigation Phantom Designs: A Systematic Review.

Authors:  Marko Švaco; Ivan Stiperski; Domagoj Dlaka; Filip Šuligoj; Bojan Jerbić; Darko Chudy; Marina Raguž
Journal:  Front Neurorobot       Date:  2020-10-23       Impact factor: 2.650

6.  Image-guided cochlear access by non-invasive registration: a cadaveric feasibility study.

Authors:  Jiang Wang; Hongsheng Liu; Jia Ke; Lei Hu; Shaoxing Zhang; Biao Yang; Shilong Sun; Na Guo; Furong Ma
Journal:  Sci Rep       Date:  2020-10-27       Impact factor: 4.379

7.  Registration-free workflow for electromagnetic and optical navigation in orbital and craniofacial surgery.

Authors:  R Schreurs; F Baan; C Klop; L Dubois; L F M Beenen; P E M H Habets; A G Becking; T J J Maal
Journal:  Sci Rep       Date:  2021-09-10       Impact factor: 4.379

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.