Usefulness of filtered unipolar electrogram morphology for evaluating transmurality of ablated lesions during pulmonary vein isolation.

Rikitake Kogawa1, Ichiro Watanabe1, Yasuo Okumura1, Hiroaki Mano1, Kazumasa Sonoda1, Koichi Nagashima1, Naoko Sasaki1, Kimie Ohkubo1, Keiko Takahashi1, Kazuki Iso1, Sayaka Kurokawa1, Toshiko Nakai1, Atsushi Hirayama1.   

Abstract

BACKGROUND: Although alteration of the amplitude and morphology of bipolar electrograms is used widely as a guide of the ablation effect, there is little information concerning unipolar electrograms. The amplitude and morphology of filtered bipolar (BP) and filtered unipolar (UP) electrograms were compared during pulmonary vein isolation in patients with atrial fibrillation.
METHODS: BP (30-250 Hz) and UP (30-100 Hz) signals from the ablation catheter were recorded before and after each ablation point at the pulmonary vein antrum in 6 patients with atrial fibrillation.
RESULTS: In the electrogram group with low-voltage amplitude in BP electrograms before ablation (<0.5 mV), the reduction in amplitude after ablation was significantly greater in the UP than in the BP electrograms, whereas the reduction was similar between the two recording methods in the electrogram group with high-voltage amplitude in BP electrograms (≥0.5 mV). Furthermore, the S wave in the UP electrograms disappeared at the sites of no pace capture after ablation, whereas no characteristic morphologic changes were observed in the BP electrograms.
CONCLUSION: Filtered UP electrograms may be useful in assessing the effectiveness of lesion formation.

Entities:  

Keywords:  Atrial fibrillation; Durable pulmonary vein isolation; Filtered unipolar electrogram; Pulmonary vein isolation

Year:  2015        PMID: 27092191      PMCID: PMC4823572          DOI: 10.1016/j.joa.2015.10.001

Source DB:  PubMed          Journal:  J Arrhythm        ISSN: 1880-4276


Introduction

Extensive encircling ipsilateral pulmonary vein (PV) isolation (EEPVI) has been validated as a curative therapy for paroxysmal and persistent atrial fibrillation (AF) [1], [2], [3]. Complete conduction block requires a contiguous line of ablation lesions [4]. Although previous studies have shown that power titration guided by amplitude reduction in the local atrial electrogram succeeds in avoiding complications [5], [6], [7], whether the amplitude reduction observed in local bipolar (BP) and unipolar (UP) electrograms recorded after an ablation can reliably predict transmural lesion formation is controversial [5], [6], [7], [8]. The effect of different catheter orientations relative to the endocardial surface, the direction of the wavefront on the BP electrogram recorded from transmural and non-transmural lesions, and a comparison of the UP and BP electrograms for the identification of transmural lesion formation at different catheter orientations were reported in an animal model [9]. However, it is difficult to monitor unfiltered UP electrograms during the application of radiofrequency (RF) energy because of electrogram drift. Therefore, we compared filtered UP electrograms and filtered BP electrograms for their usefulness in guiding the creation of transmural atrial lesions during the application of RF energy in EEPVI.

Material and methods

Patients

Six patients with drug-refractory paroxysmal AF comprised the study group. The Institutional Review Board of Nihon University Itabashi Hospital approved this study on April 15, 2012 (IRB number: 8). Written informed consent was obtained from each participant.

Mapping and RF catheter ablation

BARD LabSystemTM PRO (Murray Hill, NJ, USA) was used for intracardiac electrogram recordings and NavXTM Velocity System (St. Jude Medical, St. Paul, MN, USA) was used for creating a three-dimensional geometric map. Left and right EEPVIs were performed antrally for 30 s at each site on the atrial side around the circular mapping catheters until the occurrence of entrance block using a 3.5 mm irrigated tip catheter (ThermoCool; Biosense Webster, Inc., Diamond Bar, CA, USA), which is defined as the loss of all ipsilateral PV potentials, as well as the loss of dormant PV conduction by the administration of 30 mg of ATP. After each ablation, UP pacing (output, 10 mA; pulse width, 2 ms) was performed from the ablation catheter. Since a previous animal study demonstrated that the pacing thresholds increased significantly more at sites with transmural necrosis (unipolar: increase by 378±103% vs. 207±93%; non-transmural, P<0.001) [10], we arbitrarily defined loss of unipolar pacing capture of 10 mA, 2 ms as transmural lesion formation because pacing threshold before ablation is usually <2 mA/2 ms; ablation was repeated until the absence of left atrial (LA) capture was achieved [11].

