Literature DB >> 27042021

The outcomes of intra-aortic balloon pump usage in patients with acute myocardial infarction: a comprehensive meta-analysis of 33 clinical trials and 18,889 patients.

Zhong-Guo Fan1, Xiao-Fei Gao2, Li-Wen Chen1, Xiao-Bo Li2, Ming-Xue Shao2, Qian Ji1, Hao Zhu1, Yi-Zhi Ren1, Shao-Liang Chen2, Nai-Liang Tian2.   

Abstract

BACKGROUND: The effects of intra-aortic balloon pump (IABP) usage in patients with acute myocardial infarction remain controversial. This study sought to evaluate the outcomes of IABP usage in these patients.
METHODS: Medline, EMBASE, and other internet sources were searched for relevant clinical trials. The primary efficacy endpoints (in-hospital, midterm, and long-term mortality) and secondary endpoints (reinfarction, recurrent ischemia, and new heart failure in the hospital) as well as safety endpoints (severe bleeding requiring blood transfusion and stroke in-hospital) were subsequently analyzed.
RESULTS: Thirty-three clinical trials involving 18,889 patients were identified. The risk of long-term mortality in patients suffering from acute myocardial infarction was significantly decreased following IABP use (odds ratio [OR] 0.66, 95% confidence interval [CI]: 0.48-0.91, P=0.010). Both in-hospital and midterm mortality did not differ significantly between the IABP use group and no IABP use group (in-hospital: OR 0.87, 95% CI: 0.59-1.28, P=0.479; midterm: OR 1.12, 95% CI: 0.53-2.38, P=0.768). IABP insertion was not associated with the risk reduction of reinfarction, recurrent ischemia, or new heart failure. However, IABP use increased the risk of severe bleeding requiring blood transfusion (OR 2.05, 95% CI: 1.29-3.25, P=0.002) and stroke (OR 1.71, 95% CI: 1.04-2.82, P=0.035). In the thrombolytic therapy and cardiogenic shock subgroups, reduced mortality rates following IABP use were observed.
CONCLUSION: IABP insertion is associated with feasible benefits with respect to long-term survival rates in patients suffering from acute myocardial infarction, particularly those suffering from cardiogenic shock and receiving thrombolytic therapy, but at the cost of higher incidence of severe bleeding and stroke.

Entities:  

Keywords:  acute myocardial infarction; cardiogenic shock; intra-aortic balloon pump; meta-analysis; thrombolytic therapy

Year:  2016        PMID: 27042021      PMCID: PMC4801154          DOI: 10.2147/PPA.S101945

Source DB:  PubMed          Journal:  Patient Prefer Adherence        ISSN: 1177-889X            Impact factor:   2.711


Introduction

Patients suffering from acute myocardial infarction (AMI) are at an increased risk for high mortality, particularly in the setting of AMI complicated by cardiogenic shock (CS), although both emergency revascularization (ERV) and thrombolysis have been widely used.1 The intra-aortic balloon pump (IABP) has been widely used since it was first used clinically in 1968. It improves diastolic coronary blood flow and reduces both afterload and myocardial oxygen demand,2 changes thought to have positive effects on myocardial recovery following AMI.3,4 According to the American College of Cardiology and American Heart Association (ACC/AHA 2012) guidelines, IABP was recommended for CS and its insertion was suggested at the completion of coronary angiography and revascularization (IIa).5 The European Society of Cardiology guidelines also recommended IABP as a bridge to reperfusion for patients suffering from CS.6 In recent years, several large randomized controlled trials (RCTs)7–10 and meta-analyses11,12 have demonstrated only limited or even no benefits with respect to midterm and long-term all-cause mortality in patients with AMI complicated by CS for whom early revascularization was planned. However, two other meta-analyses13,14 have demonstrated that IABP has a positive impact with respect to all-cause mortality in these patients for whom thrombolysis was used as a preferred reperfusion strategy. These conflicting data challenged the recommendations of the current guidelines and these aforementioned meta-analyses did not include all the available relevant clinical trials. Therefore, we sought to conduct an updated, comprehensive meta-analysis involving as many clinical trials as possible to evaluate the evidence pertaining to the performance of IABP performed as an adjunct therapy in patients suffering from an AMI complicated with CS or not.

Methods

Literature search

We searched Medline, EMBASE, and the Cochrane Controlled Trials Registry from their dates of inception until May 2015 for clinical trials comparing outcomes following IABP use with outcomes in the absence of IABP use (defined as the Control group) following AMI. To be certain all relevant studies were included, the electronic databases were searched using combinations of several relevant keywords, including intra-aortic balloon pump, counterpulsation, acute myocardial infarction, and clinical trials. All potentially relevant articles and references from published reviews and meta-analyses were subsequently screened for eligibility.

