Tristan R A Lane1, Damian Kelleher1, Hayley M Moore1, Ian J Franklin1, Alun H Davies2.
Abstract
The Sapheon Venaseal Closure System (Sapheon Inc, Santa Rosa, Calif), using cyanoacrylate glue, has provided a new modality of treatment, with patients treated without both tumescent anesthesia and postoperative compression. We present the first case of great saphenous vein occlusion performed using glue while the patient was fully anticoagulated with warfarin. This was tolerated well, and the treated vein showed complete early occlusion at 8 weeks; however at 6 months, extensive recanalization was demonstrated on duplex imaging.Entities:
Year: 2013 PMID: 26992590 DOI: 10.1016/j.jvsv.2012.09.007
Source DB: PubMed Journal: J Vasc Surg Venous Lymphat Disord