A two-stage rotary blood pump design with potentially lower blood trauma: a computational study.

Bente Thamsen1, Ricardo Mevert1, Michael Lommel1, Philip Preikschat1, Julia Gaebler1, Thomas Krabatsch2, Ulrich Kertzscher1, Ewald Hennig2, Klaus Affeld1.   

Abstract

AIM: In current rotary blood pumps, complications related to blood trauma due to shear stresses are still frequently observed clinically. Reducing the rotor tip speed might decrease blood trauma. Therefore, the aim of this project was to design a two-stage rotary blood pump leading to lower shear stresses.
METHODS: Using the principles of centrifugal pumps, two diagonal rotor stages were designed with an outer diameter of 22 mm. The first stage begins with a flow straightener and terminates with a diffusor, while a volute casing behind the second stage is utilized to guide fluid to the outlet. Both stages are combined into one rotating part which is pivoted by cup-socket ruby bearings. Details of the flow field were analyzed employing computational fluid dynamics (CFD). A functional model of the pump was fabricated and the pressure-flow dependency was experimentally assessed.
RESULTS: Measured pressure-flow performance of the developed pump indicated its ability to generate adequate pressure heads and flows with characteristic curves similar to centrifugal pumps. According to the CFD results, a pressure of 70 mmHg was produced at a flow rate of 5 L/min and a rotational speed of 3200 rpm. Circumferential velocities could be reduced to 3.7 m/s as compared to 6.2 m/s in a clinically used axial rotary blood pump. Flow fields were smooth with well-distributed pressure fields and comparatively few recirculation or vortices. Substantially smaller volumes were exposed to high shear stresses >150 Pa.
CONCLUSIONS: Hence, blood trauma might be reduced with this design. Based on these encouraging results, future in vitro investigations to investigate actual blood damage are intended.

Entities:  

Mesh:

Year:  2016        PMID: 27034319     DOI: 10.5301/ijao.5000482

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  3 in total

1.  Scaling the Low-Shear Pulsatile TORVAD for Pediatric Heart Failure.

Authors:  Jeffrey R Gohean; Erik R Larson; Brian H Hsi; Mark Kurusz; Richard W Smalling; Raul G Longoria
Journal:  ASAIO J       Date:  2017 Mar/Apr       Impact factor: 2.872

2.  A Novel Toroidal-Flow Left Ventricular Assist Device Minimizes Blood Trauma: Implications of Improved Ventricular Assist Device Hemocompatibility.

Authors:  Carlo R Bartoli; Samson Hennessy-Strahs; Jeff Gohean; Maryann Villeda; Erik Larson; Raul Longoria; Mark Kurusz; Michael A Acker; Richard Smalling
Journal:  Ann Thorac Surg       Date:  2018-12-23       Impact factor: 5.102

3.  Hemolytic Performance in Two Generations of the Sputnik Left Ventricular Assist Device: A Combined Numerical and Experimental Study.

Authors:  Alexandra N Romanova; Alexander A Pugovkin; Maxim V Denisov; Ivan A Ephimov; Dmitry V Gusev; Marian Walter; Thomas Groth; Olga L Bockeria; Tatyana G Le; Anna S Satyukova; Sergey V Selishchev; Dmitry V Telyshev
Journal:  J Funct Biomater       Date:  2022-01-12
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.