Inclusion and exclusion criteria

The articles were required to meet the following criteria: 1) adult patients suffering from AMI (age from 18 to 90 years), regardless of CS and 2) clinical trials comparing an IABP group and a control group. The exclusion criteria were as follows: 1) nonhuman or ongoing studies; 2) non-English language studies; 3) reviews or meta-analyses; 4) duplicated studies or different studies using the sample; and 5) patients supported by other cardiac assist devices.

Data extraction, synthesis, and quality assessment

Two investigators (FZG and GXF) independently reviewed all relevant articles to assess their eligibility, using standardized data-abstraction forms. Disagreements were resolved by a third investigator (CLW). We extracted the following data from each included study: the name of the trial or first author, publication year, inclusion and exclusion criteria, baseline demographics, and clinical outcomes at the hospital and/or during follow-up. All included studies were divided into two subgroups based on the design of each study, described as RCTs and observational clinical trials (Obs). To account for the effects of treatment, we also grouped the studies by type of reperfusion, as follows: no reperfusion, ERV alone, including either percutaneous transcoronary angioplasty or coronary artery bypass grafting, thrombolytic therapy alone (TT alone), and ERV plus TT. However, the patients with CS were selected as another subgroup to determine whether they responded to IABP insertion. Moreover, the patients with AMI were divided into two subgroups based on whether they underwent IABP insertion before or after reperfusion to determine the best opportunity for IABP insertion. The quality of all retrieved studies was assessed to ensure minimization of bias, but no formal scoring system was used.

Study endpoints

The primary efficacy endpoint was the rate of all-cause mortality, which was evaluated across three periods based on the follow-up durations of the included trials, including in-hospital mortality, midterm mortality (defined as mortality between 30 days and 2 months), and long-term mortality (defined as mortality after 6 months), and the secondary efficacy endpoints were new heart failure, reinfarction, and recurrent ischemia during hospitalization. The incidences of stroke and severe bleeding requiring blood transfusion in-hospital were evaluated as safety endpoints. The rates of all-cause mortality and new heart failure could be replaced by cardiac death15 and pulmonary edema,16 respectively, if the included articles did not have the relevant data. The definitions of the clinical endpoints varied slightly among the included trials, and we evaluated the clinical endpoints using standardized definitions.

Statistical analysis

All endpoints were treated as dichotomous variables, which were analyzed using odds ratio (OR) with 95% confidence interval (CI). Statistical heterogeneity among the included studies was measured using the Cochrane’s Q test and the I2 statistic. When the P-value of Q test was <0.10 and/or the I2 was ≥50%, significant heterogeneity was considered and a random-effects model was selected. If not, the fixed-effects model and the Mantel–Haenszel method were used. Publication bias was examined via Egger’s test (P<0.1 for significant asymmetry).17 To assess the stability of the overall treatment effects, sensitivity analyses (exclude one study at a time) were performed. All P-values were two-tailed, and a P-value <0.05 was considered statistically significant. This meta-analysis was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analysis statements,18 and the data were analyzed using STATA 12.0 (StataCorp LP, College Station, TX, USA).

Results

Eligible studies and patient characteristics

After screening 1,841 initial articles using the electronic databases, 33 clinical trials were identified, including 15 RCTs (2,497 patients)7–10,15,19–30 and 18 Obs (16,392 patients)16,31–49 (Figure 1). The baseline characteristics of the included RCTs and Obs are summarized in Tables 1–4.
Figure 1

A flowchart depicting the selection of the studies included in this meta-analysis.

Table 1

The characteristics of randomized controlled trials pertaining to the use of an intra-aortic balloon pump in the setting of AMI

StudyPublication yearType of reperfusion for AMIInclusion criteriaExcluded CSFollow-up duration (d)
Waksman et al201993ThrombolysisAMI with CSNo360
TACTICS192005Thrombolysis, RVAMI with hypotension or severe HFNo30, 180
Li et al272007RVAMI with CSNo360
SHOCK Trial282010ERVAMI with CSNoIn-hospital
IABP-SHOCK II7,82013ERVAMI with CSNo30, 360
O’Rourke et al221981No reperfusionAMI with acute HFNA450
Flaherty et al151985No reperfusionAMI without CSYes14, 60
Ohman et al211994ERVAMI without CSYesIn-hospital
Kono et al231996Failed thrombolysisAMI without CSYes30
PAMI-II241997ERVAMI without CSYesIn-hospital
Vijayalakshmi et al262007ERVAMI without CSYes30
Gu et al292011ERVAMI without CSYes30, 180
CRISP AMI302011ERVAMI without CSYes30, 180
BCIS-1102013ERVAMI without CSYes1,530
Van’t Hof et al251999ERVAMI with ST↑/↓>20 mmNA180

