Menahem Y Rotenberg1, Hovav Gabay2, Yoram Etzion2,3, Smadar Cohen1,3,4. 1. The Avram and Stella Goldstein-Goren Department of Biotechnology Engineering, Ben-Gurion University of the Negev, Beer-Sheva, Israel. 2. Cardiac Arrhythmia Research Laboratory, Department of Physiology and Cell Biology, Ben-Gurion University of the Negev, Beer-Sheva, Israel. 3. Regenerative Medicine &Stem Cell Research Center, Ben-Gurion University of the Negev, Beer-Sheva, Israel. 4. Ilse Katz Institute for Nanoscale Science and Technology, Ben-Gurion University of the Negev, Beer-Sheva, Israel.
Abstract
A noninvasive, effective approach for immediate and painless heart pacing would have invaluable implications in several clinical scenarios. Here we present a novel strategy that utilizes the well-known mechano-electric feedback of the heart to evoke cardiac pacing, while relying on magnetic microparticles as leadless mechanical stimulators. We demonstrate that after localizing intravenously-injected magnetic microparticles in the right ventricular cavity using an external electromagnet, the application of magnetic pulses generates mechanical stimulation that provokes ventricular overdrive pacing in the rat heart. This temporary pacing consistently managed to revert drug-induced bradycardia, but could only last up to several seconds in the rat model, most likely due to escape of the particles between the applied pulses using our current experimental setting. In a pig model with open chest, MEF-based pacing was induced by banging magnetic particles and has lasted for a longer time. Due to overheating of the electromagnet, we intentionally terminated the experiments after 2 min. Our results demonstrate for the first time the feasibility of external leadless temporary pacing, using injectable magnetic microparticles that are manipulated by an external electromagnet. This new approach can have important utilities in clinical settings in which immediate and painless control of cardiac rhythm is required.
A noninvasive, effective approach for immediate and painless heart pacing would have invaluable implications in several clinical scenarios. Here we present a novel strategy that utilizes the well-known mechano-electric feedback of the heart to evoke cardiac pacing, while relying on magnetic microparticles as leadless mechanical stimulators. We demonstrate that after localizing intravenously-injected magnetic microparticles in the right ventricular cavity using an external electromagnet, the application of magnetic pulses generates mechanical stimulation that provokes ventricular overdrive pacing in the rat heart. This temporary pacing consistently managed to revert drug-induced bradycardia, but could only last up to several seconds in the rat model, most likely due to escape of the particles between the applied pulses using our current experimental setting. In a pig model with open chest, MEF-based pacing was induced by banging magnetic particles and has lasted for a longer time. Due to overheating of the electromagnet, we intentionally terminated the experiments after 2 min. Our results demonstrate for the first time the feasibility of external leadless temporary pacing, using injectable magnetic microparticles that are manipulated by an external electromagnet. This new approach can have important utilities in clinical settings in which immediate and painless control of cardiac rhythm is required.
Pacing modalities for electrical stimulation of the heart are fundamental tools in modern
cardiac electrophysiology and are used for various clinical purposes12.
Technological advances during the last three decades have led to tremendous improvements
of implanted pacing devices in terms of their sensing, pacing and computational
capabilities as well as battery size and lifetime 2. In addition, rapid
progress has been made in leadless technologies that in some cases can circumvent the
need for intravascular leads, which are considered to be the Achilles’ heel of
these devices345. Nevertheless, the typical procedure for either
temporary or permanent pacemaker implantation is invasive, takes time and is preferably
done under fluoroscopy by skilled personnel. Thus, this form of therapy encompasses
risks of bleeding, infection6789 and thrombosis810
and importantly, is not readily available in cases requiring acute and rapid application
of pacing, which can be life-saving in patients suffering from bradycardia and
hemodynamic compromise11. Thus, for emergency bradycardia cases in which
pacing should be started immediately, transcutaneous electrical pacing through the chest
wall (external pacing) is the only viable approach. However, this technique is very
painful and in most cases necessitates the use of sedative or anesthetic agents, which
may further impair the critical hemodynamic condition of the patient11.
An approach for fast, leadless and painless external pacing may be invaluable not only
in bradycardia cases, but could also be used to halt common forms of
tachyarrhythmias12. Therefore, significant efforts have been invested
at developing a totally noninvasive approach of cardiac pacing utilizing magnetic field
stimulation13. However, this approach has thus far failed to reach
practical use due to very high energy requirements and a very low yield of pacing1415.Mechano-electrical feedback (MEF) is a well-known mechanism which can be used to pace
cardiac tissue. According to this mechanism, the myocardium is electrically excited in
response to a mechanical stimulation resulting from contact, force or pressure. For
example, ectopic beats may be induced as a result of a catheter-tip approaching the
endocardium during cardiac catheterization16. Precordial thumps were
reported to successfully revert ventricular tachycardia1718 and even
early ventricular fibrillation19. Recently, high intensity focused
ultrasound was shown to induce a mechanical force that provoked single ectopic beats
noninvasively20. It should be noted that local pressure as low as
2kPa may provoke MEF-dependent pacing21, while tissue damage usually
starts when the impact energy exceeds ~250kPa22. This gap of over
two orders of magnitude gives a large therapeutic window for safe MEF-induced cardiac
pacing.We hypothesized that MEF-induced cardiac pacing could be achieved by the action of
localized magnetically responsive microparticles on the ventricle wall (Fig. 1a). We envisioned that intravenously injected microparticles can be
localized and trapped in the cavity of the right ventricle (RV) by applying an external
magnetic force. Subsequently, effective pacing could be achieved by generating an
alternating magnetic field that periodically forces the microparticles against the
ventricular wall. In this paper, we used ex-vivo and in-vivo rat models,
followed by proof-of-concept experiments in a pig model, in order to demonstrate the
feasibility of our novel approach.
Figure 1
The proposed modality for inducing cardiac pacing and features of its
components: the magnetic microparticles (IMPs) and electromagnet.
