| Literature DB >> 26994654 |
C Andrew Combs1, Alan Fishman2.
Abstract
Three steps must be followed to prevent the transmission of infection via a contaminated transvaginal ultrasound probe: cleaning the probe after every use, high-level disinfection, and covering the probe with a single-use barrier during the examination. There may be critical flaws in at least 2 of these steps as they are currently practiced. First, 2 widely used disinfectants, glutaraldehyde and orthophthalaldehyde, have recently been found to be ineffective at neutralizing human papilloma virus type 16 and type 18. Second, commercial ultrasound probe covers have an unacceptable rate of leakage (8-81%) compared to condoms (0.9-2%). We recommend the use of a sonicated hydrogen peroxide disinfectant system rather than aldehyde-type disinfectants. We recommend that the probe be covered with a condom rather than a commercial probe cover during transvaginal ultrasound examination. Combined with probe cleaning, these 2 steps are estimated to result in an 800 million- to 250 billion-fold reduction in human papilloma virus viral load, which should translate to greatly enhanced patient safety.Entities:
Keywords: disinfection; human papilloma virus; nosocomial infection; patient safety; sterilization; ultrasound safety; vaginal ultrasound
Mesh:
Substances:
Year: 2016 PMID: 26994654 DOI: 10.1016/j.ajog.2016.03.014
Source DB: PubMed Journal: Am J Obstet Gynecol ISSN: 0002-9378 Impact factor: 8.661