Literature DB >> 27007788

Accuracy of Pedicle Screw Placement in Children 10 Years or Younger Using Navigation and Intraoperative CT.

Tianyi D Luo1, David W Polly, Charles G Ledonio, Nicholas M Wetjen, A Noelle Larson.   

Abstract

STUDY
DESIGN: A consecutive case series.
OBJECTIVE: To determine the revision rate for pedicle screws placed using intraoperative CT and image-guided navigation in children 10 years or younger. SUMMARY OF BACKGROUND DATA: Screws are frequently used for spinal instrumentation in young children, although this is only by physician-directed use. This is a multicenter retrospective study of patients aged 10 years or younger, who underwent spinal screw instrumentation with image-guided navigation. We hypothesized that intraoperative navigation would result in a high rate of accuracy for screw placement.
METHODS: Between 2007 and 2013, 130 pedicle and 7 lateral mass screws were placed in 16 consecutive patients undergoing a total of 17 surgeries at 2 institutions. Mean age at surgery was 6.9 years (range, 0.8-10.9 y). Screws were placed using an open technique with intraoperative CT (O-arm) and image-guided navigation (Stealth). Procedures included: growing spine device (3), hemivertebrae excision (4), posterior fusion (7), cervical fusion (2), and vertebral column resection (1). Congenital deformity was the most common diagnosis. Primary outcome measures were need for intraoperative screw revision or complication associated with screw placement.
RESULTS: Mean number of screws used per procedure was 8.1 (range, 2-17). Screws were placed from C1 to L5. Of the 137 screws, 3 required revision to shorter screws for an overall accuracy rate of 97.8%. In 1 case, a right T3 screw was revised due to anterior penetration. In another case, left-sided T1 and T2 pedicle screws were shortened 5 mm because they had penetrated the anterior aspect of their respective vertebral bodies. There were no screw-related complications.
CONCLUSIONS: In this series, image-guided navigation resulted in accurate placement of screws in patients aged 10 years or younger with no associated intraoperative complications. The navigated accuracy rate (97.8%) is significantly higher (P=0.01) than the reported 90.9% pedicle screw accuracy rate without navigation in the same age group by Baghdadi and colleagues. Intraoperative CT and image guidance were useful in our practice for placement of screws in skeletally immature patients.

Entities:  

Mesh:

Year:  2016        PMID: 27007788     DOI: 10.1097/BSD.0000000000000230

Source DB:  PubMed          Journal:  Clin Spine Surg        ISSN: 2380-0186            Impact factor:   1.876


  4 in total

1.  Accuracy of robot-assisted pedicle screw insertion in adolescent idiopathic scoliosis: is triggered electromyographic pedicle screw stimulation necessary?

Authors:  K Aaron Shaw; Joshua S Murphy; Dennis P Devito
Journal:  J Spine Surg       Date:  2018-06

2.  Does the use of intraoperative CT scan improve outcomes in Orthopaedic surgery? A systematic review and meta-analysis of 871 cases.

Authors:  Vishal Kumar; Vishnu Baburaj; Sandeep Patel; Siddhartha Sharma; Raju Vaishya
Journal:  J Clin Orthop Trauma       Date:  2021-05-15

Review 3.  State of the art review of new technologies in spine deformity surgery-robotics and navigation.

Authors:  J Alex Sielatycki; Kristen Mitchell; Eric Leung; Ronald A Lehman
Journal:  Spine Deform       Date:  2021-09-06

4.  3D imaging and stealth navigation instead of CT guidance for radiofrequency ablation of osteoid osteomas: a series of 52 patients.

Authors:  Ran Ankory; Assaf Kadar; Doron Netzer; Haggai Schermann; Yair Gortzak; Shlomo Dadia; Yehuda Kollander; Ortal Segal
Journal:  BMC Musculoskelet Disord       Date:  2019-12-01       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.