| Literature DB >> 27047061 |
Pranoti Sharma1, C K Singh1, Deepti Narang2.
Abstract
AIM: Detection of rabies is a cause of serious concern in developing countries, where dearth of highly equipped laboratories and trained personnel to handle sophisticated investigations is felt. The availability of a diagnostic kit, which can be used in the field, is essential for diagnosis and control programs as well as for epidemiological surveillance of the prevalence of the disease. This study was planned to evaluate anigen rabies Ag test kit for its efficacy to be used for rapid diagnosis of rabies under field conditions. The test results were compared with hemi-nested reverse transcriptase polymerase chain reaction and with a gold standard fluorescent antibody test.Entities:
Keywords: antigen; heminested; immunochromatographic; rabies
Year: 2015 PMID: 27047061 PMCID: PMC4774692 DOI: 10.14202/vetworld.2015.135-138
Source DB: PubMed Journal: Vet World ISSN: 0972-8988
Figure-1Detection limit immunochromatographic diagnostic test, (a) 1:100 dilution, strong positive bands were observed both in the test line and the control line, (b) 1:1000 dilution, weak positive band was observed in the test line and strong band in control line, (c) 1:10,000 dilution, very weak positive band was observed in test line and strong band in control line, (d) only control band appeared, negative result.
Test results of FAT and immunochromatographic test kit.
| Test | Brain FAT positive | Brain FAT negative | Total |
|---|---|---|---|
| Number of animals positive by ICT kit | 22 | 0 | 22 |
| Number of animals negative by ICT kit | 2 | 10 | 12 |
| Total | 24 | 10 | 34 |
Sensitivity=22/22+2×100, Specificity=10/0+10, Accuracy=22+10/22+0 + 10+2, FAT=Fluorescent antibody test, ICT=Immunochromatographic diagnostic test
Test results of FAT and HnRT-PCR.
| Test | Brain FAT positive | Brain FAT negative | Total |
|---|---|---|---|
| Number of animals positive HnRT-PCR | 24 | 0 | 24 |
| Number of animals negative HnRT-PCR | 0 | 10 | 10 |
| Total | 24 | 10 | 34 |
Sensitivity=24/24+0×100, Specificity=10/0+10, Accuracy=24+10/24+0+10+0, FAT=Fluorescent antibody test, HnRT-PCR=Hemi-nested reverse transcriptase polymerase chain reaction