A novel suture anchor constructed of cortical bone for rotator cuff repair: a biomechanical study on sheep humerus specimens.

Qi Guo1,2, Chunbao Li3, Wei Qi3, Hongliang Li3, Xi Lu3, Xuezhen Shen3, Feng Qu3, Yujie Liu4.   

Abstract

PURPOSE: The aim of this study was to evaluate biomechanical properties of a new type of suture anchors constructed of human cortical allograft bone and compare it with the similar standard titanium screw anchor for rotator cuff tears in sheep humerus model.
METHODS: Twenty-four paired sheep humeri were harvested from 12 male sheep aged 18 months. Specimens were divided into cortical bone anchor group and titanium screw anchor group. The anchors loaded with two sutures were placed at the footprint of infraspinatus tendon. Cyclic loading test was performed from 10 to 60 N at 1 Hz for 500 cycles and followed by a load-to-failure test at 33 mm/sec. A paired t-test was used to compare the biomechanical properties of the anchors of each type.
RESULTS: No anchors failed during the cyclic phase, and the cortical bone anchors were all pulled out intact. The cyclic displacement of the cortical bone anchor was not significantly greater than that of the titanium screw anchor (P > 0.05). Student's t test showed no statistically significant difference between anchors in terms of failure load (cortical bone anchor: 304.74 ± 64.46 N versus titanium screw anchor: 328.45 ± 89.58 N; P = 0.213), ultimate load (cortical bone anchor: 325.82 ± 76.45 N versus titanium screw anchor: 345.61 ± 83.56 N; P = 0.183), yield load (cortical bone anchor: 273.78 ± 44.75 N versus titanium screw anchor: 284.72 ± 56.37 N; P = 0.326) or stiffness (cortical bone anchor: 52.97 ± 14.28 N/mm versus titanium screw anchor: 62.38 ± 18.35 N/mm; P = 0.112).
CONCLUSIONS: In vitro, this experimental study suggested no statistically significant difference in initial fixation stability between the new type anchor and titanium screw anchor at a chosen level of significance (P < 0.05). The new type of suture anchor constructed of cortical bone provides comparable initial fixation strength to a similar metallic anchor for rotator cuff repair.

Entities:  

Keywords:  Biomechanical testing; Cortical allograft bone; Rotator cuff repair; Sheep; Suture anchor

Mesh:

Year:  2016        PMID: 27055449     DOI: 10.1007/s00264-016-3185-4

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  31 in total

Review 1.  The evolution of suture anchors in arthroscopic rotator cuff repair.

Authors:  Patrick J Denard; Stephen S Burkhart
Journal:  Arthroscopy       Date:  2013-07-19       Impact factor: 4.772

2.  Biomechanical testing of a new knotless suture anchor compared with established anchors for rotator cuff repair.

Authors:  Matthias F Pietschmann; Valerie Froehlich; Andreas Ficklscherer; Bernd Wegener; Volkmar Jansson; Peter E Müller
Journal:  J Shoulder Elbow Surg       Date:  2008-04-08       Impact factor: 3.019

3.  Coating cortical bone allografts with periosteum-mimetic scaffolds made of chitosan, trimethyl chitosan, and heparin.

Authors:  Raimundo Romero; Laura Chubb; John K Travers; Timothy R Gonzales; Nicole P Ehrhart; Matt J Kipper
Journal:  Carbohydr Polym       Date:  2015-01-14       Impact factor: 9.381

4.  Reprint of: The deadman theory of suture anchors: observations along a south Texas fence line.

Authors:  Stephen S Burkhart
Journal:  Arthroscopy       Date:  2014-08       Impact factor: 4.772

5.  Biological effects of rAAV-caAlk2 coating on structural allograft healing.

Authors:  Mette Koefoed; Hiromu Ito; Kirill Gromov; David G Reynolds; Hani A Awad; Paul T Rubery; Michael Ulrich-Vinther; Kjeld Soballe; Robert E Guldberg; Angela S P Lin; Regis J O'Keefe; Xinping Zhang; Edward M Schwarz
Journal:  Mol Ther       Date:  2005-08       Impact factor: 11.454

6.  The development of suture anchors for use in soft tissue fixation to bone.

Authors:  E M Goble; W K Somers; R Clark; R E Olsen
Journal:  Am J Sports Med       Date:  1994 Mar-Apr       Impact factor: 6.202

7.  Pullout strength of all suture anchors in the repair of rotator cuff tears: a biomechanical study.

Authors:  Alexandre Galland; Stéphane Airaudi; Renaud Gravier; Sophie Le Cann; Patrick Chabrand; Jean-Noël Argenson
Journal:  Int Orthop       Date:  2013-07-09       Impact factor: 3.075

Review 8.  Arthroscopic Repair for Chronic Massive Rotator Cuff Tears: A Systematic Review.

Authors:  Patrick Henry; David Wasserstein; Sam Park; Tim Dwyer; Jaskarndip Chahal; Gerard Slobogean; Emil Schemitsch
Journal:  Arthroscopy       Date:  2015-09-11       Impact factor: 4.772

9.  Pullout strength of suture anchors in comparison with transosseous sutures for rotator cuff repair.

Authors:  Matthias F Pietschmann; Valerie Fröhlich; Andreas Ficklscherer; Jörg Hausdorf; Sandra Utzschneider; Volkmar Jansson; Peter E Müller
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2008-05       Impact factor: 4.342

10.  Experiences with sheep as an animal model for shoulder surgery: strengths and shortcomings.

Authors:  A Simon Turner
Journal:  J Shoulder Elbow Surg       Date:  2007-05-15       Impact factor: 3.019

View more
  1 in total

1.  Rotator cuff repair with biodegradable high-purity magnesium suture anchor in sheep model.

Authors:  Yudie Chen; Yu Sun; Xinhui Wu; Jie Lou; Xiaonong Zhang; Zhaoxiang Peng
Journal:  J Orthop Translat       Date:  2022-09-10       Impact factor: 4.889