Post-ablation electrogram characteristics

The filter settings were 30–250 Hz for the BP electrograms and 30–100 Hz for the UP electrograms. The proximal pole of a pentapolar catheter placed in the aorta was used as the indifferent cathodal reference for the UP electrode. BP and UP electrograms that were recorded from a total of 140 ablated points from 6 patients were assessed before and after each ablation, as well as after the absence of LA capture was achieved (Fig. 1A, B). The filtered BP and UP electrograms were analyzed for peak-to-peak amplitude, which is defined as the voltage difference between the highest and lowest deflections, as well as the qualitative analysis of the development of an R or rr′ morphology (change from RSr′ or rSr′ to R or rr′) visually in the UP electrograms.
Fig. 1

Three-dimensional mapping images of the left atrium showing representative recording sites (indicated by red dots) of bipolar and filtered unipolar electrograms. The recordings were made at 4 points on the anterior (A) and 4 points on the posterior (B) pulmonary vein antra. AP: anteroposterior; PA: posteroanterior.

Statistical analysis

Continuous variables are presented as mean±standard deviation, and the Wilcoxon signed-rank test was used to analyze differences in these variables. P<0.05 was considered statistically significant.

Results

Effect of ablation on electrogram voltage

We analyzed 5–6 electrograms per pulmonary vein, and a total of 140 electrograms were analyzed. When baseline pre-ablation BP and UP electrogram amplitudes were compared with the post-ablation electrogram amplitudes in the absence of LA capture, the voltages of the BP and UP electrograms decreased from 0.51±0.33 mV to 0.28±0.19 mV (P=0.002) and from 0.54±0.33 mV to 0.31±0.19 mV (P<0.01), respectively. We then classified the pre-ablation BP electrogram voltages into high (≥0.5 mV, n=75 electrograms) and low BP voltage groups (<0.5 mV, n=65 electrograms). In the high BP voltage group, the post-ablation reduction in voltage was similar between the BP and UP electrograms (0.59±1.2 mV vs. 0.39±0.27 mV, P=0.2565). In the low BP voltage group, however, the post-ablation reduction in voltage was significantly greater in the UP compared to the BP electrograms (0.15±0.12 mV vs. 0.10±0.12 mV, P=0.0063).

Effect of ablation on UP electrogram morphology

Local filtered UP electrograms before ablation showed an rSr′ or RSr′ pattern. At the site where the LA capture remained present, filtered UP electrograms showed an rsr′ or Rsr′ pattern. However, where LA capture was absent, UP electrograms showed an R or rr′ pattern (Fig. 2A, B and Table 1). In contrast, no characteristic morphologic changes were observed in BP electrograms.
Fig. 2

Bipolar and filtered unipolar electrograms obtained before and after ablation. Panels 1–8 refer to the recording sites shown in Fig. 1. (A) Bipolar and filtered unipolar electrogram amplitudes were measured from the peak positive point to the peak negative point. (B) TL: transmural lesion formation assessed by non-capture with 10-mA, 2-ms unipolar pacing after ablation.

Table 1

Comparison of the changes in the filtered unipolar electrograms before and after ablation.

