Literature DB >> 27040828

Forced oscillation, integer and fractional-order modeling in asthma.

Alvaro C D Faria1, Juliana Veiga1, Agnaldo J Lopes2, Pedro L Melo3.   

Abstract

The purpose of this study was to evaluate the use of fractional-order (FrOr) modeling in asthma. To this end, three FrOr models were compared with traditional parameters and an integer-order model (InOr). We investigated which model would best fit the data, the correlation with traditional lung function tests and the contribution to the diagnostic of airway obstruction. The data consisted of forced oscillation (FO) measurements obtained from healthy (n=22) and asthmatic volunteers with mild (n=22), moderate (n=19) and severe (n=19) obstructions. The first part of this study showed that a FrOr was the model that best fit the data (relative distance: FrOr=4.3±2.4; InOr=5.1±2.6%). The correlation analysis resulted in reasonable (R=0.36) to very good (R=0.77) associations between FrOr parameters and spirometry. The closest associations were observed between parameters related to peripheral airway obstruction, showing a clear relationship between the FrOr models and lung mechanics. Receiver-operator analysis showed that FrOr parameters presented a high potential to contribute to the detection of the mild obstruction in a clinical setting. The accuracy [area under the Receiver Operating Characteristic curve (AUC)] observed in these parameters (AUC=0.954) was higher than that observed in traditional FO parameters (AUC=0.732) and that obtained from the InOr model (AUC=0.861). Patients with moderate and severe obstruction were identified with high accuracy (AUC=0.972 and 0.977, respectively). In conclusion, the results obtained are in close agreement with asthma pathology, and provide evidence that FO measurement associated with FrOr models is a non-invasive, simple and radiation-free method for the detection of biomechanical abnormalities in asthma.
Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  Asthma; Constant phase element; Forced oscillation technique; Fractional-order model; Respiratory impedance; Respiratory system

Mesh:

Year:  2016        PMID: 27040828     DOI: 10.1016/j.cmpb.2016.02.010

Source DB:  PubMed          Journal:  Comput Methods Programs Biomed        ISSN: 0169-2607            Impact factor:   5.428


  6 in total

1.  Oscillation Mechanics, Integer and Fractional Respiratory Modeling in COPD: Effect of Obstruction Severity.

Authors:  Caroline Oliveira Ribeiro; Agnaldo José Lopes; Pedro Lopes de Melo
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2020-12-08

2.  Explainable machine learning methods and respiratory oscillometry for the diagnosis of respiratory abnormalities in sarcoidosis.

Authors:  Allan Danilo de Lima; Agnaldo J Lopes; Jorge Luis Machado do Amaral; Pedro Lopes de Melo
Journal:  BMC Med Inform Decis Mak       Date:  2022-10-20       Impact factor: 3.298

3.  Respiratory resistance and reactance in adults with sickle cell anemia: Part 2-Fractional-order modeling and a clinical decision support system for the diagnosis of respiratory disorders.

Authors:  Cirlene de Lima Marinho; Maria Christina Paixão Maioli; Jorge Luis Machado do Amaral; Agnaldo José Lopes; Pedro Lopes de Melo
Journal:  PLoS One       Date:  2019-03-07       Impact factor: 3.240

4.  Combined forced oscillation and fractional-order modeling in patients with work-related asthma: a case-control study analyzing respiratory biomechanics and diagnostic accuracy.

Authors:  Fábio Augusto D Alegria Tuza; Paula Morisco de Sá; Hermano A Castro; Agnaldo José Lopes; Pedro Lopes de Melo
Journal:  Biomed Eng Online       Date:  2020-12-09       Impact factor: 2.819

5.  Early Diagnosis of Respiratory Abnormalities in Asbestos-Exposed Workers by the Forced Oscillation Technique.

Authors:  Paula Morisco de Sá; Hermano Albuquerque Castro; Agnaldo José Lopes; Pedro Lopes de Melo
Journal:  PLoS One       Date:  2016-09-09       Impact factor: 3.240

6.  Forced oscillation technique for early detection of the effects of smoking and COPD: contribution of fractional-order modeling.

Authors:  Caroline Oliveira Ribeiro; Alvaro Camilo Dias Faria; Agnaldo José Lopes; Pedro Lopes de Melo
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2018-10-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.