Literature DB >> 27032932

Biomechanical performance of retrograde nail for supracondylar fractures stabilization.

Nattapon Chantarapanich1, Kriskrai Sitthiseripratip2, Banchong Mahaisavariya3, Pongwit Siribodhi4.   

Abstract

The study compared the biomechanical performance of retrograde nail used to stabilize supracondylar fracture (three different levels) by means of finite element analysis. Three different nail lengths (200, 260, and 300 mm) of stainless steel and titanium nails were under consideration. Intact femur model was reconstructed from Digital Imaging and Communications in Medicine images of Thai cadaveric femur scanned by computed tomography spiral scanner, whereas geometry of retrograde nail was reconstructed with the data obtained from three-dimensional laser scanner. The retrograde nail was virtually attached to the femur before nodes and elements were generated for finite element model. The finite element models were analyzed in two stages, the early stage of fracture healing and the stage after fracture healing. The finding indicated that purchasing proximal locking screw in the bowing region of the femur may be at risk due to the high stresses at the implant and bone. There were no differences in stress level, elastic strain at a fracture gap, and bone stress between stainless steel and titanium implant. Since the intramedullary canal requires reaming to accommodate the retrograde nail, the length of retrograde nail should be as long as necessary. However, in case that the retrograde nail can be accommodated into the intramedullary canal without reaming, the longer retrograde nail can be used.

Entities:  

Keywords:  Femoral fractures; Finite element analysis; Fracture fixation; Intramedullary nailing

Mesh:

Year:  2016        PMID: 27032932     DOI: 10.1007/s11517-016-1466-0

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  31 in total

1.  Unusual complications of supracondylar nails.

Authors:  A A Martínez; J M Pérez; A Herrera; J J Lallana
Journal:  Injury       Date:  2000-12       Impact factor: 2.586

2.  Finite element study of trochanteric gamma nail for trochanteric fracture.

Authors:  K Sitthiseripratip; H Van Oosterwyck; J Vander Sloten; B Mahaisavariya; E L J Bohez; J Suwanprateeb; R Van Audekercke; P Oris
Journal:  Med Eng Phys       Date:  2003-03       Impact factor: 2.242

3.  Postoperative weight-bearing after a fracture of the femoral neck or an intertrochanteric fracture.

Authors:  K J Koval; D A Sala; F J Kummer; J D Zuckerman
Journal:  J Bone Joint Surg Am       Date:  1998-03       Impact factor: 5.284

4.  Determination of muscle loading at the hip joint for use in pre-clinical testing.

Authors:  M O Heller; G Bergmann; J-P Kassi; L Claes; N P Haas; G N Duda
Journal:  J Biomech       Date:  2005-05       Impact factor: 2.712

5.  Retrograde intramedullary nailing in distal femoral fractures--results in a series of 46 consecutive operations.

Authors:  Lauri Handolin; Jarkko Pajarinen; Jan Lindahl; Eero Hirvensalo
Journal:  Injury       Date:  2004-05       Impact factor: 2.586

6.  Finite element comparison of retrograde intramedullary nailing and locking plate fixation with/without an intramedullary allograft for distal femur fracture following total knee arthroplasty.

Authors:  Shih-Hao Chen; Ming-Chieh Chiang; Ching-Hua Hung; Shang-Chih Lin; Hsiao-Wei Chang
Journal:  Knee       Date:  2013-04-10       Impact factor: 2.199

7.  Biomechanical analysis of retrograde intramedullary nail fixation in distal femoral fractures.

Authors:  Shih-Hao Chen; Tzai-Chiu Yu; Chih-Han Chang; Yuan-Chiao Lu
Journal:  Knee       Date:  2008-08-21       Impact factor: 2.199

8.  Closed retrograde retrieval of the distal broken segment of femoral cannulated intramedullary nail using a ball-tipped guide wire.

Authors:  Sreenivasulu Metikala; Riazuddin Mohammed
Journal:  Indian J Orthop       Date:  2011-07       Impact factor: 1.251

9.  Numerical investigations on the strain-adaptive bone remodelling in the periprosthetic femur: influence of the boundary conditions.

Authors:  Bernd-Arno Behrens; Ingo Nolte; Patrick Wefstaedt; Christina Stukenborg-Colsman; Anas Bouguecha
Journal:  Biomed Eng Online       Date:  2009-04-16       Impact factor: 2.819

10.  Retrograde nailing versus locked plating of extra-articular distal femoral fractures: comparison of 36 cases.

Authors:  Kanda Gao; Wei Gao; Jianhua Huang; Haoqing Li; Fan Li; Jie Tao; Qiugen Wang
Journal:  Med Princ Pract       Date:  2012-10-13       Impact factor: 1.927

View more
  3 in total

1.  Single lag screw and reverse distal femur locking compression plate for concurrent cervicotrochanteric and shaft fractures of the femur: biomechanical study validated with a clinical series.

Authors:  Surasak Jitprapaikulsarn; Nattapon Chantarapanich; Arthit Gromprasit; Chantas Mahaisavariya; Chawanan Patamamongkonchai
Journal:  Eur J Orthop Surg Traumatol       Date:  2021-01-08

2.  Application of a double reverse traction repositor in the retrograde intramedullary nailing of distal femur fractures.

Authors:  Xiaodong Lian; Kuo Zhao; Wei Chen; Junzhe Zhang; Junyong Li; Hongyu Meng; Zhiyong Hou; Yingze Zhang
Journal:  J Orthop Surg Res       Date:  2021-03-03       Impact factor: 2.359

3.  A systematic review of the use of titanium versus stainless steel implants for fracture fixation.

Authors:  Collin C Barber; Matthew Burnham; Ogaga Ojameruaye; Michael D McKee
Journal:  OTA Int       Date:  2021-08-18
  3 in total