An illustration of our proposed modality, and its setting (a, Figure
courtesy of Yair Vardi). The IMP morphology is viewed by SEM (b) and
their size distribution analyzed by static light scattering (c) A
color map of the magnetic induction (d) and the magnetic induction
intensity and gradient as a function of the distance from the electromagnet
tip (e) generated by the electromagnet, as obtained from the
computerized model.
Results
Features of the magnetic iron microparticles (IMPs)
Commercially-available iron microparticles (IMPs) were chosen as the magnetic
stimulators in our feasibility experiments due to their high magnetic moment
(224 A.m2/kg), which facilitates their magnetic
localization, and their low coercivity of 70 A/m23, which
prevents the particles from aggregating in the absence of a magnetic field.
According to scanning electron microscope (SEM) images, and static light
scattering (SLS) the IMPs are smooth spheres, with a mode diameter size of
7.6 μm (Fig. 1b,c). The magnetic
attraction applied by a given magnet on a magnetic microparticle is proportional
to its mass, thus to the third power of its radius, while the drag force applied
by the blood flow is proportional to the radius (supplementary information
Fig. 1). For this reason, larger particles have a stronger
tendency to become trapped under the blood flow using an external magnet. Thus,
7.6 μm is an appropriate size for such a feasibility study,
considering their magnetic properties. Moreover, although the size of the
particles may not allow them to pass in the capillaries and transport in the
blood circulation freely, they are rather close in size to particles that can do
so (<3 μm)2425.
Electromagnet Design
The electromagnet is a critical player as it has to generate a strong magnetic
induction as well as gradient in order to trap the microparticles in the cavity
of the RV. Moreover, it must allow complete control over its magnetization so it
can generate magnetic pulses that will push the particles and provoke
MEF-induced pacing. For this purpose, we designed an electromagnet composed of a
coil and permendur core (saturation induction of ~2.34 T). Due
to the low magnetic coercivity of permendur (80 A/m), the magnetic
induction generated by the core is eliminated as the electric current through
the coil is stopped. This feature allows for the precise control of the
electromagnet induction by altering the electric current through the coil. We
used COMSOL Multiphysics software to simulate the magnetic properties of the
electromagnet, and the computerized model is illustrated in Fig.
1d,e. In Supplementary Fig.
2 the model was expanded to include currents of 15 and 20 A,
which cover the full range of currents used throughout our study; it was
important to perform such an analysis as the magnetic permeability of the
electromagnet core, permendur, is not constant, leading to a magnetic induction
which is not linear to the current applied through the coil.In our setting, we need to apply a strong magnetic induction as well as gradient
at a distance of about 0.5 cm from the electromagnet tip. The magnetic
field lines, however, tend to disperse, which leads to a rapid decrease in the
magnetic induction. On the other hand, the permeability of the core is up to
7000 times higher than that of the air, so the field lines can energetically
benefit from going through the core rather than the air. Thus, beveling the core
“focuses” the induction near the tip of the core2627, increasing the field and the gradient it generates. For
this reason, the permendur core in our electromagnet was machined so that the
tip was narrowed to a diameter of 5 mm, as illustrated in Fig. 1d.Before proceeding to the animal study, we performed an evaluation of the
mechanical effect we may obtain using our setting. According to our
calculations, by using a current of 20 A, we could apply mean local
pressure of up to ~146, 41 or 11 kPa at distances of 0.5, 1 or
2 cm from the electromagnet tip, respectively (Table 1, supplementary information). These
values fall within the therapeutic window of 2–250 kPa required
for MEF induction, as previously reported22.
Localizing IMPs in the right ventricle of anesthetized rats
In order to examine the feasibility of localizing intravenously administrated
IMPs in the RV cavity, we administered IMPs through the tail vein of
anesthetized rats while the electromagnet was externally positioned against the
chest of the animal at a distance of ~0.5 cm from the heart
surface and 45° angel relative to the chest wall (Fig.
1a). After the electromagnet was positioned, IMPs were injected and
were allowed to be carried by the blood flow for one minute. Then, the heart was
arrested by intravenous injection of potassium chloride (KCl) and the animal was
frozen while a magnet was still positioned against the chest. The electromagnet
was set to generate a magnetic induction and magnetic induction gradient of
~0.2 T and ~42 T/m, respectively, by applying a
direct current (DC) of 10 A in order to maintain constant attraction.
Figure 2a shows the presence of large IMP aggregates
in the RV cavity of rats wherein the electromagnet was positioned against the
chest wall, while none were seen in sham animals where no electromagnet was
applied to the chest wall (Fig. 2b). A closer look at the
cryosections reveals that IMPs are located in a large portion of the RV
cavity.
Figure 2
Localization of IMPs in the RV cavity by applying an external magnet.
Rats were administered with 15 mg of IMPs via their tail vein, with
(a) or without (b) an electromagnet positioned against
their chest (magnetic induction of ~0.2 T and gradient of
~42 T/m at the heart location). After 1 min, while
magnet was still in action, the rats were sacrificed (KCl), frozen and the
heart was cryo-sectioned. Blue arrows indicate the IMP sediments, while the
brown-red substance that appears in the ventricles is blood. Scale bar is
1 mm. Squares in (a,b) show higher magnification view
of the content in the RV cavity.
