| Literature DB >> 27003957 |
Zsolt Varga1, Adrian Nicholl1, Maria Wallenius1, Klaus Mayer1.
Abstract
This paper describes rapid methods for the determination of the production date (age dating) of plutonium (Pu) materials by inductively coupled plasma mass spectrometry (ICP-MS) for nuclear forensic and safeguards purposes. One of the presented methods is a rapid, direct measurement without chemical separation using 235U/239Pu and 236U/240Pu chronometers. The other method comprises a straightforward extraction chromatographic separation, followed by ICP-MS measurement for the 234U/238Pu, 235U/239Pu, 236U/240Pu and 238U/242Pu chronometers. Age dating results of two plutonium certified reference materials (SRM 946 and 947, currently distributed as NBL CRM 136 and 137) are in good agreement with the archive purification dates.Entities:
Keywords: Age dating; Nuclear forensics; Nuclear safeguards; Plutonium; Radiochronometry
Year: 2015 PMID: 27003957 PMCID: PMC4779457 DOI: 10.1007/s10967-015-4418-5
Source DB: PubMed Journal: J Radioanal Nucl Chem ISSN: 0236-5731 Impact factor: 1.371
Fig. 1Measured spectrum of the unspiked SRM 946 sample using the direct method. Note that for the isotope ratio evaluation the peak centroid of the isotopes of interest (flat-top peak) was selected
Fig. 2Measured spectrum of the unspiked SRM 946 sample after the chemical separation. Note that the signals at m/z = 239 and m/z = 240 were omitted due to the high intensity (about 2 × 106 and 2 × 105 cps, respectively), and for the isotope ratio evaluation the peak centroid of the isotopes of interest was selected
Measured age dating results of the plutonium certified reference materials using the different chronometers
| Chronometers | ||||
|---|---|---|---|---|
| 234U/238Pu | 235U/239Pu | 236U/240Pu | 238U/242Pu | |
| SRM946 | ||||
| Age determined by chemical separation (years) | 44.4 ± 1.2 | 44.56 ± 0.46 | 44.07 ± 0.60 | 150 ± 110 |
| Age measured by direct analysis (years) | NA | 44.46 ± 0.47 | 43.77 ± 0.61 | NA |
| SRM947 | ||||
| Age determined by chemical separation (years) | 42.8 ± 2.5 | 44.15 ± 0.11 | 44.36 ± 0.18 | 76 ± 55 |
| Age determined by direct analysis (years) | NA | 44.57 ± 0.73 | 44.36 ± 0.19 | NA |
Uncertainties are expressed as expanded uncertainties (k = 2) (reference date: 11 March 2015)
Calculated production date results of the plutonium certified reference materials for the different chronometers
| Chronometers | |||
|---|---|---|---|
| 234U/238Pu | 235U/239Pu | 236U/240Pu | |
| SRM946 | |||
| Production date by chemical separation | October 1970 ± 1.2 years | August 1970 ± 6 months | February 1971 ± 7 months |
| Production date determined by direct analysis | NA | September 1970 ± 6 months | June 1971 ± 7 months |
| SRM947 | |||
| Production date by chemical separation | May 1972 ± 2.5 years | 15 January 1971 ± 1 month | 30 October 1970 ± 2 months |
| Production date determined by direct analysis | NA | August 1970 ± 9 months | 30 October 1970 ± 2 months |
Uncertainties are expressed as expanded uncertainties (k = 2)
Fig. 3Comparison of the measured and reported production dates for SRM 946 standard [10, 17]. Uncertainties are expressed as reported (either as 95 % confidence level or as expanded uncertainties at k = 2)