| Literature DB >> 27074463 |
Yongbok Kim1, Joseph M Modrick, Edward C Pennington, Yusung Kim.
Abstract
The objective of this work is to present commissioning procedures to clinically implement a three-dimensional (3D), image-based, treatment-planning system (TPS) for high-dose-rate (HDR) brachytherapy (BT) for gynecological (GYN) cancer. The physical dimensions of the GYN applicators and their values in the virtual applicator library were varied by 0.4 mm of their nominal values. Reconstruction uncertainties of the titanium tandem and ovoids (T&O) were less than 0.4 mm on CT phantom studies and on average between 0.8-1.0 mm on MRI when compared with X-rays. In-house software, HDRCalculator, was developed to check HDR plan parameters such as independently verifying active tandem or cylinder probe length and ovoid or cylinder size, source calibration and treatment date, and differences between average Point A dose and prescription dose. Dose-volume histograms were validated using another independent TPS. Comprehensive procedures to commission volume optimization algorithms and process in 3D image-based planning were presented. For the difference between line and volume optimizations, the average absolute differences as a percentage were 1.4% for total reference air KERMA (TRAK) and 1.1% for Point A dose. Volume optimization consistency tests between versions resulted in average absolute differences in 0.2% for TRAK and 0.9 s (0.2%) for total treatment time. The data revealed that the optimizer should run for at least 1 min in order to avoid more than 0.6% dwell time changes. For clinical GYN T&O cases, three different volume optimization techniques (graphical optimization, pure inverse planning, and hybrid inverse optimization) were investigated by comparing them against a conventional Point A technique. End-to-end testing was performed using a T&O phantom to ensure no errors or inconsistencies occurred from imaging through to planning and delivery. The proposed commissioning procedures provide a clinically safe implementation technique for 3D image-based TPS for HDR BT for GYN cancer.Entities:
Mesh:
Year: 2016 PMID: 27074463 PMCID: PMC5874852 DOI: 10.1120/jacmp.v17i2.5818
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Reconstruction of titanium T&O applicator on MR images: (a) T2‐weighted MRI with 3 mm slice thickness; (b) T1‐weighted MRI with clear flange information; (c) fused MRI (T1‐ and T2‐weighted); (d) X‐ray prior to MR scan; (e) X‐ray following MR scan; (f) tandem tip of reconstructed applicator is described in detail on MR images with MR intensity profile. The flange position in (d) and (e) enables to calculate Δd. From the maximum intensity change point, we account for 1 mm artifacts and another 1 mm for catheter thickness.
Figure 2In‐house independent HDR dose calculation and HDR plan quality verification tool compared to the TPS plan for vagina cylinder HDR.
Ten clinical T & O cases to compare volume‐based optimization with line‐based optimization. Three parameters were compared between two plans: total reference air KERMA (TRAK), volume receiving 100% prescription dose , and Point A dose averaged between left and right points. The values of volume‐based optimized plan were normalized by those of line‐based optimized plan
|
|
|
|
|
|
|
|
|
|
|
|
|---|---|---|---|---|---|---|---|---|---|---|
| Prescription dose (Gy) | 7 | 7 | 7 | 5.5 | 7.4 | 7 | 7 | 5.5 | 5.5 | 6 |
| Active tandem length (cm) | 7 | 6 | 6 | 7 | 5.5 | 7 | 6.5 | 6.5 | 6.5 | 6.5 |
| Ovoid size (cm) | 2.5 | 2 | 1.6 | 2 | 2 | 2 | 2 | 2 | 2.6 | 2 |
|
| ||||||||||
| TRAK(cGy&mdot;cm2) | 1.01 | 1.00 | 0.99 | 0.99 | 1.02 | 1.01 | 1.02 | 0.99 | 1.02 | 1.03 |
|
| 1.13 | 1.02 | 0.86 | 1.01 | 1.05 | 1.05 | 1.05 | 1.03 | 1.06 | 1.07 |
| Average Point A Dose | 0.99 | 0.99 | 0.98 | 0.94 | 1.01 | 1.00 | 1.00 | 1.00 | 1.00 | 1.00 |
Figure 3Comparison of dose distributions of different volume optimization techniques of graphical optimization (a), pure inverse optimization (b), and hybrid inverse optimization (c) when compared to that of conventional Point A planning technique (d) for a T&O case. Optimization times spent for each technique are presented (e).