Provoking MEF-induced pacing in a rat model ex-vivo and
in-vivo
We first tested the capability of IMPs to provoke cardiac pacing in response to
an alternating magnetic field in an ex vivo Langendorff perfused heart
model. Such model could unravel the set of conditions required for MEF-induced
pacing, with no disturbances of blood flow which can carry away the IMPs from
the RV with time. Additionally, the electromagnet can be easily positioned in
close proximity to the heart. Further, in this preparation we could mechanically
ablate the atrioventricular node in order to obtain low beating rates
originating from infra-nodal regions. Thereafter, IMPs (15 mg) were
injected to the RV cavity through the tricuspid orifice, while the electromagnet
was positioned at a distance of ~0.3 cm, directly against the
front portion of the RV. The efficacy of cardiac pacing was assessed by
measuring the left ventricular (LV) pressure waveform using a balloon inserted
into the LV cavity. By applying magnetic pulses on the accumulated IMPs, they
were abruptly subjected to a magnetic attraction in a pulsatile manner. Thus, we
expected to obtain effective MEF-induced overdrive pacing that would synchronize
the pressure wave with the magnetic pulses.Indeed, square waveforms (5 Hz, 10 A, 20% duty cycle) induced
remarkable overdrive pacing (Fig. 3c). One can see that
before the magnetic pulses were applied (red line), the infra nodal heart rate
was extremely slow (blue line). However, once magnetic pulses were applied, the
heart rate synchronized with the pulses and a heart rate of 300 bpm was
obtained. To visualize this we marked by plus (+) signs the heartbeats that were
synchronized with the magnetic pulses. This MEF-induced pacing indicates a
‘banging’ effect of the IMPs on the RV wall as a result of the
action of the alternating magnetic field. Indeed, immediately after terminating
the magnetic pulses, the heart resumed the slow infra-nodal beating rate.
Overall, we managed to provide effective overdrive pacing in five hearts using
square magnetic pulses.
Figure 3
MEF-induced pacing in a Langendorff perfused heart model.
Illustration of the perfusion setting of the Langendorff perfused heart model
(a) The Tyrode’s buffer is continually oxygenated by
bubbling. The perfusion flow is set by the peristaltic pump to give
perfusion pressure of 80–100 mmHg
(10–15 ml/min). The left ventricular pressure is measured by
a water filled balloon connected to a pressure transducer. A photograph of
an isolated heart with the electromagnet next to it is shown (b). The
photograph was taken for illustration purposes, during mechanical pacing the
electromagnet was positioned against the apex, while the heart was held in
place by a plastic tube. Bradycardia was induced in a Langendorff perfused
heart model by removing the right atrium and applying local pressure on the
atrioventricular node. IMPs were inserted directly into the RV, and magnetic
pulses were then applied (c) The blue line indicates the left
ventricular pressure (LVP), and the red line indicates the current through
the electromagnet coil, which correlates with the magnetic induction
generated by the electromagnet. Plus (+) signs indicate heart beats that are
synchronized with the magnetic pulses.
We also investigated the ability to induce pacing using sine and ramp waveforms
of the same amplitude. These waveforms, however, were far less effective than
square pulses in creating MEF-induced pacing (not shown). Thus, it appears that
the ability to induce pacing with IMPs depends not only on the amount of
accumulated IMPs and the intensity of the magnetic force, but also on the
kinetics by which IMPs are banged on the heart wall. The gradual effect of sine
and ramp waveforms turned out to be insufficient to provoke pacing, whereas the
pulsatile nature of square waveform consistently provoked MEF-induced
pacing.Next, we evaluated the applicability of the new pacing modality in whole rat
model. For this purpose, arterial pressure waves were invasively recorded in the
tail artery. This measurement allowed us to monitor heart beats as well as blood
pressure over time. In this experimental setting, bradycardia was induced by the
alpha-2-adrenergic agonist xylazine. Thereafter, IMPs were injected into the
tail vein while the electromagnet was positioned against the lower part of the
rat sternum, as described in the previous section (Fig.
1a). The electromagnet was first set to accumulate and localize the IMPs
in the RV cavity by constant magnetic induction (10 A for
1 min). Next, square magnetic pulses were applied (5Hz, 20% duty cycle,
and amplitude of AC 0–20A through the coil). These conditions
consistently provoked transient overdrive pacing lasting
4–20 sec. Figure 4a illustrates an example
of such MEF-induced overdrive pacing. Before application of the magnetic pulses,
the heart rate was constant and relatively slow (~240 bpm). Upon
application of magnetic pulses, the heart rate was immediately synchronized with
the pulses, resulting in an overdrive pacing that lasted over 5 s.
Importantly, the MEF pacing counteracted the hemodynamic effect of xylazine,
thereby elevating the blood pressure as also desired clinically. Overall, an
effective pacing of more than 4 s was successfully obtained in 13 out of
15 tested animals. Of note, we did not record ECG in this setting since the
magnetic field distorts this signal during the application of alternating
magnetic fields.
Figure 4
MEF-induced cardiac pacing in an in vivo rat model.
(a) Bradycardia was induced by xylazine and IMPs were injected into
the tail vein. First, the IMPs were captured in the RV by the electromagnet
operating on DC mode. Then, magnetic pulses were applied in order to
mechanically stimulate the heart. The blue line indicates the arterial
pressure (AP), and the red line indicates the current through the
electromagnet coil, which correlates with the magnetic induction generated
by the electromagnet. Plus (+) signs indicate heart beats that are
synchronized with the magnetic pulses (see more detailed information of the
pacing in A in Fig. 3 in supplementary information). To examine the
fading of pacing, similar experiments were repeated, but rats were
sacrificed before (b) or after (c) the magnetic pulses were
applied. Thereafter, animals were frozen and the hearts were cryo-sectioned
to visualize the content of the RV. The IMP sediments are marked with blue
arrows. High magnification (squares) clearly shows that the red substance in
the ventricle of c is blood and it is clear that the amount of IMPs in b is
much larger than that in (c) Scale bar is 1 mm.
