The evidence for clinically significant bias in plasma glucose between liquid and lyophilized citrate buffer additive.

Gordana Juricic1, Andrea Saracevic2, Lara Milevoj Kopcinovic2, Ana Bakliza3, Ana-Maria Simundic4.   

Abstract

OBJECTIVES: Citrate buffer additive has been suggested to be of supreme performance in inhibiting glycolysis. However, there is little evidence in the literature regarding the comparability of glucose concentrations in liquid and lyophilized citrate buffer containing tubes. The aim of this study was to compare glucose concentrations in tubes containing liquid (Glucomedics) and lyophilized citrate buffer (Terumo VENOSAFE™ Glycemia) additive, measured immediately after centrifugation. DESIGN AND METHODS: Blood was collected from forty volunteers into both Glucomedics and Venosafe Glycemia tubes. Blood was centrifuged within 15min from venipuncture and glucose concentration was measured immediately after centrifugation, on the Abbott Architect analyzer. Differences between glucose concentrations in Glucomedics and Terumo tubes were tested using the paired t-test. Mean bias was calculated and compared to recommended quality specification for glucose (i.e. 2.2%).
RESULTS: Glucose concentration in Terumo tubes was 3.4% lower than in Glucomedics tubes (P<0.001). The mean bias was clinically significant.
CONCLUSIONS: There is a clinically significant difference between glucose concentrations in liquid and lyophilized citrate buffer additive tubes (Glucomedics vs. Terumo tubes) measured immediately after centrifugation. This difference may affect the patient outcome due to the misclassification of diabetes. Copyright Â
© 2016 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Diabetes; Glucomedics; Glycolysis inhibitors; Preanalytical phase; Venosafe Glycemia

Mesh:

Substances:

Year:  2016        PMID: 27040902     DOI: 10.1016/j.clinbiochem.2016.03.006

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  1 in total

1.  Preanalytical issues related to routine and diagnostic glucose tests: Results from a survey in Spain.

Authors:  Isabel García-Del-Pino; Josep M Bauça; Carolina Gómez; Andrea Caballero; María Antonia Llopis; Mercedes Ibarz; Débora Martínez; Montserrat Ventura; Itziar Marzana; Juan J Puente; Marta Segovia; Paloma Salas; Rubén Gómez-Rioja
Journal:  Biochem Med (Zagreb)       Date:  2019-12-15       Impact factor: 2.313

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.