Abbreviations: AMI, acute myocardial infarction; CS, cardiogenic shock; d, days; ERV, emergency revascularization; HF, heart failure; NA, not available; RV, revascularization, including percutaneous transcoronary angioplasty and coronary artery bypass grafting; ST, ST-segments.

Table 2

The characteristics of observational trials pertaining to the use of an intra-aortic balloon pump in the setting of AMI

StudyPublication yearType of reperfusion for AMIInclusion criteriaExcluded CSFollow-up duration (d)
Moulopoulos et al311986No reperfusionAMI with CSNoIn-hospital
Bengtson et al321992NAAMI with CSNoIn-hospital
Stomel et al331994Thrombolysis, ERVAMI with CSNoIn-hospital
Kovack et al351997ThrombolysisAMI with CSNo30, 360
GUSTO-I341997Thrombolysis, RVAMI with CSNo30, 360
SHOCK Trial Registry382000Thrombolysis, ERVAMI with suspected CSNoIn-hospital
NRMI-2392001Thrombolysis, ERVAMI with CSNoIn-hospital
French et al402003Thrombolysis, ERVAMI with CSNo360
Gu et al412010ERVAMI with CSNo30, 180
Stub et al432011RVACS with CSNo30
EHS PCI Registry422011ERVAMI with CSNoIn-hospital
AMC CS492012ERVAMI with CSNo30
ALKK-PCI Registry452013ERVAMI with CSNoIn-hospital
Dziewierz et al462014ERVAMI with CSNo30, 360
Ohman et al161991ThrombolysisAMI without CSYesIn-hospital
Mahmoudi et al442012ERVAMI without CSYesIn-hospital
Brodie et al361999ERVAMI without prior TTNA30
Kumbasar et al371999ThrombolysisAMI ≤6 hNAIn-hospital

Abbreviations: AMI, acute myocardial infarction; CS, cardiogenic shock; d, days; ERV, emergency revascularization; h, hours; NA, not available; RV, revascularization, including percutaneous transcoronary angioplasty and coronary artery bypass grafting; TT, thrombolytic therapy.

Table 3

The baseline characteristics of randomized controlled trials pertaining to the use of an IABP in the setting of AMI

StudyPatients, N (IABP/control)Mean age, years (IABP/control)Male, n (IABP/control)Hypertension, n (IABP/control)Diabetes, n (IABP/control)Prior MI, n (IABP/control)
Waksman et al2024/2166.8/67.814/1514/10NA/NA15/11
TACTICS1930/2768*/67*23/20NA/NA9/312/5
Li et al2720/1967.4/64.912/12NA/NA9/8NA/NA
SHOCK Trial2819/2162.1/66.114/178/1010/104/5
IABP-SHOCK II8299/29670*/69*202/211213/199105/9071/61
O’Rourke et al2214/1660/5412/12NA/NANA/NA2/5
Flaherty et al1510/1052/539/8NA/NANA/NA4/4
Ohman et al2196/8656/5571/6547/3716/1519/20
Kono et al2323/2254/6020/1610/136/6NA/NA
PAMI-II24211/22664.7/63.7158/170116/12645/3345/49
Vijayalakshmi et al2617/1657.5/5914/146/63/42/4
Gu et al2951/5567.4/66.629/3635/3318/192/3
CRISP AMI30156/17356.1*/57.7*132/14439/6027/360/0
BCIS-19,10151/15071/71122/11795/9156/50113/108
Van’t Hof et al25118/12059/5699/101NA/NA12/917/16

Note:

Median.

Abbreviations: AMI, acute myocardial infarction; IABP, intra-aortic balloon pump; MI, myocardial infarction; NA, not available.

