The Pullout Strength of Pedicle Screws Following Redirection After Lateral Wall Breach or End-plate Breach.

Yuichiro Goda1, Kosaku Higashino1, Shunichi Toki1, Daisuke Suzuki2, Takuma Kobayashi3, Tetsuya Matsuura1, Mineko Fujimiya2, Williams C Hutton4, Yoshihiro Fukui5, Koichi Sairyo1.   

Abstract

STUDY
DESIGN: Using fresh cadavers, the biomechanical testing were used to examine the pullout strength of each pedicle screw.
OBJECTIVE: The aim of this study was to evaluate pullout strength of (1) a redirected pedicle screw following lateral wall breach; (2) a redirected pedicle screw following end-plate breach; and (3) a pedicle without redirection after end-plate breach without redirection. SUMMARY OF BACKGROUND DATA: Screw malposition, such as lateral wall breach or end-plate breach, is one of the main pitfalls of inserting pedicle screws.
METHODS: From 17 fresh spines, 54 vertebrae were harvested. In each vertebra on one pedicle, the screw was inserted correctly down the axis of the pedicle, while on the other pedicle, the screw was inserted to breach the lateral wall or the end-plate. The 18 pedicle screws that breached the lateral wall were then removed and redirected along the correct axis of the pedicle. The 18 pedicle screws that breached the end-plate were removed and redirected along the correct axis of the pedicle. The 18 other pedicle screws that had breached the end-plate were not removed. The pullout force of pedicle screws was measured.
RESULTS: First, the mean pullout strength for the redirected screws following lateral wall breach was 24.0% less as compared with the correctly aligned screws. Second, the mean pullout strength for the redirected screws following end-plate breach was 23.3% less as compared with the correctly aligned screws. Third, the mean pullout strength for the pedicle screws end-plate breach was 7.6% less as compared with the correctly aligned screws.
CONCLUSION: The pullout strength of redirected pedicle screws after either a lateral pedicle breach or end-plate breach is significantly less than the pullout strength of correctly aligned screw. A pedicle screw that is not redirected after end-plate breach is weaker than a pedicle screw correctly aligned; however, the difference is not significant. LEVEL OF EVIDENCE: N/A.

Entities:  

Mesh:

Year:  2016        PMID: 27046637     DOI: 10.1097/BRS.0000000000001600

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  4 in total

1.  Biomechanical properties of posterior transpedicular-transdiscal oblique lumbar screw fixation with novel trapezoidal lateral interbody spacer: an in vitro human cadaveric model.

Authors:  Ai-Min Wu; Jonathan A Harris; John C Hao; Sean M Jenkins; Yong-Long Chi; Brandon S Bucklen
Journal:  Eur Spine J       Date:  2017-04-06       Impact factor: 3.134

2.  Comparison of accuracy and safety between robot-assisted and conventional fluoroscope assisted placement of pedicle screws in thoracolumbar spine: A meta-analysis.

Authors:  Chuntao Li; Wenyi Li; Shangju Gao; Can Cao; Changren Li; Liang He; Xu Ma; Meng Li
Journal:  Medicine (Baltimore)       Date:  2021-09-24       Impact factor: 1.817

3.  Micro-computed tomography analysis of the lumbar pedicle wall.

Authors:  Tomoyo Y Irie; Tohru Irie; Alejandro A Espinoza Orías; Kazuyuki Segami; Norimasa Iwasaki; Howard S An; Nozomu Inoue
Journal:  PLoS One       Date:  2021-07-08       Impact factor: 3.240

4.  Can Postoperative CT Imaging in Spine Surgery Be Replaced by Intraoperative 3D Rotation With the C-Arm?: Results of a Prospective Single Center Cohort Study.

Authors:  Mohammed Banat; Johannes Wach; Abdallah Salemdawod; Lisa Domurath; Jasmin Scorzin; Hartmut Vatter
Journal:  Front Surg       Date:  2021-07-14
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.