We postulated that the fading phenomenon may be attributed to the escape of IMPs
from the RV cavity during the intervals between the applied magnetic pulses;
during these intervals, no field-induced attraction is acting on the
microparticles and they can escape the RV cavity with blood flow. To
substantiate this possibility, we investigated the relation between the amount
of localized IMPs in the RV cavity and their ability to provoke pacing. To this
end, IMPs were localized in the RV using a constant magnetic force. Then, before
or after MEF-induced pacing was applied, the animals were sacrificed, frozen,
and the IMPs content of the RV cavity was visualized. Indeed, when animals were
sacrificed before the induction of pacing, most of the RV cavity was filled with
IMPs (Fig. 4b). However, when animals were sacrificed
after the magnetically-induced pacing faded (Fig. 4c),
only small sediments of IMPs were present in the RV cavity.In an attempt to counteract the escape of IMPs from RV cavity, two strategies
were investigated. In one strategy, a low level of ~5 A DC was
maintained during the intervals between the magnetic pulses. It appears that the
employment of this magnetic force reduced the ability of the pulsatile magnetic
field to induce cardiac pacing (Fig. 5a). To validate
this, we counted the number of heart beats synchronized with the pulses with the
addition of low DC and compared it to the regular 20% duty cycle from Fig. 4. Figure 5b shows that the mean
number of synchronized heartbeats was less than 1 when DC was added and
~23.5 without it. Thus, when DC is added to the time intervals between
pulses, the application of a pulse is not efficient enough to provoke MEF. The
meaning is that the mechanical stimulation applied by the IMPs on the RV wall
must be removed for the consecutive pulse to be efficient. Although this
strategy impaired the ability to provoke MEF-induced pacing, it effectively
managed to prevent the IMPs from being carried away by the blood flow. This can
be realized from the restored ability to provoke transient overdrive pacing
immediately after DC was eliminated indicating the presence of IMPs which were
not carried away during DC application (Fig. 5a, marked
with a green arrow).
Figure 5
Low DC added to the magnetic pulses and high vs. low duty cycle.
Previous experiment (presented in Fig. 4) was repeated
with different wave forms. In (a) we added low DC
(~5 A) to the AC waveform. In (b) we switched from
high (80%) to low (20%) duty cycles (see more detailed information of the
pacing in B in Fig. 4 in supplementary information). The blue line
indicates the arterial pressure (AP), and the red line indicates the current
through the electromagnet coil. Plus (+) signs indicate heart beats that are
synchronized with the magnetic pulses. Note the absence of effective pacing
using both protocols, but effective transient pacing after switching to low
duty cycle without DC, indicating that both protocols retained the IMPs in
the RV (see text for more details). The number of consecutive synchronized
heart beats were counted and the mean + SEM for each group
is plotted (c). Bars are SEM, *p < 0.05.
The second strategy we tested for overcoming the escape of IMPs from the RV,
included the application of high duty cycles, meaning prolonged application of
an attracting magnetic field separated by short intervals without magnetic field
attraction. Figure 5c shows the effect of elevating the
duty cycle of the magnetic pulses from low (20%) to high (80%). The high duty
cycle protocol could generate some extra beats, but did not result in overdrive
pacing. The mean number of heart beats synchronized with the pulses with the
high duty cycle was ~2.2, significantly lower than the ~23.5
synchronized beats counted for the regular 20% duty cycle (Fig.
5b). It thus seems that time intervals between pulses is a key factor
for successful pacing. As mentioned previously, the pulsatile manner of the
magnetic stimulation is of great importance, and in this setting the time
interval between two consecutive pulses (40 ms) was not long enough to
allow the accumulated IMPs to loosen from the RV wall, so that the next pulse
could apply sufficient stimulation. Nonetheless, it did seem to withhold the
IMPs from being carried away by the blood flow, as seen from the transient
overdrive pacing recorded right after the high duty cycle was changed to low
duty without any addition of IMPs (Fig. 5c, marked with a
green arrow). Figure 5b shows that the number of
synchronized pulses when applying the 20% duty cycle was ~10 and
~30 times higher than that that of 80% duty cycle and low DC,
respectively. This prominent difference shows that the 20% duty cycle is the
most efficient wave form we investigated in this study.
Provoking MEF-induced pacing in an in-vivo pig model
To substantiate the feasibility of MEF-based pacing in the large mammalian heart,
we conducted proof-of-concept experiments in pigs. Based on the characterization
of the waveform we found to be effective in the rat model, we used square
waveform with duty cycles of 5–10%, which corresponds to pulses duration
of 30–60 ms when using frequencies of 1.6 Hz
(96 bpm). First, using isolated blood-perfused heart preparations, we
verified that MEF-based pacing can be induced in the large mammalian heart using
our electromagnetic apparatus (Supplementary Fig. 6). Next, in two anesthetized pigs with open chest
(Fig. 6a), IMPs were injected into the femoral vein
while the electromagnet was directed to the apical portion of the RV. We found
that when IMPs were injected while the electromagnet was constantly applying
magnetic pulses, sporadic MEF-based single ectopic beats were induced (Fig. 6b), presumably they were provoked by magnetic
particles passing through the RV and attracted by the pulses. This effect has
faded after a few seconds in a manner consistent with particle washout by the
blood flow. In contrast, when IMPs were injected to the blood flow under
constant magnetic attraction for ~1 min, the application of
magnetic pulses thereafter induced rather long lasting MEF-based overdrive
pacing (Fig. 6c) of over 2 min. Although some
beats coming from the normal pacing system were also noted, MEF-based pacing was
clearly evident in both animals as shown in Fig. 6. After
two minutes of pacing, due to overheating of our electromagnet (which does not
include an efficient cooling strategy), we had to terminate the experiment.
Figure 6
MEF-induced cardiac pacing in an anesthetized pig.
(a) A photograph of the experimental setting. (b) IMPs
(100 mg) were injected to the femoral vein while pulses were applied
through the electromagnet (96 bpm). In this setting only sporadic
events of pacing were noted, presumably since IMPs did not accumulate in the
RV. (c) When IMPs (200 mg) were injected and allowed to
accumulate in the RV for ~1 min by a DC current, overdrive
MEF-based pacing was noted and maintained until the pulses were terminated
due to overheating of the electromagnet. The blue line indicates the
arterial pressure (AP), and the red line indicates the current through the
electromagnet coil, which correlates with the magnetic induction generated
by the electromagnet. Plus (+) signs indicate heart beats that are
synchronized with the magnetic pulses.
