Wenjie Cheng1, Xiang Gao2, Weili Wang1, Min Zhi2, Jian Tang2, Yan-Ling Wen1, Junli Yu1, Yao Chen1, Xiaoyin Liu1, Chuan Yang1, Pinjin Hu2, Guangjian Liu3.
Abstract
To assess influencing factors for quantitative analysis of contrast-enhanced ultrasound (CEUS) in Crohn's disease (CD), dynamic CEUS examinations from 77 consecutive CD patients were recorded. Peak intensity (PI) values were calculated using the pre-installed quantification software of the ultrasound scanner. The influence of depth, pressure from the ultrasound probe and intraluminal gas was analyzed. The PI value of the anterior wall was lower than that of the posterior wall when the depth was ≤3.4 cm (17.9 dB vs. 21.3 dB; p < 0.05) or evident pressure was exerted (19.1 dB vs. 22.5 dB; p < 0.01). In the presence of intraluminal gas, the PI of the anterior wall was higher than that of the posterior wall (20.7 dB vs. 18.8 dB; p < 0.05). Nevertheless, no significant difference was found between the PI value of anterior and posterior walls when the depth was >3.4 cm (19.8 dB vs. 20.3 dB), moderate pressure was exerted (20.5 dB vs. 21.1 dB) or luminal gas was excluded between the two bowel walls (18.9 dB vs. 21.2 dB; p ≥ 0.05). The factors of depth, pressure from the ultrasound probe and intraluminal gas can affect the quantification results of CEUS. It is preferable to place the region of interest in the posterior wall when luminal gas is absent and in the anterior wall when luminal gas is present. In the latter case, more attention should be paid to reducing pressure by the ultrasound probe.Entities:
Keywords: Contrast-enhanced ultrasound; Crohn's disease; Influencing factors; Peak intensity; Quantification; Time intensity curves
Mesh:
Substances:
Year: 2016 PMID: 27087694 DOI: 10.1016/j.ultrasmedbio.2016.03.001
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998