Are ligament-tensioning devices interchangeable? A study of femoral rotation.

F Basselot1, T Gicquel2, H Common2, A Hervé2, E Berton3, M Ropars4, D Huten2.   

Abstract

BACKGROUND: During total knee arthroplasty (TKA), femoral rotation can be adjusted either in relation to bony landmarks or by tensioning the ligaments with the knee in 90° of flexion. The primary objective of this study was to compare femoral rotations achieved using various ligament-tensioning devices. The secondary objective was to compare these femoral rotations to that indicated by the transepicondylar axis (TEA).
MATERIAL AND METHODS: We performed 13 posterior-stabilised TKA procedures using HiFit (Ceraver(®)) on cadaver knees. Before performing the posterior condyle cut, we used an original method to measure the femoral rotation induced by five different ligament-tensioning devices (2 with a ratchet mechanism, 1 with screws, 1 force-sensing device, and 1 with spacer blocks) and the central tibio-femoral distance (CTFD).
RESULTS: Both ratchet tensioners provided significantly greater mean external rotation values (P=0.002), of 4.94° and 4.46°, respectively, compared to the force-sensing and spacer tensioners. Significant differences were found across devices for CTFD, with a mean difference of about 2mm between the ratchet and screw tensioners versus the force-sensing and spacer tensioners. The mean differences in rotations obtained using the tensioners versus the TEA were close to 0° but with standard deviations greater than 4°.
CONCLUSION: Femoral rotation was dependent on the distraction force applied to the joint. Tensioners that did not measure the distraction force were associated with greater distraction force and external rotation values. The TEA criterion did not reliably indicate good ligament balance. LEVEL OF EVIDENCE: Experimental study.
Copyright © 2016. Published by Elsevier Masson SAS.

Entities:  

Keywords:  Femoral rotation; Gap balancing; Ligament-tensioning device; Total knee arthroplasty

Mesh:

Year:  2016        PMID: 27055932     DOI: 10.1016/j.otsr.2016.02.001

Source DB:  PubMed          Journal:  Orthop Traumatol Surg Res        ISSN: 1877-0568            Impact factor:   2.256


  3 in total

Review 1.  Robotic technology in total knee arthroplasty: a systematic review.

Authors:  Babar Kayani; Sujith Konan; Atif Ayuob; Elliot Onochie; Talal Al-Jabri; Fares S Haddad
Journal:  EFORT Open Rev       Date:  2019-10-01

2.  Functional stability: an experimental knee joint cadaveric study on collateral ligaments tension.

Authors:  Bernardo Innocenti; Edoardo Bori; Thomas Paszicsnyek
Journal:  Arch Orthop Trauma Surg       Date:  2021-05-27       Impact factor: 3.067

3.  Robotic total knee arthroplasty: clinical outcomes and directions for future research.

Authors:  B Kayani; F S Haddad
Journal:  Bone Joint Res       Date:  2019-11-02       Impact factor: 5.853

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.