Table 4

The baseline characteristics of observational clinical trials pertaining to the use of an IABP in the setting of AMI

StudyPatients, N (IABP/control)Mean age, years (IABP/control)Male, n (IABP/control)Hypertension, n (IABP/control)Diabetes, n (IABP/control)Prior MI, n (IABP/control)
Moulopoulos et al3135/1460.3/61.129/13NA/NANA/NA6/16
Bengtson et al3299/10164/67NA/NANA/NANA/NANA/NA
Stomel et al3351/1366/6623/831/611/316/2
Kovack et al3527/1962/6416/1210/117/56/1
GUSTO-I3462/24864*/68*42/15423/9914/5719/67
Shock Registry38439/41765.2/73.4294/250203/233137/141154/188
NRMI-2394,215/4,456~67/74.1~60.5%/50.5%~47%/49%~29%/32%~26%/30%
French et al40260/4165.3/67.4175/30124/1783/1084/14
Gu et al4143/4870.4/67.927/3130/3217/17NA/NA
Stub et al43251/15965.7/67.7189/110142/9971/3954/38
EHS PCI Registry42162/49165.3/65.4110/33595/31644/13858/136
AMC CS49199/9364.7/61.5136/6169/2936/1660/22
ALKK-PCI Registry45487/1,42667.7/69.9372/970359/1,084159/462171/549
Dziewierz et al4630/2164.5*/72*25/8NA/NA8/18/4
Ohman et al1685/72558/5669/58040/29717/10221/87
Mahmoudi et al4470/70~59/60.6~71.4%/69.2%~73.9%/80%~21.9%/27%~12.4%/9.6%
Brodie et al36213/1,277NA/NA126/916NA/NA23/8556/229
Kumbasar et al3725/2053.4/53.522/179/1010/5NA/NA

Note:

Median.

Abbreviations: AMI, acute myocardial infarction; IABP, intra-aortic balloon pump; MI, myocardial infarction; NA, not available.

Primary efficacy endpoint

In-hospital mortality

As shown in the Figure 2A, the overall risk of in-hospital mortality was not reduced significantly following IABP use (OR 0.87, 95% CI: 0.59–1.28, P=0.479; P<0.001, I2=90.6%); similar results were observed for both the RCTs and Obs (RCT: OR 0.89, 95% CI: 0.54–1.49, P=0.669; P=0.298, I2=15.7%; Obs: OR 0.84, 95% CI: 0.52–1.35, P=0.467; P<0.001, I2=94.4%). The results of the Egger’s test indicted that no publication bias was encountered (P=0.406, 0.325, 0.175 for Obs, RCT, and overall, respectively).
Figure 2

Forest plots of the primary efficacy endpoint of the included trials.

Notes: The odds ratio (OR) of in-hospital all-cause mortality (A), midterm all-cause mortality (B), and long-term all-cause mortality (C), associated with IABP use compared with no IABP use, stratified by different dual regimens. Weights are from random effects analysis.

Abbreviations: CI, confidence interval; IABP, intra-aortic balloon pump; RCT, randomized controlled trial; ID, identification.

Midterm mortality

The comparison between IABP use and no IABP use demonstrated no significant differences with respect to the risk of midterm mortality (OR 1.12, 95% CI: 0.53–2.38, P=0.768; P<0.001, I2=93.4%; Figure 2B), as well as no significant reductions in risk in the RCTs and Obs (RCT: OR 0.84, 95% CI: 0.43–1.64, P=0.609; P=0.122, I2=45.0%; Obs: OR 1.18, 95% CI: 0.40–3.47, P=0.760; P<0.001, I2=95.5%). No publication bias was observed based on the results of the Egger’s test (P=0.135, 0.843, 0.813 for Obs, RCT, and overall, respectively).

Long-term mortality

A significantly lower incidence of long-term mortality rate was observed in the IABP group compared with the control group (OR 0.66, 95% CI: 0.48–0.91, P=0.010; P=0.025, I2=47.4%; Figure 2C), without publication bias (Egger’s test: P=0.132). A similar result was observed in the RCTs (OR 0.71, 95% CI: 0.50–0.99, P=0.046; P=0.177, I2=30.2%; Egger’s test: P=0.191), and no significant difference was observed in Obs (OR 0.60, 95% CI: 0.31–1.19, P=0.145; P=0.012, I2=68.9%; Egger’s test: P=0.359). A sensitivity analysis confirmed the beneficial effects of IABP with respect to the long-term mortality.

Secondary efficacy endpoints

Reinfarction

The impact of IABP insertion did not significantly differ from no IABP use with respect to the reduction of risk of reinfarction during hospitalization (OR 1.10, 95% CI: 0.68–1.78, P=0.706; P=0.094, I2=34.3%; Figure 3A), and the results from the RCTs and Obs were similar (Obs: OR 1.25, 95% CI: 0.48–3.25, P=0.644; P=0.027, I2=55.6%; RCT: OR 0.91, 95% CI: 0.60–1.40, P=0.680; P=0.580, I2=0.0%). No publication bias was observed for reinfarction (Egger’s test: P=0.548, 0.667, 0.837 for Obs, RCT, and overall, respectively).
Figure 3

Forest plots of the secondary efficacy endpoint of the included trials.

