Optimization of the filter parameters in (99m)Tc myocardial perfusion SPECT studies: the formulation of flowchart.

Takayuki Shibutani1,2, Masahisa Onoguchi3, Tomoki Yamada4, Hiroki Kamida5, Kohei Kunishita6, Yuuki Hayashi7, Tadashi Nakajima4, Seigo Kinuya2.   

Abstract

Myocardial perfusion single photon emission computed tomography (SPECT) is typically subject to a variation in image quality due to the use of different acquisition protocols, image reconstruction parameters and image display settings by each institution. One of the principal image reconstruction parameters is the Butterworth filter cut-off frequency, a parameter strongly affecting the quality of myocardial images. The objective of this study was to formulate a flowchart for the determination of the optimal parameters of the Butterworth filter for filtered back projection (FBP), ordered subset expectation maximization (OSEM) and collimator-detector response compensation OSEM (CDR-OSEM) methods using the evaluation system of the myocardial image based on technical grounds phantom. SPECT studies were acquired for seven simulated defects where the average counts of the normal myocardial components of 45° left anterior oblique projections were approximately 10-120 counts/pixel. These SPECT images were then reconstructed by FBP, OSEM and CDR-OSEM methods. Visual and quantitative assessment of short axis images were performed for the defect and normal parts. Finally, we formulated a flowchart indicating the optimal image processing procedure for SPECT images. Correlation between normal myocardial counts and the optimal cut-off frequency could be represented as a regression expression, which had high or medium coefficient of determination. We formulated the flowchart in order to optimize the image reconstruction parameters based on a comprehensive assessment, which enabled us to perform objectively processing. Furthermore, the usefulness of image reconstruction using the flowchart was demonstrated by a clinical case.

Entities:  

Keywords:  Butterworth filter; Cut-off frequency; Flowchart; Myocardial perfusion SPECT; Optimization; Standardization

Mesh:

Substances:

Year:  2016        PMID: 27052439     DOI: 10.1007/s13246-016-0433-2

Source DB:  PubMed          Journal:  Australas Phys Eng Sci Med        ISSN: 0158-9938            Impact factor:   1.430


  1 in total

1.  Optimization of Ordered Subset Expectation Maximization Parameters for Image Reconstruction in Tc-99m Methoxyisobutylisonitrile Myocardial Perfusion SPECT and Comparison with Corresponding Filtered Back Projection-Reconstructed Images.

Authors:  Pankaj Dheer; Priyanka Gupta; Sameer Kamalakar Taywade; Averilicia Passah; Anil Kumar Pandey; Chetan Patel
Journal:  Indian J Nucl Med       Date:  2021-03-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.