Renal Stent Crushed During Open Aneurysmorrhaphy for Endoleak After Fenestrated EVAR.

Islah Ud Din1, Usman Shaikh2, Robert K Fisher3, Richard G McWilliams2.   

Abstract

PURPOSE: To report a complication of renal stent crushing during open aneurysmorrhaphy performed 10 years after fenestrated endovascular aneurysm repair (FEVAR). CASE REPORT: A 67-year-old male patient underwent elective FEVAR of a juxtarenal aortic aneurysm. Uncovered balloon-expandable stents were placed through the fenestrations for the superior mesenteric and right renal arteries; the left renal artery received a Jostent covered balloon-expandable stent. The uncovered right renal stent was noted to be crushed on the first imaging after FEVAR, but was left untreated; the right kidney was thereafter significantly smaller than the left. Over several years, the aneurysm expanded, and type II endoleak involving the lumbar arteries was embolized with coils 7 years after FEVAR. Despite this, the aneurysm continued to enlarge and reached a diameter of 12.8 cm. Open aneurysmorrhaphy and selective ligation of lumbar arteries was performed 10 years after FEVAR. The first surveillance imaging after aneurysmorrhaphy demonstrated a new finding of a crushed left renal stent, which was presumed to be related to surgical instrumentation. The stent was successfully redilated percutaneously, and renal function remained stable. Computed tomography demonstrated a normal appearance of the left renal covered stent after re-expansion. Doppler ultrasound after 7.5 months showed normal renal perfusion.
CONCLUSION: Balloon-expandable visceral artery stents are susceptible to crush injury during aortic surgery. Consideration should be given to early imaging after such surgery in FEVAR patients.
© The Author(s) 2016.

Entities:  

Keywords:  aneurysmorrhaphy; balloon-expandable stent; complication; crushed stent; endovascular aneurysm repair; fenestrated stent-graft; juxtarenal aortic aneurysm; stent deformation

Mesh:

Year:  2016        PMID: 27044269     DOI: 10.1177/1526602816642199

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  1 in total

1.  Self-expandable Graft Stenting in an Iatrogenic Fistula between Common Carotid Artery and Internal Jugular Vein.

Authors:  Sung Chul Jin; Dong Hyuk Lee; Chae Wook Huh
Journal:  J Cerebrovasc Endovasc Neurosurg       Date:  2017-09-30