Literature DB >> 27091353

In Vitro Comparison of Cyclic Fatigue Resistance of ProTaper Next, HyFlex CM, OneShape, and ProTaper Universal Instruments in a Canal with a Double Curvature.

Hüseyin Sinan Topçuoğlu1, Gamze Topçuoğlu2, Ahmet Akti3, Salih Düzgün3.   

Abstract

INTRODUCTION: The aim of this study was to compare the resistance to cyclic fatigue of ProTaper Next X2 (PTN X2; size 25, 0.06 taper), Hyflex CM (HCM; size 25, 0.06 taper), OneShape (OS; size 25, 0.06 taper), and ProTaper Universal F2 (PTU F2; size 25, 0.08 taper) nickel-titanium files in an artificial root canal with a double (S-shaped) curvature.
METHODS: A total of 160 new PTN X2, OS, HCM, and PTU F2 files were tested in an artificial stainless steel canal with a double curvature. Forty files from each system were rotated until fracture to calculate the number of cycles to failure. The length of each fractured fragment was recorded. Data were analyzed by using one-way analysis of variance and Tukey post hoc tests.
RESULTS: The resistance to cyclic fatigue of the PTN X2 and HCM instruments was significantly greater than the OS and PTU F2 instruments in the apical curvature (P < .05). There was no statistical difference in the cyclic fatigue resistance of the PTN X2 and HCM instruments in the apical curvature (P > .05). In addition, there was no statistical difference between the OS and PTU F2 instruments in the apical curvature (P > .05). PTN X2, OS, HCM, and PTU F2 instruments showed similar cyclic fatigue resistance values in the coronal curvature (P > .05).
CONCLUSIONS: This study showed that PTN X2 and HCM instruments exhibit greater resistance to cyclic fatigue than OS and PTU F2 instruments in the apical curvature of an artificial canal with a double curvature.
Copyright © 2016 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cyclic fatigue; Hyflex CM; NiTi files; ProTaper Next; S-shaped canal

Mesh:

Substances:

Year:  2016        PMID: 27091353     DOI: 10.1016/j.joen.2016.03.010

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  16 in total

1.  Cyclic fatigue resistance of XP-endo Shaper compared with different nickel-titanium alloy instruments.

Authors:  Amr Elnaghy; Shaymaa Elsaka
Journal:  Clin Oral Investig       Date:  2017-10-13       Impact factor: 3.573

2.  Comparison of the cyclic fatigue resistance of One Curve, F6 Skytaper, Protaper Next, and Hyflex CM endodontic files.

Authors:  Charlotte Gouédard; Laurent Pino; Reza Arbab-Chirani; Shabnam Arbab-Chirani; Valérie Chevalier
Journal:  Restor Dent Endod       Date:  2022-03-04

3.  Bending resistance and cyclic fatigue life of Reciproc Blue, WaveOne Gold, and Genius files in a double (S-shaped) curved canal.

Authors:  Taha Özyürek; Mustafa Gündoğar; Koray Yılmaz; Gülşah Uslu
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-12-13

4.  In vitro comparison of the cyclic fatigue resistance of HyFlex EDM, One G, and ProGlider nickel titanium glide path instruments in single and double curvature canals.

Authors:  Koray Yılmaz; Gülşah Uslu; Taha Özyürek
Journal:  Restor Dent Endod       Date:  2017-10-30

5.  Comparison of the Shaping Ability of WaveOne Reciprocating Files with or without Glide Path in Simulated Curved S-shaped Root Canals.

Authors:  Ayca Yilmaz; Enver Sedat Kucukay; Murat Istektepe; Selen Nihal Sisli; Handan Ersev; Isil Karagoz-Kucukay
Journal:  J Int Soc Prev Community Dent       Date:  2017-06-20

6.  CBCT Assessment of Root Dentine Removal by Gates-Glidden Drills and Two Engine-Driven Root Preparation Systems.

Authors:  Azade Harandi; Fatemeh Mohammadpour Maleki; Ehsan Moudi; Maryam Ehsani; Soraya Khafri
Journal:  Iran Endod J       Date:  2017

Review 7.  Rotary science and its impact on instrument separation: A focused review.

Authors:  Sandhya Anand Khasnis; Prem Prakash Kar; Apoorva Kamal; Jayaprakash D Patil
Journal:  J Conserv Dent       Date:  2018 Mar-Apr

8.  Cyclic fatigue, bending resistance, and surface roughness of ProTaper Gold and EdgeEvolve files in canals with single- and double-curvature.

Authors:  Wafaa A Khalil; Zuhair S Natto
Journal:  Restor Dent Endod       Date:  2019-04-26

9.  Cyclic Fatigue Resistance of Heat-Treated Nickel-Titanium Instruments.

Authors:  Mário Tanomaru-Filho; Camila Galletti Espir; Ana Carolina Venção; Nathaly Macedo-Serrano; Jader Camilo-Pinto; JulianeMaria Guerreiro-Tanomaru
Journal:  Iran Endod J       Date:  2018

10.  Effect of Sterilization on Cyclic Fatigue Resistance of Proflexendo Endodontic Rotary Files.

Authors:  Emad O Alshwaimi
Journal:  Saudi J Med Med Sci       Date:  2019-02-08