Literature DB >> 27020966

Absorbed dose measurements in the build-up region of flattened versus unflattened megavoltage photon beams.

Annemieke De Puysseleyr1, Wolfgang Lechner2, Wilfried De Neve3, Dietmar Georg2, Carlos De Wagter3.   

Abstract

This study evaluated absorbed dose measurements in the build-up region of conventional (FF) versus flattening filter-free (FFF) photon beams. The absorbed dose in the build-up region of static 6 and 10MV FF and FFF beams was measured using radiochromic film and extrapolation chamber dosimetry for single beams with a variety of field sizes, shapes and positions relative to the central axis. Removing the flattening filter generally resulted in slightly higher relative build-up doses. No considerable impact on the depth of maximum dose was found.
Copyright © 2016. Published by Elsevier GmbH.

Keywords:  Ausgleichsfilterfreie Photonenstrahlung; Dosis in der Aufbauregion; Flattening filter-free; build-up dose; surface dose

Mesh:

Year:  2016        PMID: 27020966     DOI: 10.1016/j.zemedi.2016.02.005

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  1 in total

1.  Key clinical beam parameters for nanoparticle-mediated radiation dose amplification.

Authors:  Alexandre Detappe; Sijumon Kunjachan; Pascal Drané; Shady Kotb; Marios Myronakis; Douglas E Biancur; Thomas Ireland; Matthew Wagar; Francois Lux; Olivier Tillement; Ross Berbeco
Journal:  Sci Rep       Date:  2016-09-23       Impact factor: 4.379