Validation of a phospholipase A2 receptor antibody ELISA in an Australian cohort with membranous glomerulonephritis.

Lawrence Ong1, Roger Silvestrini2, Jeremy Chapman3, David A Fulcher4, Ming Wei Lin4.   

Abstract

A commercial PLA2R Ab ELISA was validated by examining its ability to distinguish primary from secondary membranous nephropathy, correlating results with clinical markers of disease activity, and comparing its performance with an indirect immunofluorescence test (IIFT). PLA2R Ab levels were measured in 77 patients with biopsy proven membranous nephropathy, divided into either idiopathic (n = 61) or secondary groups (n = 6). In the idiopathic group, measures of contemporaneous disease activity (proteinuria, serum creatinine) were compared between seropositive and seronegative subjects. ELISA values were then compared with semi-quantitative results from an IIFT using PLA2R transfected HEK293 cells as substrate. The PLA2R Ab ELISA was positive in only 15 of 61 (25%) patients with idiopathic membranous nephropathy (IMN), but there was a significant negative relationship with time since diagnosis. Thus, in a subgroup of patients diagnosed within 6 months of analysis, the sensitivity was 6/15 (55%), rising to 6/8 (75%) in those recently-diagnosed patients who had not been treated. In the entire cohort, there was a significant positive correlation between ELISA values and degree of proteinuria, but our analysis did not control for variation of both variables with time. The PLA2R Ab ELISA also showed very high agreement with IIFT (96%). Therefore, the PLA2R Ab ELISA is a highly specific test for distinguishing primary from secondary membranous nephropathy that is most sensitive in newly diagnosed patients who have not received immunosuppression. Antibody levels correlated with degree of proteinuria, but this relationship was not shown to be independent of time. Both IIFT and ELISA platforms performed comparably.
Copyright © 2016 Royal College of Pathologists of Australasia. Published by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Phospholipase A2 receptor antibody; enzyme linked immunosorbent assay; indirect immunofluorescence test; membranous nephropathy; sensitivity; specificity

Mesh:

Substances:

Year:  2016        PMID: 27020499     DOI: 10.1016/j.pathol.2016.02.001

Source DB:  PubMed          Journal:  Pathology        ISSN: 0031-3025            Impact factor:   5.306


  2 in total

1.  Comparison of measurements of anti-PLA2R antibodies in Japanese patients with membranous nephropathy using in-house and commercial ELISA.

Authors:  Hajime Kaga; Atsushi Komatsuda; Soh Yamamoto; Tadashi Kikuchi; Mika Kamata; Akiko Sato; Masafumi Odaka; Shin-Ichi Yokota; Naoto Takahashi; Hideki Wakui
Journal:  Clin Exp Nephrol       Date:  2019-02-13       Impact factor: 2.801

2.  Diagnostic Test Accuracy of Serum Anti-PLA2R Autoantibodies and Glomerular PLA2R Antigen for Diagnosing Idiopathic Membranous Nephropathy: An Updated Meta-Analysis.

Authors:  Weiying Li; Yuliang Zhao; Ping Fu
Journal:  Front Med (Lausanne)       Date:  2018-04-26
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.