Discussion
The present study is a proof-of-concept, showing the feasibility of a novel and
unique approach to induce noninvasive cardiac pacing utilizing injectable magnetic
microparticles and low energy magnetic waveforms applied to the chest wall. This is
to the best of our knowledge, the first demonstration of a selective leadless
approach that can induce in vivo overdrive pacing of the heart without the
use of high electrical energy levels that can be painful and harmful to adjacent
tissues. Moreover, compared to an alternative experimental method of MEF-pacing
using high focused ultrasound, our results demonstrate the actual induction of
overdrive pacing rather than single sporadic ectopic beats20. Thus,
the current approach seems far more attractive in clinical terms.The physiological principle underlying our new modality is based on the MEF
mechanism, according to which the myocardium responds to a mechanical stimulation
resulting from contact, force or pressure. Although MEF research has gained much
interest in the past years, the underlying mechanism is still not completely
known28. The coupling between mechanical stimulation and
electrical activity in the cardiac tissue was first attributed to stretch-activated
ion channels29, but recently, there is growing evidence pointing at
the role of intracellular calcium handling3031. Regardless of the
molecular mechanism, in this study we hypothesized that magnetic microparticles may
be manipulated by an external electromagnet to stay in the RV cavity and to activate
the innate MEF mechanism. This hypothesis was based on the findings of Zoll et
al.32 showing that MEF may be utilized to induce mechanical
pacing. For our proof-of-concept, we performed both ex vivo and in
vivo studies in rat and pig models and proved the feasibility of this new
leadless pacing approach using IMPs.In a step wise manner, at first we proved that IV injected IMPs can be localized in
significant amounts in the RV cavity of intact rats when an external electromagnet
is positioned against the rat’s chest wall. Although we used IMPs that were
rather large in size, it is important to realize that the mechanical stimulation
induced by the magnetic pulses is generated by an aggregate of particles rather than
single particles, so the actual size of the particle has no ramifications in this
context. On the other hand, larger particles are easier to be localized under
magnetic attraction (supplementary
information
Fig. 1). However, those particles were successfully localized
in the RV even when the current was set on 5 A DC (as in Fig.
5a). Thus, it is presumable that IMPs with diameter less than
3 μm may be localized in this setting, as the current through the
coil may be easily increased up to 40 A. Moreover, magnetic particles tend
to aggregate in the presence of magnetic field, which will further aid their
localization.Next, we demonstrated that the application of magnetic pulses from the same
electromagnet can provoke pacing in an isolated heart model. We found that the
approach is viable but is highly dependent on the shape of the applied pulses. When
a pulse of magnetic attraction is applied on an aggregate of IMPs, they initially
accelerate towards the soft tissue, resulting in its deformation; eventually they
are stopped by the tissue when the decelerating force it applies overcomes the
attracting force applied by the electromagnet. The peak force is the maximal force
during the decelerating period that is applied by the tissue on the aggregate, thus
by the aggregate on the tissue as well. The deformation rate and the peak force,
rather than the mean pressure, were shown to play a significant role in provoking
MEF in the cardiac tissue2122. It is clear that these factors are
closely related to the kinetics by which the magnetic pulse is applied. When the
magnetic attraction is gradually applied, the IMP aggregate will deform the tissue
mildly, thus the deceleration force applied on it is minimal. In accordance with
these theoretical considerations, we found that sine and ramp waveforms were not
able to provoke MEF-induced pacing in our experimental system.Next, we tested the feasibility of our pacing approach in an intact rat model.
Settings were based on successful findings in the previous steps. Thus, we used DC
of 10 A through the electromagnet coil to localize the IMPs in the RV and
square waveforms with a frequency of 5 Hz to provoke MEF-induced pacing. Due
to the fact that in the intact animal, the distance of the heart from the
electromagnet tip was slightly larger, we increased the level of the AC amplitude
from 10 to 20 A. The ability of the electromagnet to produce sufficient
levels of magnetic induction and a magnetic induction gradient in the target
location is attributed to the focusing of the magnetic field lines by the narrowed
tip of the core, as illustrated in the computerized simulation of the electromagnet.
Indeed, our results show that the ability to induce non-invasive overdrive pacing is
consistent under these conditions and thus we can conclude that the approach is a
viable one. Although we provided a clear proof of concept to our modality, the
fading of the effect in the present experimental setting is an obvious limitation in
applicable terms.We found evidence that fading most likely results from the escape of the IMPs from
the rat RV cavity in the time durations between the pulses. During these timeframes
there was no magnetic attraction and the IMPs were free to be carried away with the
blood flow. When IMPs are subjected to magnetic pulses, the attracting force depends
on the magnetic moment, which is proportional to the mass. For this reason, the
amount of IMPs is a crucial factor for sufficient magnetically induced mechanical
stimulation. Although we cannot totally exclude the possibility that fading is due
to use dependency of the MEF mechanism, this option seems unlikely since there are
multiple evidence in the literature showing that MEF pacing can last at least
several hours. For instance, in a series of studies33343536
externally applied thumbs on patients’ chest enabled pacing for up to
2 hours and 45 minutes37. In addition, in a preliminary
experiment of the present study an isolated heart was stimulated by a magnetic bead
(2 mm thick and 3 mm diameter) that was inserted into the RV. This
configuration led to MEF-based pacing that was eliminated only when we stopped the
pulses after ~100 seconds (data not shown). This data strongly
suggests that the prominent fading we observed with IMPs was indeed the results of
escape of the IMPs from the RV rather than a use dependent mechanism.Attempting to counteract the escape of the IMPs from the rat RV we used high duty
cycles or low DC that was added to the magnetic pulses. Doing so, we demonstrated
that the magnetic attraction before the application of a magnetic pulse must be
eliminated for a sufficient amount of time in order to generate tissue deformation
and a peak force that will generate a banging effect and provoke MEF. This
conclusion is inferred from the fact that addition of DC current or reducing the
intervals between pulses, as approaches to prevent the escape of IMPs, strongly
impeded the ability to provoke pacing under the current experimental conditions.Although the main focus of the present study was on initial proof-of-concept in rats,
applicability to the human heart is naturally an important issue. In this regard,
there are several considerations that may be relevant. Our modality is based on
mechanical stimulation that correlates to the amount of IMPs localized in the
ventricles. This amount is strongly limited by the small size of the ventricles in
rats. In humans, however, the large size of the RV allows the accumulation of larger
amounts of IMPs that will result in higher levels of mechanical stimulation.