Notes: The odds ratio (OR) of reinfarction (A), recurrent ischemia (B), and new heart failure (C) associated with IABP use versus no IABP use stratified by different dual regimens. Weights are from random effects analysis.

Abbreviations: CI, confidence interval; IABP, intra-aortic balloon pump; RCT, randomized controlled trial; ID, identification.

Recurrent ischemia

As depicted in Figure 3B, there was no significant difference between IABP use and no IABP use with respect to the risk of in-hospital ischemia recurrence (OR 0.87, 95% CI: 0.36–2.12, P=0.754; P=0.001, I2=75.7%), in either the RCTs or Obs (RCT: OR 0.49, 95% CI: 0.18–1.31, P=0.155; P=0.060, I2=59.4%; Obs: OR 2.31, 95% CI: 1.20–4.46, P=0.013; P=0.744, I2=0.0%). No publication bias was observed (P=0.855, 0.836 for RCTs and overall, respectively).

New heart failure

The overall incidence of new heart failure was similar between the two subgroups (OR 1.40, 95% CI: 0.72–2.72, P=0.320; P<0.001, I2=82.3%; Figure 3C), as well as in the RCTs (OR 0.89, 95% CI: 0.63–1.28, P=0.541; P=0.404, I2=2.0%), whereas the risk of new heart failure increased significantly in the Obs (OR 3.29, 95% CI: 1.29–8.38, P=0.013; P=0.005, I2=87.5%). The results from the Egger’s test suggested no publication bias regarding the incidence of new heart failure (P=0.875, 0.866 for RCTs and overall, respectively).

Safety endpoints

Severe bleeding requiring blood transfusion

The incidence of severe bleeding requiring blood transfusion was higher in the IABP group than in control group (OR 2.05, 95% CI: 1.29–3.25, P=0.002; P=0.001, I2=78.5%; Figure 4A); a similar result was observed in the Obs (OR 3.48, 95% CI: 2.09–5.79, P<0.001; P=0.003, I2=65.1%), whereas no significant difference was observed in RCTs (OR 1.14, 95% CI: 0.87–1.49, P=0.338; P=0.804, I2=0.0%). The Egger’s test was not suggestive of publication bias (P=0.582, 0.640, 0.880 for Obs, RCTs, and overall, respectively). The sensitivity analysis demonstrated that the inferior effects of IABP insertion in the setting of AMI on severe bleeding requiring blood transfusion were always observed by omitting a single study at a time.
Figure 4

Forest plots of the safety endpoints of the included trials.

Notes: The odds ratio (OR) of severe bleeding requiring blood transfusion (A) and stroke (B) associated with IABP use versus no IABP use stratified by different dual regimens. Weights are from random effects analysis.

Abbreviations: CI, confidence interval; IABP, intra-aortic balloon pump; RCT, randomized controlled trial; ID, identification.

Stroke

IABP usage was associated with a higher in-hospital incidence of stroke (OR 1.71, 95% CI: 1.04–2.82, P=0.035, Figure 4B) without any heterogeneity (P=0.636, I2=0.0%), which contrasted with the results of RCTs and Obs (OR 1.70, 95% CI: 0.73–3.97, P=0.220; P=0.302, I2=17.3% for RCTs; Obs: OR 1.72, 95% CI: 0.93–3.19, P=0.085; P=0.704, I2=0.0% for Obs). No publication bias was noted via Egger’s test (P=0.662, 0.142, 0.248 for Obs, RCTs, and overall, respectively).

Subgroup analysis

In the subgroup of patients suffering from AMI complicated by CS, significantly lower risks of in-hospital and long-term mortality associated with IABP use versus no IABP use were noted (OR 0.68, 95% CI: 0.46–0.99, P=0.045, Figure 5A; OR 0.64, 95% CI: 0.41–0.996; P=0.048, Figure 5B), but no significant difference compared with midterm mortality was observed (OR 0.86, 95% CI: 0.56–1.35, P=0.521).
Figure 5

Forest plots of the subgroup analysis of the included trials.

