Physical and composition characteristics of clinical secretions compared with test soils used for validation of flexible endoscope cleaning.

M J Alfa1, N Olson2.   

Abstract

AIM: To determine which simulated-use test soils met the worst-case organic levels and viscosity of clinical secretions, and had the best adhesive characteristics.
METHODS: Levels of protein, carbohydrate and haemoglobin, and vibrational viscosity of clinical endoscope secretions were compared with test soils including ATS, ATS2015, Edinburgh, Edinburgh-M (modified), Miles, 10% serum and coagulated whole blood. ASTM D3359 was used for adhesion testing. Cleaning of a single-channel flexible intubation endoscope was tested after simulated use.
RESULTS: The worst-case levels of protein, carbohydrate and haemoglobin, and viscosity of clinical material were 219,828μg/mL, 9296μg/mL, 9562μg/mL and 6cP, respectively. Whole blood, ATS2015 and Edinburgh-M were pipettable with viscosities of 3.4cP, 9.0cP and 11.9cP, respectively. ATS2015 and Edinburgh-M best matched the worst-case clinical parameters, but ATS had the best adhesion with 7% removal (36.7% for Edinburgh-M). Edinburgh-M and ATS2015 showed similar soiling and removal characteristics from the surface and lumen of a flexible intubation endoscope.
CONCLUSIONS: Of the test soils evaluated, ATS2015 and Edinburgh-M were found to be good choices for the simulated use of endoscopes, as their composition and viscosity most closely matched worst-case clinical material.
Copyright © 2016 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Cleaning; Clinical secretions; Gastrointestinal endoscopes; Simulated use; Validation

Mesh:

Substances:

Year:  2016        PMID: 27021399     DOI: 10.1016/j.jhin.2016.01.023

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  2 in total

1.  Turbulent fluid flow is a novel closed-system sample extraction method for flexible endoscope channels of various inner diameters.

Authors:  Seo Yean Sohn; Michelle J Alfa; Richard Lai; Yacoob Tabani; Mohamed E Labib
Journal:  J Microbiol Methods       Date:  2019-11-20       Impact factor: 2.363

2.  Sterile Reverse Osmosis Water Combined with Friction Are Optimal for Channel and Lever Cavity Sample Collection of Flexible Duodenoscopes.

Authors:  Michelle J Alfa; Harminder Singh; Zoann Nugent; Donald Duerksen; Gale Schultz; Carol Reidy; Patricia DeGagne; Nancy Olson
Journal:  Front Med (Lausanne)       Date:  2017-11-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.