Moreover, the distance between the apical portion of the RV to the Tricuspid valve,
where abrupt changes in blood flow are present, is much larger in humans; thus, one
may expect that the IMPs will be efficiently trapped in the RV cavity as they will
be protected from the blood flow. Of note, despite the larger size of the human
chest, the distance between the RV cavity and the outside of the chest wall is only
~3 cm in humans compared to ~0.5 cm in rats. In
addition, the significantly slower heart rate in humans should allow us to extend
the time gap during which the magnetic force can be diminished or even eliminated
during diastole. During the long diastole of more than 600 ms in human, the
blood fills the ventricles, so the magnetic attraction may be eliminated without
losing the IMPs to the blood flow. Thus, although we showed that the significantly
more effective waveform of 20% duty cycle contributes to the loss of IMPs to the
blood flow in rats, this is likely to be different in humans.Our deduction was substantiated by the results obtained with the pig model. Although
the obtained data (Fig. 6) are still far from clinical
utility, they demonstrate that following accumulation of particles in the RV by
application of DC magnetic force, a durable and rather long lasting MEF-based pacing
was achieved in pig model in comparison to the rat model.Further work is needed in order to optimize the new modality and make it efficient,
long lasting and biocompatible. In regard to the latter, alternative
superparamagnetic particles that are biocompatible exist in the market3839. Other safety issues such as the possibility of calcium overload
as a side effect of the method cannot be excluded at present. However, dealing with
such possibilities is beyond the scope of the present feasibility study and will
have to be done further on. We believe that the proof of concept for our modality,
presented here, provides solid ground to further investigate its feasibility using
magnetite nanoparticles, and using a stronger electromagnet with an efficient
cooling system that will enable usage of this novel modality in the intact pig
model. Of note, there are electromagnets reported in the literature that may
overcome both the challenges of distance to the RV, and overheating26. As described herein, it appears that large mammalian model should allow us to
accumulate larger amounts of magnetic particles and to maintain them longer in the
RV so that the proposed modality will be more durable. In terms of pacing
efficiency, our findings indicate that the selected waveform that is used has a
critical effect on the ability to induce pacing. In this context, the slower heart
rate of the large mammalian heart allows greater flexibility in designing the
optimal waveform. Therefore, a full and more quantitative description of the
dependence of pacing on the selected waveform will be important in order to optimize
the modality and its efficacy.In summary, this study demonstrates the feasibility of a novel and unique approach to
obtain noninvasive cardiac pacing utilizing injectable magnetic microparticles and
low energy magnetic waveforms applied to the chest wall. The encouraging findings
presented herein may open a window of opportunities for the use of non-invasive
pacing in several important clinical scenarios such as symptomatic bradyarrhythmias,
acute treatment of reentry-dependent tachyarrhythmias, temporary pacing in the
setting of pacemaker extraction due to bacteremia, and more. The results of this
study should lay the foundations for further refinement of this novel modality in
the near future.
Methods
Iron magnetic microparticles (IMPs) were purchased from Sigma-Aldrich (Rechovot,
Israel). The IMPs’ size distribution was determined by static light
scattering using a Fritsch Analysette 22 MicroTec Plus (Oberstein, Germany)
laser diffraction analyzer. Their morphology was determined by scanning electron
microscopy (SEM; JEOL model JSM-35CF, Tokyo, Japan).
Electromagnet design
An electromagnet comprising a coil and a metal core was custom-made for this
study. The core was made out of permendur (Goodfellow, Huntingdon, UK), a soft
ferromagnetic iron-cobalt-vanadium (49:49:2) alloy with extremely high magnetic
induction at saturation (~2.34 T) and low coercivity
(80 A/m), which is necessary for generating alternating magnetic fields.
The coil body was made of a fabric-base laminate composed of 900 windings of
isolated copper wire (diameter 2 mm). The core and the coil were 13 and
10 cm long, respectively. The 3 remaining cm were narrowed to a
5 mm diameter tip, as illustrated in Fig. 1d, in
order to focus the magnetic induction stream lines (core and coil machining was
performed in the Ben-Gurion University workshop, Beer Sheva, Israel).The magnetic induction generated by the electromagnet was simulated using COMSOL
Multiphysics software (COMSOL Inc., Burlington, MA). The simulation was
performed by solving equations 1 and 2.Where H is the magnetic field (A/m), B is the magnetic flux density (T),
Je is the current density (A/),
σ is the conductivity, υ is the velocity and A is the magnetic
vector potential. The properties of the air and copper wire were obtained from
the software material library, while B-H curve data for permendur were taken
from Cobb et al.40, and the current through the coil was
set to be 10 A. The coil was connected to a power amplifier (AE Techron
7224, Elkhart, IN) operating in current control mode and connected to a function
generator (GW Instek, Taiwan). The compensation of the amplifier was performed
by Erantel Electronics, Kfar Saba, Israel.
Animal studies
Rodent studies were performed using adult Sprague Dawley rats (300–450gr)
of both sexes, with the approval and according to the guidelines of the
Institutional Animal Care and Use Committee of the Ben-Gurion University,
Israel. Pig experiment were done in L.R.I., Comprehensive Pre-clinical Services
facility, Kibbutz Lahav, Israel using farm pigs weighing ~60 kg.