Measurements of physical tip thickness on CT images and their comparison with nominal values and values in applicator library for T&O and vaginal cylinder (VC) applicator
|
|
| ||||||||
|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
| |
| Nominal thickness (NT) (mm) | 0.95 | 0.95 | 0.95 | 0.3 | 5.4 | 6.5 | 7.7 | 9.2 | 11.1 |
| Applicator library thickness (ALT) (mm) | 0.6 | 0.9 | 0.9 | 0.3 | 5.3 | 6.4 | 7.6 | 9.2 | 11.1 |
| Measured thickness (MT) (mm) | 1.0 | 1.2 | 1.3 | 0.6 | 5.5 | 6.6 | 7.8 | 9.4 | 11.2 |
| Difference: (NT)‐(ALT) (mm) | 0.35 | 0.05 | 0.05 | 0.0 | 0.1 | 0.1 | 0.1 | 0 | 0 |
| Difference: (NT)‐(MT) (mm) |
|
|
|
|
|
|
|
|
|
Figure 4The values of for 19 T&O plans (a) between pre‐X‐ray and MRI; (b) between MRI and post‐X‐ray; (c) between pre‐X‐ray and post‐X‐ray The overall reconstruction uncertainties of T&O on MRI are on average less than 1.1 mm regardless of control datasets (i.e., pre‐ or post‐X‐ray). In addition, panel (c) demonstrates the overall reconstruction uncertainties of T&O between two control X‐ray datasets are on average less than 0.6 mm.
Figure 5Imported CT ((a) and (b)) and high resolution (3 Tesla) MRI ((c) and (d)) dataset's integrity was checked with known dimensions using QA phantom.
Figure 6Comparison of rectal dose‐volume histograms between BrachyVision TPS and PlanUNC TPS.
Optimization sensitivity to optimization times for 10 s, 30 s, 1 min, and 1.5 min relative to the reference (Ref) optimization time of 2 min. Four parameters are compared: each dwell time, source activity, total air Kerma strength, and total treatment time. % difference is defined as: )
|
| ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
| |
| Tandem | 119 | 63.7 | 64.3 | 63.9 | 63.8 | 63.8 | 0.2% |
|
| 0.0% |
| 118.5 | 62 | 63.1 | 62.6 | 62.6 | 62.6 | 1.0% |
| 0.0% | 0.0% | |
| 118 | 59.8 | 60.7 | 60.9 | 60.9 | 61 | 2.0% | 0.5% | 0.2% | 0.2% | |
| 117.5 | 57.1 | 58.8 | 59.1 | 59.1 | 59.1 | 3.4% | 0.5% | 0.0% | 0.0% | |
| 117 | 54.7 | 57 | 57.3 | 57.3 | 57.4 | 4.7% | 0.7% | 0.2% | 0.2% | |
| 116.5 | 53.2 | 55.2 | 55.7 | 55.8 | 55.8 | 4.7% | 1.1% | 0.2% | 0.0% | |
| 116 | 53.7 | 54.4 | 54.5 | 54.6 | 54.6 | 1.6% | 0.4% | 0.2% | 0.0% | |
| 115.5 | 54.6 | 53.8 | 53.7 | 53.8 | 53.8 |
| 0.0% | 0.2% | 0.0% | |
| 115 | 55.8 | 53.8 | 53.4 | 53.4 | 53.4 |
|
| 0.0% | 0.0% | |
| 114.5 | 56.7 | 53.4 | 53.5 | 53.5 | 53.4 |
| 0.0% |
|
| |
| Right Ovoid | 120 | 34.6 | 35.5 | 34.7 | 34.6 | 34.6 | 0.0% |
|
| 0.0% |
| 119.5 | 34.3 | 35.1 | 34.8 | 34.8 | 34.7 | 1.2% |
|
|
| |
| 119 | 35 | 35.1 | 35 | 35 | 34.9 |
|
|
|
| |
| 118.5 | 35.7 | 35.1 | 35.2 | 35.2 | 35 |
|
|
|
| |
| Left Ovoid | 120 | 32.7 | 32.8 | 32.8 | 32.8 | 32.8 | 0.3% | 0.0% | 0.0% | 0.0% |
| 119.5 | 32.5 | 32.9 | 32.9 | 32.9 | 33 | 1.5% | 0.3% | 0.3% | 0.3% | |
| 119 | 31.9 | 32.8 | 33.1 | 33.1 | 33.1 | 3.6% | 0.9% | 0.0% | 0.0% | |
| 118.5 | 31.3 | 33.1 | 33.2 | 33.2 | 33.3 | 6.0% | 0.6% | 0.3% | 0.3% | |
|
| ||||||||||
|
|
|
|
|
|
|
|
|
| ||
| 4670.6 | 4670.6 | 4670.6 | 4670.6 | 4670.6 | 0.0% | 0.0% | 0.0% | 0.0% | ||
|
| ||||||||||
|
|
|
|
|
|
|
|
|
| ||
| 4388.3 | 4428.0 | 4424.9 | 4425.4 | 4425.9 | 0.9% | 0.0% | 0.0% | 0.0% | ||
|
| ||||||||||
|
|
|
|
|
|
|
|
|
| ||
| 839.3 | 846.9 | 846.3 | 846.4 | 846.5 | 0.9% | 0.0% | 0.0% | 0.0% | ||
