Preanalytics: The (in-)stability of volatile POCT parameters and the homogeneity of blood in syringes at the market.

Michael A Gruber1, Stefan Felbermeir2, Regina Lindner2, Martin Kieninger2.   

Abstract

BACKGROUND: In between calibrations, quality controls and so on reduce the availability of blood gas analyzers. Despite the intended urgency of the measurements, these inhibitory circumstances cause delays of measurements. The comparison of changes (primarily PO2 and PCO2, secondarily ctHb) in syringes from different suppliers and the embedding into guideline ranges are the objectives of this study.
METHODS: Five matrices were tonometered at two levels within the German guideline (RiLi-BAEK) controlled range. Matrices were aspired into each syringe and blood gas analysis accomplished after 0 to 30min. The procedure was repeated 5 to 8-fold for the gas concentrations and every condition which consisted of syringe type (7), matrices (5) and time delay (7).
RESULTS: Syringes produce almost identical effects on partial pressures during the 30min observation period. Initial fast changes arise when partial pressure (PP) differs widely from the atmospheric pressure and are followed by diffusion, gas production or consumption and ctHb depending effects.
CONCLUSIONS: Only normal arterial blood gases without "atmospheric pollution" are not affected by the syringe materials within 30min. The homogeneity of a de-aerated and capped ICU-patients' sample is hardly to be provided after more than 10min of resting without mechanical assistance.
Copyright © 2016 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Hemoglobin; Partial pressures; Point-of-care testing; Syringes; Tonometry

Mesh:

Substances:

Year:  2016        PMID: 27012513     DOI: 10.1016/j.cca.2016.03.010

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  2 in total

1.  Can we rely on out-of-hospital blood samples? A prospective interventional study on the pre-analytical stability of blood samples under prehospital emergency medicine conditions.

Authors:  Johannes Prottengeier; Nicola Jess; Frank Harig; Christine Gall; Joachim Schmidt; Torsten Birkholz
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2017-03-04       Impact factor: 2.953

2.  Stability of pH, Blood Gas Partial Pressure, Hemoglobin Oxygen Saturation Fraction, and Lactate Concentration.

Authors:  Ariadna Arbiol-Roca; Claudia Elizabeth Imperiali; Dolors Dot-Bach; José Valero-Politi; Macarena Dastis-Arias
Journal:  Ann Lab Med       Date:  2020-06-17       Impact factor: 3.464

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.