| Literature DB >> 27020966 |
Annemieke De Puysseleyr1, Wolfgang Lechner2, Wilfried De Neve3, Dietmar Georg2, Carlos De Wagter3.
Abstract
This study evaluated absorbed dose measurements in the build-up region of conventional (FF) versus flattening filter-free (FFF) photon beams. The absorbed dose in the build-up region of static 6 and 10MV FF and FFF beams was measured using radiochromic film and extrapolation chamber dosimetry for single beams with a variety of field sizes, shapes and positions relative to the central axis. Removing the flattening filter generally resulted in slightly higher relative build-up doses. No considerable impact on the depth of maximum dose was found.Keywords: Ausgleichsfilterfreie Photonenstrahlung; Dosis in der Aufbauregion; Flattening filter-free; build-up dose; surface dose
Mesh:
Year: 2016 PMID: 27020966 DOI: 10.1016/j.zemedi.2016.02.005
Source DB: PubMed Journal: Z Med Phys ISSN: 0939-3889 Impact factor: 4.820