Patient #Non-transmural pattern
Transmural pattern
Pre-ablationPost-ablationPre-ablationPost-ablation
1rSr′: 3rsr′: 2rSr′:10R: 8
Rsr′: 1rr′:2
RSr′: 2rsr′: 1RSr′: 8R: 5
Rsr′: 1rr′:3











2rSr′: 4rsr′: 2rSr′: 9R: 6
Rsr′: 2rr′:3
RSr′:2rsr′: 1RSr′: 9R: 5
Rsr′: 1rr′: 4











3rSr′: 3rsr′: 2rSr′: 10R:7
Rsr′: 1rr′: 3
RSr′:3rsr′: 3RSr′: 10R: 6
Rsr′: 0rr′: 4











4rSr′: 1rsr′: 1rSr′: 8R: 4
Rsr′: 0rr′: 4
RSr′: 2rsr′: 0RSr′: 9R: 6
Rsr′: 2rr′: 3











5rSr′: 4rsr′: 3rSr′: 7R: 5
Rsr′: 1rr′: 2
RSr′: 2rsr′: 0RSr′: 8R: 6
Rsr′: 2rr′: 2











6rSr′: 3rsr′: 2rSr′: 10R: 6
Rsr′: 1rr′: 4
RSr′: 4rsr′: 1RSr′: 9R: 6
Rsr′: 3rr′: 3

Discussion

Main findings

Filtered BP and UP LA electrogram voltages recorded from the ablation catheter decreased significantly after ablation at the sites of no pace capture, and only the UP electrograms showed a significant reduction in voltage after ablation at the sites where the pre-ablation BP voltage was <0.5 mV. Furthermore, filtered UP electrogram morphology showed an rsr′ or Rsr′ pattern at the pace-capture sites, which changed to an R or rr′ pattern in the absence of LA capture. Otomo et al. [9] showed that, in porcine atria, transmural atrial ablation lesions were associated with loss of the terminal S wave in UP electrograms. Michowitz et al. [12] reported that in UP electrograms recorded in humans from a circumferential mapping catheter after achievement of entrance block, a UP terminal S wave of >0.2 mV was associated with post-ablation capture with an overall accuracy of 58%.

Clinical implications

Otomo et al. [9] used filter settings for UP electrogram recordings of 0.5 Hz and 100 Hz, and Michowitz et al. [12] used settings of 0.5–150 Hz or 0.5–180 Hz. However, we used filter settings of 30–100 Hz. If a filter setting was not used, it was impossible to record UP electrograms during ablation because of the large amount of electrical noise and electrogram drift, but it was possible to record UP electrograms during RF energy delivery with the 30–100 Hz filter setting. Although measurement of electrogram voltages before and after each ablation is time consuming, it allows the evaluation of lesion formation in real time during ablation. Since this is performed by observation of the changes in filtered UP electrogram morphology, insufficient ablation and unnecessarily high-power and long-duration ablations can be avoided. The absence of the characteristic changes in the filtered bipolar electrogram morphology on ablation may be due to several reasons. Since the filtered bipolar electrogram showed multiple spiky deflections, it was difficult to characterize morphological changes before and after ablation. Additionally, because bipolar electrogram morphology is influenced by wavefront orientation, its alteration with respect to the ablated lesion makes the morphology comparison more difficult, especially at the preablation low voltage sites.

Study limitations

We did not investigate the rate of AF recurrence after ablation using this method, nor did we compare the sites of dormant conduction unveiled by bolus administration of ATP with the sites of presence and absence of LA capture. Furthermore, transmural lesion formation was speculated from the absence of UP capture with high-output pacing. Finally, we did not include the orientation of the tip of the ablation catheter to the results. Further studies are needed to validate our findings.

Conclusions

The combined use of changes in real-time filtered UP electrogram amplitude and morphology to guide PVI by RF ablation is feasible and may aid in assessing the effectiveness of lesion formation during EEPVI.

Disclosures

None.

Conflict of interest

All authors declare no conflict of interest related to this study.

Financial support

Departmental resources supported this study.
  12 in total

1.  Atrial linear ablations in pigs. Chronic effects on atrial electrophysiology and pathology.

Authors:  L Gepstein; G Hayam; S Shpun; D Cohen; S A Ben-Haim
Journal:  Circulation       Date:  1999-07-27       Impact factor: 29.690

2.  Electrogram-guided radiofrequency catheter ablation of atrial tissue comparison with thermometry-guide ablation: comparison with thermometry-guide ablation.

