Literature DB >> 27006327

Prediction accuracy of preinstalled formulas on 2 optical biometers.

David L Cooke1, Timothy L Cooke2.   

Abstract

PURPOSE: To evaluate how well partial coherence interferometry (PCI) (IOLMaster) and optical low-coherence reflectometry (OLCR) (Lenstar LS 900) predict postoperative refractions using only the formulas that come preinstalled on the machines.
SETTING: Private practice, Saint Joseph, Michigan, USA.
DESIGN: Retrospective consecutive case series.
METHODS: Eyes were measured with 2 biometers before cataract surgery. Six formulas were ranked by machine. Formulas were also ranked for extremely long and short eyes by averaging the ranks of 6 statistics (mean error, mean absolute error, standard deviation [SD], maximum error, and percentage of eyes within ±0.5 diopter [D] and ±1.0 D of prediction).
RESULTS: Formulas were ranked by the SD of the prediction errors. The OLCR device outperformed the PCI device using the preinstalled formulas. The Olsen formula performed the best (0.378) but was preinstalled on the OLCR device only. Other formulas had the following SDs (OLCR device first, PCI device second): Haigis (0.393, 0.401), Holladay 1 (0.408, 0.414), Hoffer Q (0.428, 0.432), SRK/T (0.433, 0.440), and SRK II (0.623, 0.633). Rankings for long eye were (first to last) were Olsen, Haigis, SRK/T, Hoffer Q, and Holladay 1. Rankings for short eyes were Olsen, Haigis, Holladay 1, SRK/T, and Hoffer Q.
CONCLUSIONS: The OLCR device outperformed the PCI device using the Olsen formula. The Olsen formula also ranked first for short eyes and long eyes. Other formulas performed about the same on both machines. The SRK II formula should be avoided.
Copyright © 2016 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2016        PMID: 27006327     DOI: 10.1016/j.jcrs.2015.11.040

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Comparison of the Barrett Universal II, Kane and VRF-G formulas with existing intraocular lens calculation formulas in eyes with short axial lengths.

Authors:  Oleksiy V Voytsekhivskyy; Larysa Tutchenko; Diogo Hipólito-Fernandes
Journal:  Eye (Lond)       Date:  2022-01-15       Impact factor: 3.775

2.  Ray Tracing versus Thin-Lens Formulas for IOL Power Calculation Using Swept-Source Optical Coherence Tomography Biometry.

Authors:  Reza Ghaffari; Parisa Abdi; Alireza Moghaddasi; Somayeh Heidarzadeh; Hossein Ghahvhechian; Maryam Kasiri
Journal:  J Ophthalmic Vis Res       Date:  2022-04-29

3.  An innovative approach for determining the customized refractive index of ectatic corneas in cataractous patients.

Authors:  Shiva Pirhadi; Keivan Maghooli; Khosrow Jadidi
Journal:  Sci Rep       Date:  2020-10-07       Impact factor: 4.379

  3 in total