Abstract
A simple, sensitive, and rapid detection of glutathione by cyclic voltammetry using a bare glassy carbon electrode is reported in which glutathione forms a 1 : 1 complex compound with copper(ii) ions. This complex compound is adsorbed onto the electrode surface and undergoes electrochemical oxidation at a characteristic oxidation potential of ca. -0.20 V vs. the standard mercury/mercurous sulphate reference electrode, which is used to detect the glutathione concentration. The linear dynamic range is obtained for a glutathione concentration from 1 μM to 12.5 μM, and the sensitivity is found to be 0.1 ± 0.002 μA μM(-1). A low limit of detection (n = 3) of 0.14 μM and a precision of 1.8% are achieved using a simple, unmodified electrode. The robustness of the present methodology is demonstrated by the successful quantitative analysis of glutathione in the presence of cysteine.Entities:
Year: 2016 PMID: 27074944 DOI: 10.1039/c6an00550k
Source DB: PubMed Journal: Analyst ISSN: 0003-2654 Impact factor: 4.616