Authors:  D Schwartzman; J J Michele; C T Trankiem; J F Ren
Journal:  J Interv Card Electrophysiol       Date:  2001-09       Impact factor: 1.900

3.  Identification of transmural necrosis along a linear catheter ablation lesion during atrial fibrillation and sinus rhythm.

Authors:  Javier E Sanchez; G Neal Kay; Michael E Benser; Jeffrey A Hall; Gregory P Walcott; William M Smith; Raymond E Ideker
Journal:  J Interv Card Electrophysiol       Date:  2003-02       Impact factor: 1.900

4.  Simultaneous creation and evaluation of linear radiofrequency lesions.

Authors:  Hennie van Rensburg; Rik Willems; Patricia Holemans; Wim Anné; Hugo Ector; Hein Heidbüchel
Journal:  J Interv Card Electrophysiol       Date:  2002-07       Impact factor: 1.900

5.  Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surface.

Authors:  Kiyoshi Otomo; Kikuya Uno; Hideomi Fujiwara; Mitsuaki Isobe; Yoshito Iesaka
Journal:  Heart Rhythm       Date:  2010-06-09       Impact factor: 6.343

6.  Techniques, evaluation, and consequences of linear block at the left atrial roof in paroxysmal atrial fibrillation: a prospective randomized study.

Authors:  Mélèze Hocini; Pierre Jaïs; Prashanthan Sanders; Yoshihide Takahashi; Martin Rotter; Thomas Rostock; Li-Fern Hsu; Frédéric Sacher; Sylvain Reuter; Jacques Clémenty; Michel Haïssaguerre
Journal:  Circulation       Date:  2005-12-13       Impact factor: 29.690

7.  Monitoring the local electrogram at the ablation site during radiofrequency application for common atrial flutter.

Authors:  K Azegami; S Satake; K Okishige; T Sasano; H Ohira; K Yamashita
Journal:  Jpn Circ J       Date:  1998-08

8.  Complete isolation of left atrium surrounding the pulmonary veins: new insights from the double-Lasso technique in paroxysmal atrial fibrillation.

Authors:  Feifan Ouyang; Dietmar Bänsch; Sabine Ernst; Anselm Schaumann; Hitoshi Hachiya; Minglong Chen; Julian Chun; Peter Falk; Afsaneh Khanedani; Matthias Antz; Karl-Heinz Kuck
Journal:  Circulation       Date:  2004-10-04       Impact factor: 29.690

9.  Electrical connections between pulmonary veins: implication for ostial ablation of pulmonary veins in patients with paroxysmal atrial fibrillation.

Authors:  Atsushi Takahashi; Yoshito Iesaka; Yoshihide Takahashi; Ryoko Takahashi; Kenzaburo Kobayashi; Katsumasa Takagi; Osamu Kuboyama; Takeo Nishimori; Hidenobu Takei; Hiroshi Amemiya; Hideomi Fujiwara; Masayasu Hiraoka
Journal:  Circulation       Date:  2002-06-25       Impact factor: 29.690

10.  Loss of pace capture after radiofrequency application predicts the formation of uniform transmural lesions.

Authors:  Ioanna Kosmidou; Haley Houde-Walter; Lori Foley; Gregory Michaud
Journal:  Europace       Date:  2012-11-21       Impact factor: 5.214

View more
  2 in total

1.  Unipolar electrogram-guided radiofrequency catheter ablation in paroxysmal atrial fibrillation: electrogram patterns and outcomes.

Authors:  Guohua Fu; Bin He; Binhao Wang; Jing Liu; Yibo Yu; Xianfeng Du; Mingjun Feng; Fang Gao; He Jin; Renyuan Fang; Huimin Chu
Journal:  J Interv Card Electrophysiol       Date:  2018-10-17       Impact factor: 1.900

2.  Continuous and discontinuous radiofrequency energy delivery on the atrial free wall: Lesion transmurality, width, and biophysical characteristics.

Authors:  Daniel J Friedman; Jed A Overmann; Jeffrey M Fish; Stephen A Gaeta; John H Tranter; Riki Thao; Jonathan P Piccini
Journal:  Heart Rhythm O2       Date:  2021-11-06