Notes: The odds ratio (OR) of in-hospital all-cause mortality (A) and long-term all-cause mortality (B) among the patients suffering from AMI complicated by CS, as well as midterm all-cause mortality (C) and the long-term mortality associated with different opportunities of IABP insertion vs no IABP use (D) according to reperfusion strategy and IABP insertion associated with IABP use versus no IABP use stratified by different dual regimens. (C) #: Different groups of patients from the TACTICS trial received the corresponding reperfusion; ##: different groups of patients from the GUSTO-I trial received the corresponding reperfusion. Weights are from random effects analysis.

Abbreviations: AMI, acute myocardial infarction; CI, confidence interval; ERV, emergency revascularization; TT, thrombolytic therapy; CS, cardiogenic shock; IABP, intra-aortic balloon pump; ID, identification; NA, not available.

In the subgroup of reperfusion, the combination of IABP and TT alone resulted in a lower midterm mortality rate (OR 0.42, 95% CI: 0.23–0.76, P=0.004, Figure 5C), although no significant differences were observed with respect to in-hospital and long-term mortality (in-hospital: OR 0.55, 95% CI: 0.21–1.39, P=0.206; long-term: OR 0.57, 95% CI: 0.28–1.17, P=0.128). Moreover, the combination of IABP and ERV alone was associated with a lower long-term mortality (OR 0.71, 95% CI: 0.49–1.03, P=0.072). In the analysis pertaining to the opportunity for IABP insertion, a reduced risk of long-term mortality was observed among the patients suffering AMI for whom IABP was inserted before reperfusion compared with no IABP use (OR 0.49, 95% CI: 0.34–0.71, P<0.001, Figure 5D), although no significant difference was observed between no IABP use and IABP insertion after reperfusion (OR 0.90, 95% CI: 0.35–2.31, P=0.825).

