Kevin Kang1, Robert Ferraro2, Richard Petrella3, Gurjaipal Kang4.
Abstract
Perclose Closure Device (Abbott-Vascular) was attempted for femoral artery (FA) access site closure by an experienced operator. The device was felt to snap as it was pushed through the previously scarred skin and subcutaneous tissue. Perclose had broken into two pieces with the proximal portion outside the body and the distal portion embolizing to lie in the FA to distal aorta. The left FA was accessed through 8F sheath and 4-8mm Basket snare passed through a 6F multipurpose guiding catheter. The tip of the broken Perclose was caught with the snare and pulled out. The Perclose has a distal flexible sheath that is overmolded directly onto the proximal rigid guide. This junction of sheath and guide may be a weak point. Accidental pressure applied at this weak point in any Perclose device during deployment can cause breakage. Awareness of this rare complication is important for safety.Entities:
Keywords: Femoral artery; Intraoperative complication; Vascular closure devices
Mesh:
Year: 2015 PMID: 26995426 PMCID: PMC4799013 DOI: 10.1016/j.ihj.2015.08.024
Source DB: PubMed Journal: Indian Heart J ISSN: 0019-4832
Fig. 1Long broken piece of Perclose device seen next to venous sheath.
Fig. 2Broken Perclose sheath in distal aorta grabbed with a snare.
Fig. 3Broken Perclose sheath being pulled out with snare.
Fig. 4Broken device at the junction of distal sheath with proximal guide.