The pig experiments were approved by the National Animal Care and Use Committee
of Israel, and were carried out in accordance with the approved guidelines. At
the end of all in vivo experiments the animals were sacrificed by
intravenous KCl administration under deep anesthesia.
Localizing IMPs in the RV cavity using an in-vivo rat
model
To verify the feasibility of localizing IMPs in the rat RV by an external magnet,
animals were anesthetized (ketamine, 100 mg/kg IP), and the
electromagnet was positioned against the apical portion of their heart. The
location of the apex was determined according to an external manual assessment.
The alignment of the electromagnet was 45° to all planes (transverse,
frontal, and sagittal). After the electromagnet was set to maintain constant
magnetic attraction by applying constant current through the electromagnet coil
(10 A), IMPs were injected into the tail vein (15 mg in
1 ml phosphate buffered saline; PBS; Biological Industries, Beit
Ha-emek, Israel). Then, the IMPs were allowed to be localized in the RV by the
external electromagnet for ~1 min. Thereafter, the rats were
sacrificed by injection of KCl solution (0.5 ml of 15% KCl w/v)
and frozen at −20 °C while a neodymium N-52 magnet
(5 mm diameter and 25 mm long; axially magnetized, K&J
Magnetics, Pipersville, PA) was positioned instead of the electromagnet to
prevent the IMPs from moving. The frozen body was opened and the heart was
removed and fixed in optimal cutting temperature (OCT) compound, and
cryosections of the heart were made to visualize the IMPs in the ventricles.
Negative controls were performed by injecting the IMPs suspension without the
electromagnet.Of note, although cryosections are often washed from blood in order to make the
picture clearer, the heart sections in our experiments could not be washed due
to the fact that the entities of interest (i.e. IMPs) in the ventricular cavity
were not bound to the tissue.
Langendorff-perfused heart model
A rat was heparinized (1000 IU/kg IP, Kamada, Beit Kama, Israel) and
anesthetized (pentobarbital, 60 mg/kg IP, CTS, Hod-Hasharon, Israel).
The heart was removed and the aorta was cannulated. Oxygenated Hepes-buffered
Tyrode’s solution (containing, in mM: NaCl 126, KCl 5.4, Glucose 10,
Hepes 10, MgCl2 1, CaCl22, MgSO4 1.2,
NaH2PO4 0.39; bubbled with 99.5% O2,
Maxima, Ashdod, Israel; pH tittered to 7.4 by 2M NaOH; (all salts from Sigma)
was perfused through the cannulated aorta. Perfusion rate was controlled by a
peristaltic pump to obtain perfusion pressure of 80–100 mmHg
(~10–15 ml/min). The left atrium was removed and a
saline filled latex balloon was inserted into the LV and inflated to an
end-diastolic pressure of 5–10 mmHg (see Fig.
3a). Pressure signal was recorded digitally (2 kHz) by
connecting the intraventricular balloon to a pressure recording system (ETH-256C
amplifier and BP-100 probe, iWorx, Dover, NH), and the resulting pressure
recordings were interfaced with a PC using an A/D converter (USB-6008, National
Instruments, Austin, TX) and a homemade program developed by Y.E. (using LabView
7.1, National Instruments) to control signal acquisition, data saving and
off-line analysis41.
Provoking MEF-induced pacing in an isolated heart model
After the Langendorff setting was performed, the right atrium was removed and the
RV was exposed. IMPs (15 mg in 1 ml PBS) were injected directly
to the RV. In order to prevent the IMPs from leaking out of the RV, a neodymium
magnet was placed against the apex. After the IMPs were localized in the RV, an
alternating magnetic field was applied using the electromagnet. In order to
prevent the heart from moving due to the magnetic pulses, it was held in place
by an external plastic tube. Several different waveforms (sin, square and ramp)
were used to investigate the optimal setting for stimulating the heart. The LV
pressure (LVP) was measured by the pressure probe in order to verify
synchronization between the magnetic pulses and the heart pacing. The waveform
of the magnetic pulses stimulating the heart was set by a function generator
connected to the power amplifier. For the ex-vivo Langendorff perfused
heart model (Fig. 3c) we used a low duty cycle (20%)
square waveform, where the current in the coil was 10 A for
40 ms followed by 160 ms without current.
Provoking MEF-induced pacing in an In-Vivo rat model
To prove the feasibility of the proposed methodology we used an in vivo
rat model. Rats were anesthetized (ketamine, 100 mg/kg IP,
Vétoquinol, France) and an intravenous (IV) line was inserted into the
tail vein. Then, the tail artery was exposed by dissection and the artery was
cannulated for arterial pressure (AP) measurements. Bradycardia was induced by
administering xylazine (10 mg/kg IP, Eurovet, Bladel, The Netherlands). The
electromagnet was positioned against the apical portion of the RV while the
location of the cardiac apex was determined according to an external manual
assessment. IMPs (15 mg in 1 ml PBS) were administered via the
IV line while the current in the electromagnet coil was set to be constant,
yielding constant magnetic attraction. After the IMPs accumulated in the RV
(30–60 s) the current in the electromagnet was switched to give
a pulsatile waveform, resulting in magnetic pulses.AP recording in the tail artery was done using ETH-256C amplifier and BP-100
probe (iWorx, Dover, NH). The pressure signals were interfaced with a PC using
an A/D converter (USB-6008, National Instruments, Austin, TX) and a homemade
program developed by Y.E. as described for the Landerhoff model. The waveform of
the magnetic pulses stimulating the heart was set by a function generator
connected to the power amplifier. For the in vivo experiments of Fig. 4a, a DC of 10 A was applied until the IMPs accumulated
in the heart, followed by alternating current of low duty cycle (20%) square
waveform (0–20 A).In order to visualize the IMP content after the MEF-induced pacing faded (Fig. 4b,c), the same experimental protocol was used.
