Abstract
Human immunodeficiency virus type 1 (HIV-1) p24 protein is the most abundant viral protein of HIV-1. This protein is secreted in blood serum at high levels during the early stages of HIV-1 infection, making it a biomarker for early diagnosis. In this study, a colloidal gold immunochromatographic assay (GICA) was established for detecting p24 protein using mouse monoclonal antibodies (mAbs). The HIV-1 p24 protein was expressed in E. coli strain BL21 and the purified protein was used to immunize mice. Stable hybridoma cell lines secreting anti-p24 monoclonal antibodies were obtained after ELISA screening and subcloning by limiting dilution. 34 different capture and labeling mAb pairs were selected by a novel antibody-capture indirect sandwich ELISA and then applied in GICA to detect p24 protein. The GICA method has a limit of detection (LOD) of 25 pg/mL and could detect p24 protein in all 10 positive samples obtained from the National Reference of HIV-1 p24 antigen. Out of 153 negative samples tested, 3 false positives results were obtained. The overall specificity of this test was 98.03%. The good sensitivity and specificity of this method make it a suitable alternative to provide a more convenient and efficient tool for early diagnosis of HIV infection.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27069923 PMCID: PMC4812187 DOI: 10.1155/2016/6743904
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Preparation and identification of recombinant p24. (a) Expression and purification of recombinant p24. 1: whole cell lysate without IPTG induction, 2: whole cell lysate with IPTG induction, and 3: flow through solution; lanes 4 to 7 are four different solutions eluted by 50 mM, 150 mM, 300 mM, and 500 mM imidazole buffer, respectively. (b) Indirect ELISA analysis of the recombinant p24. A: antip24 antibody as the primary antibody, B: blank without primary antibody, C: blank without coating antigen, and D: blank without primary antibody and coating antigen.
Figure 2Sensitivity of GICA to r-p24 antigen. GICA test strips were treated with different concentrations (1000 pg/mL, 200 pg/mL, 100 pg/mL, 50 pg/mL, 25 pg/mL, and 10 pg/mL) of r-p24 antigen. T and C represent the test and control lines, respectively.
The detection results of National Reference for HIV-1 p24 antigen by GICA strips.
| Sample | Derivation of references | Genotype | Detection level |
|---|---|---|---|
| P1 | P24 positive Sample 34# | AE | Faint |
| P2 | P24 positive Sample 106# | B | Faint |
| P3 | P24 positive Sample 116# | B | Medium |
| P4 | P24 positive Sample 119# | B | Weak |
| P5 | 500 fold-diluted laboratory-grown virus sample A | Not given | Medium |
| P6 | 100 fold-diluted laboratory-grown virus sample A | Not given | Weak |
| P7 | 1000 fold-diluted laboratory-grown virus sample L | B | Faint |
| P8 | 10000 fold-diluted laboratory-grown virus sample L | B | Faint |
| P9 | 50 fold-diluted HIV-1 window period Sample P | AE | Faint |
| P10 | WHO Standard p24 25 U/mL | B | Medium |
| L1 | WHO Standard p24 20 U/mL | B | Medium |
| L2 | WHO Standard p24 10 U/mL | B | Weak |
| L3 | WHO Standard p24 5 U/mL | B | Faint |
| L4 | WHO Standard p24 2.5 U/mL | B | Negative |
| L10 | Buffer for L Series | B | Negative |
Detection level: Band intensity was read independently by two individuals.
Comparison of GICA with some published HIV-1 p24 assays in recent years.
| Assay | Support | p24 detection limit | Execution time |
|---|---|---|---|
| Colloidal gold immunochromatographic assay (GICA) Commercial assays | Lateral flow test strip | 25 pg/mL | 20 min |
| Abbott ARCHITECT HIV Ag/Ab Combo (Chemiluminescentmicroparticle immunoassay) | ARCHITECT automated analyzer | 18.4 pg/mL (SFTS standard) | 26 min [ |
| Roche Elecsys® HIV combi (Chemiluminescentmicroparticle immunoassay) | Modular automated analyzer | 0.9 IU/mL (WHO standard) | 18 min [ |
| Biomérieux VIDAS HIV DUO Ultra (Enzyme linked fluorescence assay) | VIDAS automated analyzer | 11.5 pg/mL (SFTS standard) | 120 min [ |
| Bio-Rad GS HIV Combo Ag/Ab (Enzyme linked immunosorbent assay) | Bio-Rad EVOLIS | 12.7 pg/mL (SFTS standard) | >120 min [ |
| Bio-Rad Genscreen | Microplate | 13 pg/mL (SFTS standard) | 120 min [ |
| Laboratory assays | |||
| Ultrasensitive capacitive immunosensor assay | Capacitive immunosensor system | 7.9 × 10−8 pg/mL | 20 min [ |
| Amperometricimmunosensor assay based on direct gold electroplating-modified electrode | Amperometricimmunosensor system | 8 pg/mL | 20 min [ |
| Amperometricimmunosensor assay based on acetone-extracted propolis | Amperometricimmunosensor system | 6.4 pg/mL | 10 min [ |
| Boosted ELISA based on immune complex dissociation and amplified signal | Microplate | 0.5 pg/mL | >120 min [ |
| Nanoparticle-based biobarcode amplification assay | NanosphereVerigene ID Reader | 0.1 pg/mL | >120 min [ |
| Magnetic immuno-chromatography assay | MagnaBiosciences magnetic assay reader | 30 pg/mL | 40 min [ |
SFTS standard: French national reference of HIV-1 p24 Ag obtained from the French Society of Blood Transfusion.