Percent deviation of DVH parameters for the plans obtained with graphical, pure inverse, and hybrid inverse optimizations from a conventional Point A plan. Compared to conventional Point A plan, cells in blue color represent better plan while cells in red color present worse plan with respect to either target coverage or OARs sparing. Hence, blue cell means higher dose for the target and lower dose for OARs
|
|
|
|
| ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| 1 | 37.8 | 4.7% (3.7 Gy) | 0.6% (0.5 Gy) | 0% | 0% | 7.8% (6.1Gy) | 2.7% (2.0Gy) | 0% | 0% | 3.9% (3.1Gy) | 0% | 0% | 0% |
| 2 | 40.6 | 5.8% (4.1 Gy) | 3.6% (2.7 Gy) | 0% | 0% | 4.3% (3.1Gy) | 2.9% (2.2Gy) | 0% | 0% | 3.4% (2.4Gy) | 1.5% (1.1Gy) | 0% | 0% |
| 3 | 37.2 |
|
| 0% |
| 6.5% (4.9Gy) | 2.5% (1.9Gy) | 7.7% (6.9Gy) |
| 2.4% (1.8Gy) |
| 0% |
|
| 4 | 10.2 | 0% |
| 0% | 0% | 0% |
| 0% | 0% | 0% |
| 0% | 0% |
| 5 | 3.9 | 0% | 0% | 0% |
| 0% | 0% | 0% |
| 0% | 0% | 0% |
|
| 6 | 28.7 | 0% | 0% |
| 0% | 0% | 0% | 5.7% (5.8Gy) | 0% | 0% | 0% |
| 0% |
| 7 | 99.7 | 2.9% (2.0 Gy) |
| 0% | 0% | 5.2% (3.6Gy) | 14.3% (11.0Gy) | 32.6% (24.8Gy) | 0% | 3.4% (2.4Gy) |
| 0% | 0% |
| 8 | 57.5 | 4.6% (3.6 Gy) | 0% | 0% |
| 6.7% (5.2Gy) | 0% | 0% |
| 2.9% (2.3Gy) | 0% | 0% |
|
a Rectum, Bladder, and Sigmoid dose to 2cc volume of rectum, bladder, and sigmoid, respectively.
HR‐CTV minimum dose to cover 90% of high risk CTV volume.
Summary of tests for commissioning of 3D image‐based HDR brachytherapy TPS for VC and T&O cases of gynecological cancer treatment
|
|
|
|
|
|---|---|---|---|
|
| |||
| D1 Remote after loading unit (RAU) setup | Properly setup | (20) | |
| D2 HDR source setup | Properly setup | (20) | |
| D2.1 TG43 parameters check | Properly setup | (28) | |
| D2.2 TG43 calculations |
| (15) | |
| D3 Isodose display test & setup | Functional | ||
| D4 3D dose display & DVH test | Functional | (20) | |
| D5 T&O dose calculation: consistency |
| (15) | |
| D5.1 Dwell times (total & each dwell position) |
| (15) | |
| D5.2 TRAK (Total air KERMA strength) |
| (15) | |
| D5.3 Dose calculation reproducibility |
| (15) | |
| D5.4 DVH comparison |
| (20) | |
| D6 VC dose calculation: consistency |
| (15) | |
| D6.1 Dwell times (total & each dwell position) |
| (15) | |
| D6.2 TRAK (Total air KERMA strength) |
| (15) | |
| D6.3 Dose calculation reproducibility |
| (15) | |
| D7 Independent dose calculation consistency | |||
| (’HDRCalculator’ as in‐house software) |
| (15) | |
|
| |||
| N1 Configure DICOM image transfer | Functional | (20) | |
| N2 Imaging import (C‐arm / CT / MRI) | Functional | (20,5) | |
| N3 CT/MR imaging geometry consistency |
| ||
| N4 CT import check | Functional | (20,5) | |
| N5 Digitizer test |
| ||
| N6 Creating a plan with CT or MRI | Functional | (20,5) | |
| N7 Applicator library template: accuracy test | Functional | (5) | |
| N8 Reference point entry test | Functional | (20) | |
| N9 Printer test | Functional | (20) | |
| N10 Genetic‐text print format (clinical print & DVH) | Functional | ||
| N11 Print setup: PDF | Functional | ||
| N12 Plan in TPS: unlock and lock check | Functional | ||
| N13 User login setup | Functional | ||
| N14 Current date & time check | Functional | ||
| N15 T&O and VC plan template check | Functional | ||
| N16 T&O and VC optimization template check | |||
|
| (20) | ||
| E1 VC | |||
| E2 CT based T&O planning and delivery | (5) | ||
| E3 MRI based T&O planning and delivery | (5) |