Early Complications and Secondary Procedures in Transfibular Total Ankle Replacement.

Eric W Tan1, Camilla Maccario2, Paul G Talusan3, Lew C Schon4.   

Abstract

BACKGROUND: A new transfibular total ankle arthroplasty (TAA) system has not been assessed for potential early complications.
METHODS: We retrospectively assessed prospectively collected data on the initial cohort of patients undergoing TAA with this implant. We evaluated visual analog scale (VAS) pain and function, range of motion, and early radiographic outcomes.
RESULTS: Twenty consecutive TAAs (19 patients) were treated with the implant from January 2013 through June 2014. Average patient age was 63.7 (range, 41-80) years, with an average follow-up of 18 (range, 12-27) months. No fibular nonunion or implant failure was found at 12 months postoperatively. One patient had asymptomatic mild tibial lucency. Four of 20 TAAs underwent additional surgery for anterior impingement (1 ankle), deep infection and symptomatic fibular hardware (1 ankle), and symptomatic fibular hardware (2 ankles).
CONCLUSION: Of 20 ankles treated with a new transfibular arthroplasty system, no fibular nonunion, delayed union, or implant failure was noted at 12 months postoperatively. Two complications were resolved with secondary treatment, and 2 other ankles underwent secondary surgery for symptomatic fibular hardware with good outcome. The findings suggest that this total ankle system is safe and effective at short-term follow-up. LEVEL OF EVIDENCE: Level IV, retrospective case series.
© The Author(s) 2016.

Entities:  

Keywords:  ankle arthritis; ankle replacement; arthroplasty; fibula osteotomy; fixed bearing; lateral approach; trabecular metal

Mesh:

Year:  2016        PMID: 27098127     DOI: 10.1177/1071100716644817

Source DB:  PubMed          Journal:  Foot Ankle Int        ISSN: 1071-1007            Impact factor:   2.827


  6 in total

1.  Treatment of Ankle Osteoarthritis with Total Ankle Replacement Through a Lateral Transfibular Approach.

Authors:  Federico G Usuelli; Riccardo D'Ambrosi; Luigi Manzi; Camilla Maccario; Cristian Indino
Journal:  J Vis Exp       Date:  2018-01-24       Impact factor: 1.355

2.  Metal artifact reduction MRI of total ankle arthroplasty implants.

Authors:  Cesar de Cesar Netto; Lucas F Fonseca; Benjamin Fritz; Steven E Stern; Esther Raithel; Mathias Nittka; Lew C Schon; Jan Fritz
Journal:  Eur Radiol       Date:  2017-12-07       Impact factor: 5.315

3.  Total ankle replacement through a lateral approach: surgical tips.

Authors:  Federico Giuseppe Usuelli; Cristian Indino; Camilla Maccario; Luigi Manzi; Vincenzo Salini
Journal:  SICOT J       Date:  2016-11-18

4.  Contact stresses, pressure and area in a fixed-bearing total ankle replacement: a finite element analysis.

Authors:  Nicolo Martinelli; Silvia Baretta; Jenny Pagano; Alberto Bianchi; Tomaso Villa; Gloria Casaroli; Fabio Galbusera
Journal:  BMC Musculoskelet Disord       Date:  2017-11-25       Impact factor: 2.362

Review 5.  The Clinical Application of Porous Tantalum and Its New Development for Bone Tissue Engineering.

Authors:  Gan Huang; Shu-Ting Pan; Jia-Xuan Qiu
Journal:  Materials (Basel)       Date:  2021-05-18       Impact factor: 3.623

6.  The Infinity Total Ankle System: Early Clinical Results With 2- to 4-Year Follow-up.

Authors:  Murray Penner; W Hodges Davis; Kevin Wing; Thomas Bemenderfer; Feras Waly; Robert B Anderson
Journal:  Foot Ankle Spec       Date:  2018-06-04
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.