Literature DB >> 27083689

Optimization of the pediatric head computed tomography scan image quality: Reducing dose with an automatic tube potential selection in infants.

Baptiste Morel1, Aurélien Bouëtté2, Pierre Lévy3, Guillemette Antoni4, François Chalard2, Eleonore Blondiaux2, Hubert Ducou Le Pointe2.   

Abstract

PURPOSE: The objective of our study was to evaluate the impact of an automatic tube potential selection (ATPS) on the delivered dose and image quality in unenhanced head computed tomography (CT) scans of infants.
MATERIALS AND METHODS: Unenhanced head CT scans were acquired before and after the introduction of an ATPS in full automatic mode in two groups of 20 patients under one year of age. The delivered dose (CDTIvol) as the quantitative (contrast-to-noise ratio) and qualitative (based on the European CT criteria) image quality were compared on the supra- and infratentorial regions by three senior pediatric radiologists. Mann-Whitney and Fisher exact tests were performed. An interobserver Fleiss's kappa agreement was calculated for each criterion.
RESULTS: The use of an ATPS allowed a significant reduction in the delivered dose (-21%, p=0.0005) with no significant difference of the contrast-to-noise ratio in supra- (-5%, p=0.21) and infratentorial regions (+16%, p=0.96). In all cases, dose reduction was obtained with the same value of 100kV. It maintained a good qualitative image quality (e.g., differentiation between gray and white matter in supra-tentorial region: p=0.470). The interobserver Fleiss's kappa agreements were good to excellent.
CONCLUSION: ATPS is a tool that can significantly reduce the delivered dose by choosing the most appropriate tube voltage while maintaining image quality in unenhanced head CT scans of infants.
Copyright © 2016 Elsevier Masson SAS. All rights reserved.

Entities:  

Keywords:  Automatic tube potential selection; Image quality; Pediatric head CT scan; Radiation exposure

Mesh:

Substances:

Year:  2016        PMID: 27083689     DOI: 10.1016/j.neurad.2016.03.005

Source DB:  PubMed          Journal:  J Neuroradiol        ISSN: 0150-9861            Impact factor:   3.447


  1 in total

1.  A Preliminary Study of Personalized Head CT Scan in Pediatric Patients.

Authors:  Bian Bingyang; Wang Gang; Shao Zhiqing; Nan Li; BoXu Zhou; ShuJia Xu; Dan Li
Journal:  Dose Response       Date:  2021-03-04       Impact factor: 2.658

  1 in total