Comparison Between an Automated and Manual Extraction for the Determination of Immunosuppressive Drugs Whole Blood Concentrations by Liquid Chromatography Tandem Mass Spectrometry.

Guillaume Deslandes1, Matthieu Grégoire1, Christian Renaud1, Catherine Monteil-Ganière1, Claudine Azoulay1, Alain Pineau1, Pascale Jolliet1, Eric Dailly2.   

Abstract

BACKGROUND: The whole blood extraction for liquid chromatography tandem mass spectrometry (LC-MS/MS) of simultaneous quantification of cyclosporine A (Cys A), tacrolimus (Tacrs), sirolimus (Siros), and everolimus (Evers) is still performed manually in many laboratories. The analytical results obtained with an automated method using a liquid handler versus a classical manual preparation were compared.
METHODS: Cys A (n = 36), Tacrs (n = 50), Siros (n = 34), Evers (n = 38) whole blood samples of patients were analyzed by LC-MS/MS assay after manual preparation and automated process using a liquid handling platform including a centrifugation step.
RESULTS: The comparison between manual and automated extraction investigated by a linear regression showed a high correlation between results [(Tacrs "automated") = 1.0927 × (Tacrs "manual") - 0.36; (Cys A "automated") = 1.0284 × (Cys A "manual") + 0.0312; (Siros "automated") = 0.9923 × (Siros "manual") + 0.4001; (Evers "automated") = 1.0000 × (Evers "manual") - 0.0600].
CONCLUSION: The results obtained by the automated and manual preparation are consistent. The automated method is applied for high-throughput therapeutic drug monitoring of immunosuppressive drugs in routine practice, leading to an increase in quality.
© 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  automation; cyclosporine A; everolimus; sirolimus; tacrolimus

Mesh:

Substances:

Year:  2016        PMID: 27086934      PMCID: PMC6807118          DOI: 10.1002/jcla.21958

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  7 in total

1.  Liquid chromatography-tandem mass spectrometry or automated immunoassays: what are the future trends in therapeutic drug monitoring?

Authors:  Gunnar Brandhorst; Michael Oellerich; Gregory Maine; Paul Taylor; Gerhard Veen; Pierre Wallemacq
Journal:  Clin Chem       Date:  2012-01-17       Impact factor: 8.327

2.  Validation of a fully automated robotic setup for preparation of whole blood samples for LC-MS toxicology analysis.

Authors:  David Andersen; Brian Rasmussen; Kristian Linnet
Journal:  J Anal Toxicol       Date:  2012-04-02       Impact factor: 3.367

Review 3.  The current role of liquid chromatography-tandem mass spectrometry in therapeutic drug monitoring of immunosuppressant and antiretroviral drugs.

Authors:  Paul J Taylor; Chun-Hui Tai; Michael E Franklin; Peter I Pillans
Journal:  Clin Biochem       Date:  2010-07-01       Impact factor: 3.281

4.  Hand and shoulder ailments among laboratory technicians using modern plunger-operated pipettes.

Authors:  M G Björksten; B Almby; E S Jansson
Journal:  Appl Ergon       Date:  1994-04       Impact factor: 3.661

5.  Automated processing of whole blood samples for the determination of immunosuppressants by liquid chromatography tandem-mass spectrometry.

Authors:  Michael Vogeser; Ute Spöhrer
Journal:  Clin Chem Lab Med       Date:  2006       Impact factor: 3.694

6.  An automated SPE/LC/MS/MS method for the analysis of cocaine and metabolites in whole blood.

Authors:  Eshwar Jagerdeo; Madeline A Montgomery; Marc A Lebeau; Martin Sibum
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2008-09-07       Impact factor: 3.205

7.  Immunosuppressant therapeutic drug monitoring by LC-MS/MS: workflow optimization through automated processing of whole blood samples.

Authors:  Mariela Marinova; Carlo Artusi; Laura Brugnolo; Giorgia Antonelli; Martina Zaninotto; Mario Plebani
Journal:  Clin Biochem       Date:  2013-09-05       Impact factor: 3.281

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.