However, after pacing ceased to be achieved, the rat was scarified by KCl
injection (0.5 ml, 15% w/w), and frozen while a neodymium N-52
magnet was positioned instead of the electromagnet to prevent the IMPs from
moving. The frozen body was opened and the heart was removed and fixed in OCT;
cryosections of the heart were made to visualize the IMPs in the ventricles.
Visualizing the IMPs content before applying MEF-induced pacing was performed by
injecting the KCl solution while the electromagnet was on DC mode, and before
any pulses were applied. Cryosections were made and viewed as described for the
preliminary IMP localizing experiment.For the in-vivo experiments of Fig. 5a, a DC of
10 A was applied until the IMPs accumulated in the heart, and then was
reduced to 5 A as it was during the intervals. Thereafter, an
alternating current of low duty cycle (20%) square waveform
(5–25 A) was applied on the accumulated IMPs. After no pacing
was achieved the DC current was stopped, resulting in a low duty cycle (20%)
square waveform ranging from 0 to 20 A. For the in-vivo
experiments of Fig. 5c, a DC of 10 A was applied until the
IMPs accumulated in the heart, followed by an alternating current of high duty
cycle (80%) square waveform (0–20 A). After no overdrive pacing
was achieved the current was set back to the 10 A DC, followed by alternating
current of low duty cycle (20%) square waveform (0–20 A). The
waveforms are visually illustrated in the relevant figures. Of note, the main
reason that we did not routinely try to use duty cycles below 20% for the pulses
is since we used an amplifier that was limited to a voltage of 180 V.
Due to the high inductance of the electromagnet’s coil (900 windings and
a ferromagnetic core with high magnetic permeability) the voltage limitation
results in a current change that was limited to approximately 1 A/ms.
Therefore, in order to apply a current pulse with an amplitude of 20 A
the rise time alone, as well as the decay time, was ~20 ms.
Thus, generating very short pulses was not feasible in our setting. However, in
some cases we did try to use duty cycle of 10% and found that it was feasible to
pace using such pulses of shorter duration (not shown).
Counting the number of synchronized pulses
In order to determine whether a heartbeat was synchronized with the magnetic
pulses we measured the time gap between consecutive heartbeats using Microsoft
Excel software. Due to the xylazine induced bradycardia, the base line heart
rate was slower than 5 Hz. For this reason we determined any consecutive
heartbeats with an interval of less than 210 ms as capture of pacing
(the plus signs in Figs 4, 5 and
6). To determine the number of synchronized
pulses we manually counted the number of such consecutive pulses. Due to the
fact that in some cases the set of consecutive synchronized heartbeats was
separated by random occurrence of unsynchronized heartbeats, we continued
counting if the next synchronized pulse occurred no more than 1 s after the last
one. The group sizes for the low duty cycle, high duty cycle, and low DC added
were 18, 6 and 4 respectively; they were all biological duplicates.
Provoking MEF-induced pacing in an In-Vivo pig model
Pig studies were done on ventilated deeply anesthetized animals (pig weighing
60 kg) under continuous isoflurane inhalation. Arterial line and
intravenous port were inserted in the femoral artery and vein, respectively.
Thereafter, the chest was opened and the electromagnet was positioned against
the RV. In order to prevent direct contact of the electromagnet tip with the
heart the heart was protected by a transparent plastic shield that prevented
direct contact of the tip with the heart, which could possibly induce
MEF-pacing. IMPs were injected into the IV line (100 mg for Fig.
6b and 200 mg for Fig. 6c) and the
electromagnet was set to apply pulses or constant current according to the
experimental protocol. In this setting, we used two amplifiers (AE Techron 7224,
Elkhart, IN) operating in parallel on a voltage control mode and connected to
the same function generator (GW Instek, Taiwan). The heart rate, O2
saturation, arterial pressure and body temperature were continuously monitored
throughout the experiment (VitaLogic 6000, Mennen Medical Ltd, Yavne, Israel).
Electrical current through the coil and pressure signal obtained from the
monitor were recorded digitally (200 Hz) with a PC using an A/D
converter (PCI-6024E, National Instruments, Austin, TX) and a homemade program
developed by Y.E. (using LabView 7.1, National Instruments) to control signal
acquisition, data saving and off-line analysis41. Pacing was
induced by applying short square pulses of 10% duty cycle.
Statistical analysis
Statistical analysis was performed with GraphPad Prism version 5.03 for Windows
(GraphPad Software, San Diego, CA). All variables are expressed as
mean + SEM. The number of synchronized heartbeats were compared
by one-way ANOVA with Tukey’s post-hoc test. P < 0.05
was considered statistically significant.
Additional Information
How to cite this article: Rotenberg, M. Y. et al. Feasibility of
Leadless Cardiac Pacing Using Injectable Magnetic Microparticles. Sci. Rep.
6, 24635; doi: 10.1038/srep24635 (2016).
Authors: Anna Pia Pastore; Alfredo De Leo; Roberto De Leo; Giuseppe Della Chiara; Valter Mariani Primiani; Franco Moglie; Graziano Cerri Journal: IEEE Trans Biomed Eng Date: 2010-08-09 Impact factor: 4.538
Authors: P R Ros; P C Freeny; S E Harms; S E Seltzer; P L Davis; T W Chan; A E Stillman; L R Muroff; V M Runge; M A Nissenbaum Journal: Radiology Date: 1995-08 Impact factor: 11.105
Authors: Petra Dames; Bernhard Gleich; Andreas Flemmer; Kerstin Hajek; Nicole Seidl; Frank Wiekhorst; Dietmar Eberbeck; Iris Bittmann; Christian Bergemann; Thomas Weyh; Lutz Trahms; Joseph Rosenecker; Carsten Rudolph Journal: Nat Nanotechnol Date: 2007-07-22 Impact factor: 39.213