Discussion

The major finding of this comprehensive meta-analysis was that IABP insertion was associated with reduced long-term mortality in patients suffering from AMI compared with no IABP use at the cost of potential high risk of stroke and severe bleeding requiring blood transfusion. The subgroup analyses demonstrated that IABP insertion before reperfusion may be an optimal treatment for patients suffering from AMI complicated by CS or patients receiving TT. Since IABP was first reported for clinical application in 1968,4 it has been widely used in patients suffering from AMI, particularly patients following AMI complicated by CS.2,50 Both the ACC/AHA (2012)5 and the European Society of Cardiology guidelines6 strongly recommended IABP as a bridge to reperfusion for patients suffering from AMI complicated by CS, recommendations derived from several observational clinical trials.31,34,39 However, these recommendations have been challenged because of several recent meta-analyses12 and RCTs,7–10,19 which demonstrated that IABP for patients with AMI complicated by CS was not associated with reduced mortality, but with high potential risks of major bleeding and stroke. The IABP-SHOCK II trial, a randomized, open-label, multicenter trial involving 600 patients with CS following AMI who underwent early revascularization, demonstrated that IABP did not increase either 6- or 12-month survival rates compared with no IABP use.7,8 Moreover, the IABP SHOCK trial28 demonstrated that IABP insertion exerted only modest or even no effects on the acute physiology and chronic health evaluation II score as a marker of disease severity, improvements in cardiac index, reduced inflammation, or reduced plasma brain natriuretic peptide levels compared with medical therapy alone. In summary, the current data did not support IABP as a routine treatment for patients with AMI regardless of whether the patients suffered from AMI complicated by CS, which called the above-mentioned guidelines into question. On the other hand, all these over-mentioned meta-analyses for IABP application did not include all available relevant citations and there was none of them grouped meticulously which might cover the beneficial efficacy of IABP on special patients. Therefore, we carried out this updated, comprehensive meta-analysis to explore which patients could benefit mostly from IABP insertion. In spite of the advances in early revascularization,28 the mortality rate of patients suffering from AMI complicated by CS remains high; IABP was empirically preferred with respect to the treatment of these patients. The thrombolysis and counterpulsation to improve survival in myocardial infarction (TACTICS) trial19 demonstrated that IABP insertion for patients with AMI complicated by either hypotension or severe heart failure who were receiving thrombolysis was not associated with a significant risk reduction on 6-month mortality, but was associated with increased survival rates for patients in Killip classes III or IV (39% for combined therapy versus 80% for fibrinolysis alone). The data from the present study demonstrated the superior effects of IABP in patients suffering from AMI, particularly those complicated with CS, findings consistent with results of two previous meta-analyses.13,14 IABP may be the optimal choice for specific patients with AMI, but not for all patients with AMI. In fact, the observed reductions in mortality secondary to IABP usage may be balanced by the increased rates of stroke and severe bleeding. More and more evidence, including the results from our study, has demonstrated that IABP insertion is associated with a higher incidence of severe bleeding and stroke. Four risk factors for complications following IABP insertion were identified from the Benchmark Registry:51 age ≥75 years, female sex, peripheral arterial disease, and body surface area <1.65 m2. In the meta-analysis published by Sjauw et al,13 a significantly increased rate of stroke secondary to IABP use was observed in patients suffering from AMI. Restricting activity due to IABP insertion for long periods may promote the development of venous thrombosis, resulting in an increased risk of all-cause death among affected patients. Moreover, several studies52,53 have demonstrated that major bleeding was associated with an increased risk of death, indicating that bleeding was dangerous not only because of the hemorrhaging itself but also because it forced the discontinuation of the IABP and necessary antiplatelet therapy, resulting in higher rates of thrombotic events. Thiele et al7,8 pointed out that the higher in-hospital mortality in the AMI patients complicated with CS might depend on hemodynamic deterioration, multiorgan dysfunction, or systemic inflammatory response syndrome. One more possible related point was that some of the included patients were too serious to receive the implantation of IABP before their deaths or even had no access to implant this equipment in time if they were initially treated in a peripheral center. In contrast, the lower long-term mortality among these patients might be due to the successful insertion of IABP which positively affected myocardial recovery following AMI. Therefore, it may be more useful to determine which patients benefit most from IABP insertion, as opposed to studying the complications of IABP use. In contrast to the recommendations of the ACC/AHA (2012) guidelines, the SHOCK trial28 demonstrated that a higher rate of in-hospital mortality was observed among the patients suffering from AMI complicated by CS associated with IABP insertion following reperfusion compared with no IABP use (36.8% vs 28.6%). However, the balloon pump-assisted coronary intervention study (BCIS-1) trial10 demonstrated that IABP insertion before reperfusion may result in a significantly reduced risk of long-term mortality among these patients. Therefore, a subgroup analysis was conducted in our study to address this controversy and determine the ideal opportunity for IABP insertion, as well as which patients may benefit most from IABP insertion. IABP usage was associated with a lower risk of mortality when inserted before reperfusion, a finding consistent with those of previous studies.10,27 Early IABP insertion promoted hemodynamic stability and reduced myocardial oxygen demand, which may be important in patients suffering from AMI complicated by CS.2 Our results indicated that IABP insertion was associated with feasible benefits with respect to long-term survival in patients with AMI, particularly patients suffering from AMI complicated by CS and patients receiving thrombolytic therapy. Several questions remained unanswered. First, there were not enough data to assess the optimal duration of IABP insertion for patients without severe complications, as well as for which patients the IABP may be removed at an earlier time. Most of current literature regarding the efficacy of IABP use consented to a duration of ~48 hours, although no absolute benefits regarding survival rates were demonstrated. A second dilemma involved the ideal opportunity for IABP insertion in patients with AMI complicated by hemodynamic compromise. The ACC/AHA (2012) guidelines recommend implantation at the completion of reperfusion.5 Third, divergence regarding specifics about the reperfusion strategies was observed. Both ERV alone and TT alone demonstrated potential benefits in patients with AMI in this meta-analysis. Therefore, powerful randomized clinical trials comparing IABP use and no IABP use in patients suffering from AMI complicated by CS focusing on the optimal duration of IABP, most suitable time for IABP insertion, and more precise reperfusion strategies must be performed to guide clinical decision making.

Limitations

This study had several limitations. First, this meta-analysis was not based on individual patient data, and the small sample size of several included RCTs also made the evaluation of IABP’s efficacy easily influenced. Second, different studies used different endpoint definitions, which may have been the source of bias. Third, the lack of other potential confounding factors, such as the time of IABP insertion, duration of IABP placement, insertion details, including technology and choice of sheath with different sizes, as well as reperfusion strategies, did not allow us to explore the effects of IABP on patients with AMI or the underlying mechanisms of these effects. Additionally, there was no uniform or clear follow-up period.

Conclusion

IABP insertion is associated with feasible benefits with respect to long-term survival in patients with AMI, particularly those suffering from CS and receiving thrombolytic therapy, at the cost of higher risk of severe bleeding and stroke.
  52 in total

1.  Intra-aortic balloon pump improves clinical prognosis and attenuates C-reactive protein level in acute STEMI complicated by cardiogenic shock.

Authors:  Jun Gu; Wei Hu; Hongbing Xiao; Xiaodi Feng; Yueguang Chen; Dadong Zhang
Journal:  Cardiology       Date:  2010-10-05       Impact factor: 1.869

2.  Intra-aortic balloon counterpulsation in cardiogenic shock. Report of a co-operative clinical trial.

Authors:  S Scheidt; G Wilner; H Mueller; D Summers; M Lesch; G Wolff; J Krakauer; M Rubenfire; P Fleming; G Noon; N Oldham; T Killip; A Kantrowitz
Journal:  N Engl J Med       Date:  1973-05-10       Impact factor: 91.245

3.  Use and impact of intra-aortic balloon pump on mortality in patients with acute myocardial infarction complicated by cardiogenic shock: results of the Euro Heart Survey on PCI.

Authors:  Uwe Zeymer; Timm Bauer; Christian Hamm; Ralf Zahn; Franz Weidinger; Ricardo Seabra-Gomes; Matthias Hochadel; Jean Marco; Anselm Gitt
Journal:  EuroIntervention       Date:  2011-08       Impact factor: 6.534

Review 4.  The outcome of intra-aortic balloon pump support in acute myocardial infarction complicated by cardiogenic shock according to the type of revascularization: a comprehensive meta-analysis.

Authors:  Francesco Romeo; Maria Cristina Acconcia; Domenico Sergi; Alessia Romeo; Saverio Muscoli; Serafina Valente; Gian Franco Gensini; Flavia Chiarotti; Quintilio Caretta
Journal:  Am Heart J       Date:  2013-03-26       Impact factor: 4.749

5.  Use of intraaortic balloon counterpulsation in patients presenting with cardiogenic shock: observations from the GUSTO-I Study. Global Utilization of Streptokinase and TPA for Occluded Coronary Arteries.

Authors:  R D Anderson; E M Ohman; D R Holmes; I Col; A L Stebbins; E R Bates; R J Stomel; C B Granger; E J Topol; R M Califf
Journal:  J Am Coll Cardiol       Date:  1997-09       Impact factor: 24.094

6.  Concomitant use of intraaortic balloon counterpulsation and streptokinase in acute anterior myocardial infarction.

Authors:  S D Kumbasar; E Semiz; O Sancaktar; S Yalçinkaya; C Ermiş; N Değer
Journal:  Angiology       Date:  1999-06       Impact factor: 3.619

7.  The use of intraaortic balloon pumping as an adjunct to reperfusion therapy in acute myocardial infarction. The Thrombolysis and Angioplasty in Myocardial Infarction (TAMI) Study Group.

Authors:  E M Ohman; R M Califf; B S George; P J Quigley; D J Kereiakes; L Harrelson-Woodlief; R J Candela; C Flanagan; R S Stack; E J Topol
Journal:  Am Heart J       Date:  1991-03       Impact factor: 4.749

8.  Aortic counterpulsation may improve late patency of the occluded coronary artery in patients with early failure of thrombolytic therapy.

Authors:  T Kono; H Morita; T Nishina; M Fujita; H Onaka; Y Hirota; K Kawamura; A Fujiwara
Journal:  J Am Coll Cardiol       Date:  1996-10       Impact factor: 24.094

9.  A randomized comparison of intra-aortic balloon pumping after primary coronary angioplasty in high risk patients with acute myocardial infarction.

Authors:  A W van 't Hof; A L Liem; M J de Boer; J C Hoorntje; H Suryapranata; F Zijlstra
Journal:  Eur Heart J       Date:  1999-05       Impact factor: 29.983

10.  In patients with ST-segment elevation myocardial infarction with cardiogenic shock treated with percutaneous coronary intervention, admission glucose level is a strong independent predictor for 1-year mortality in patients without a prior diagnosis of diabetes.

Authors:  Marije M Vis; Krischan D Sjauw; René J van der Schaaf; Jan Baan; Karel T Koch; J Hans DeVries; Jan G P Tijssen; Robbert J de Winter; Jan J Piek; José P S Henriques
Journal:  Am Heart J       Date:  2007-09-12       Impact factor: 4.749

View more
  1 in total

Review 1.  Stroke After Coronary Artery Bypass Grafting and Percutaneous Coronary Intervention: Incidence, Pathogenesis, and Outcomes.

Authors:  Mario Gaudino; Dominick J Angiolillo; Antonino Di Franco; Davide Capodanno; Faisal Bakaeen; Michael E Farkouh; Stephen E Fremes; David Holmes; Leonard N Girardi; Sunao Nakamura; Stuart J Head; Seung-Jung Park; Michael Mack; Patrick W Serruys; Marc Ruel; Gregg W Stone; Derrick Y Tam; Michael Vallely; David P Taggart
Journal:  J Am Heart Assoc       Date:  2019-06-27       Impact factor: 5.501

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.